5×10-4 … 1×10-3 mbar ±50% of reading 1×10-3 … 100 mbar ±15% of reading 100 … 950 mbar ±5% of reading 950 … 1050 mbar ±2.5% of reading
Repeatability (N2)
1×10-3 … 1100 mbar ±2% of reading
Output signal (measurement signal)
Voltage range 0 … +10.23 V Measurement range +0.61 … +10.23 V Error signal 0 V (default) Voltage vs. pressure 1.286 V/decade, logarithmic
Output impedance
2 × 4.7 Ω, short circuit-proof
Load impedance
>10 kΩ
Response time
<30 ms
Gauge identification
71.5 kΩ
HV adjustment
at <10-5 mbar
ATM adjustment
at >100 mbar
Switching functions
SP1, SP2 Setting range 5.0×10-5 … 1500 mbar Hysteresis 2) 10% of threshold Switching characteristics 2) Low Trip Point Type 1 floating contact (n.o.) per switching function Contact rating closed <30 VAC/DC, ≤1 A resistive open LED lit solid Switching time LED off <30 ms
1) Crossover range for air, O2, CO and N2 10 mbar, 100 mbar in heavy gases.
Table 2 Technical Specifications (continued)
Specifications
Value
Diagnostic port
Jack connector 2.5 mm, 3-pin
Supply
WARNING The gauge may only be connected to power supplies, instruments, or control devices that conform to the requirements of a grounded protective extra-low voltage (SELV). The connection to the gauge has to be fused. 3).
Supply voltage at the gauge
+15 … +30 VDC
Ripple
≤1 Vpp
Power consumption without fieldbus
≤2.5 W
Power consumption with fieldbus
≤3 W
Fuse to be connected 3)
1 AT
Electrical connection
Sensor cable
shielded 0.14 mm2/conductor
Cable length RS232C operation
≤100 m
Cable length FCC 68
≤30 m
Grounding concept Vacuum connection to signal common
See Paragraph "Power Connection” connected via 10 Ω, 10 nF
RS232C
Transmission rate
57600 baud (default)
Data format
binary 8 data bits one stop bit no parity bit no handshake
see Paragragh “Power Connection"
For further information on the RS232C interface, see Paragragh “Diagnostic Port (RS232C interface)”
Profibus interface
Specification, data format, communication protocol
RS485
Interface, physical
2) The hysteresis and the switching characteristics can be programmed via the serial interface or the diagnostic port. 3) Agilent controllers fulfill this requirement.
Table 4 LCD
LCD
Meaning
off
no supply voltage
lit green
measurement / parameter mode
lit red
error
The display can be rotated by 180 ° via the serial interface.
Instructions for Use
Раздел
Страница
About this manual
6
Validity
6
Definitions and terminology
7
Definition of Caution, Warning and Note
7
Warning Symbols
8
Safety
9
Proper use
9
Improper use
10
Safety guideline for PCG-750, PCG-752
11
Disposal
12
Service
13
Engineering Facts
12 facts · every claim linked to a page
Selection
The gauge covers 5×10-5 to 1500 mbar, so it suits both rough and fine vacuum measurement in one device.
Confirms the gauge spans the full required pressure range, avoiding two separate sensors.
Source
“designed for vacuum measurement of gases in the pressure range of 5×10-5 … 1500 mbar”
Safety
Do not use the gauge to measure flammable or combustible gases mixed with oxidants inside the explosion range.
Defines an absolute process-media restriction that must be checked before selection.
Source
“It must not be used for measuring flammable or combustible gases in mixtures containing oxidants”
Selection
Choose the PCG-752 nickel-plated filament version for corrosive applications instead of the tungsten PCG-750.
Drives the material choice between the two gauge variants for aggressive media.
Source
“recommended for corrosive applications”
Installation
The gauge must be electrically connected to a grounded vacuum chamber per EN 61010, using a conductive metallic clamping ring for KF flanges.
Protective grounding is mandatory for safe fault behavior and compliance.
Source
“Electrically connect the gauge to the grounded vacuum chamber.”
Installation
Connect the gauge only to SELV-compliant supplies and fuse the connection.
Defines the electrical supply safety requirement for wiring the gauge.
Source
“The gauge may only be connected to power supplies, instruments, or control devices”
Installation
Ground the cable shield on one side only via chassis ground and use a differential measurement input to avoid ground loops.
Prevents EMC and potential-difference errors that corrupt the measurement signal.
Source
“Connect the cable shield to ground on one side via the chassis ground. Do not”
Operation
Allow at least 10 minutes stabilization after power-up and preferably run the gauge continuously.
Sets realistic expectations for measurement readiness and drift behavior.
Source
“Allow a stabilization period of at least 10 minutes.”
Operation
The setpoints SP1 and SP2 are factory set to the lower measurement range limit and therefore do not switch until configured.
Prevents false assumption that relays work out of the box; must be programmed.
Source
“The setpoints SP1 and SP2 are factory set to the lower measurement range limit”
Operation
The Profibus variant must not be operated via the RS232C interface because it causes data transmission errors.
Avoids a wiring/commissioning mistake on Profibus units.
Source
“This gauge must not be operated with the RS232C interface.”
Maintenance
Zero drift can occur over time or from contamination, so periodically check and adjust the zero under the same ambient conditions and mounting orientation.
Establishes a routine maintenance task to keep accuracy within spec.
Source
“Periodically check the zero and adjust it if necessary”
Maintenance
Gauge failures from contamination or wear, and expendable parts like seals and filament, are not covered by warranty.
Informs the user about warranty limits and the need for careful handling.
Source
“Gauge failures due to contamination or wear and tear, as well as expendable parts”
Installation
The ceramic sensor can be damaged by impacts, so do not drop the product or allow shocks.
Handling precautions during mounting protect the sensitive diaphragm sensor.