No photo available yet

Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 280Z GFAAS

Agilent Technologies 280Z GFAAS — Atomic spectroscopy.

The Agilent 280Z GFAAS is a Zeeman graphite furnace atomic absorption spectrometer configured with a GTA 120 graphite tube atomizer and PSD 120 programmable sample dispenser, as documented for cadmium determination in wheat flour. It operates with an UltrAA cadmium hollow cathode lamp at 4 mA, 0.5 nm slit, and 228.8 nm wavelength, using 10 µL sample and 5 µL chemical modifier. This modification is the application-validated 280Z platform for trace Cd in food matrices; the sibling 240Z is the alternative series member.

Agilent Technologies Graphite Furnace Atomic Absorption Instruments Series — Model Comparison & Selection

Model Wavelength rangeSample volume, milliliterSoftwareElementsLamp current, mAInert gas purity, %
280Z GFAAS Current 228.80.01method-driven, guides usCd499.99
240Z GFAAS 283.30.025.5Pb1099.99

Full specifications

Core Performance & Functional Specifications
Wavelength range 228.8 nm
Analysis time seconds
Consumables none required
Elements Cd
Inert Gas Purity 99.99 %
Modifier Volume 5 μL
Sample volume 10 μL
Software method-driven, guides user in each step
Measurement mode mfr non-destructive multipoint mapping mapping
Instrument portability mfr portable, easy-to-use use
Matrix mfr Water Water
Chemical modifier mfr 4
Instrumentation mfr 240 AAS
Hollow Cathode Lamp mfr Cadmium Cadmium
Graphite tube platform mfr Omega Omega
Mode mfr Peak Area Area
Chemical modifier volume mfr 1 g/L
Lamp type mfr UltrAA Lamp Lamp
Replicates mfr 2
Wheat Flour mfr 5.22
rinse solution nitric acid concentration mfr 2 % v/v
SRC nitrogen pressurization mfr 40 bar
hydrogen peroxide concentration mfr 30
nitric acid concentration mfr 60
digest final volume mfr 20 mL
hot block reflux duration mfr 1 hour
water boil-off duration mfr 1 hour
de-ionized water resistivity mfr 18
Student t factor for MDL mfr 3.143
MDL confidence level mfr 99 % confidence
stability test interval mfr after every five samples samples
stability test duration mfr seven-hour period period
Dimensions, Weight & Mechanical Design
Slit Width 0.5 nm
Orientation dependence mfr independent of orientation of data collection co
sample weight for microwave digestion mfr 1.0 g
sample weight for wet digestion mfr 0.5 g
Electrical, Power & Drive Specifications
Lamp Current 4 mA
Interfaces, Connections & Control
Reflectance interface angle mfr 45 °
Operating Conditions, Environment & Protection
Ash Temperature (sample) 568 °C
Atomization Temperature 2100 °C
Atomize Temperature (sample) 1577 °C
Standards, Safety & General Data
Ash Temperature (standard) 570 °C
Atomize Temperature (standard) 1533 °C
Standard Solution 2.50 Cd
Standard Reference Material mfr 1 Cd
bulk standard Pb concentration mfr 100 μg/L
bulk standard Cd concentration mfr 5 μg/L
bulk standard Cr concentration mfr 20 μg/L
Pb MDL standard concentration mfr 0.3 ppb
Cd MDL standard concentration mfr 0.05 ppb
Cr MDL standard concentration mfr 0.1 ppb
Ni MDL standard concentration mfr 0.5 ppb
stability test Pb standard concentration mfr 5 standard

Description

Construction and series positioning

Behavior under load and in the furnace program

Control, installation and operation

Integration & Operating Notes

Process Limits
Inert gas must meet 99.99 % purity for the furnace program

The documented Cd method specifies 99.99 % inert gas purity. Lower-purity argon or nitrogen introduces oxygen and moisture into the graphite tube, affecting ash/atomization behavior and the blank; verify the gas supply certificate before running the temperature program.

Application note: Determination of Pb, Cd, Cr and Ni in Grains via Zeeman GFAAS
Control & I/O
Method-driven software sequences each furnace step

The software guides the operator through each step of the method rather than requiring manual program entry. Plan operator training around the method workflow, and confirm that the installed software revision matches the method file used for the grain application.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the 280Z GFAAS rather than the 240Z sibling for a trace-cadmium food method?

Choose the 280Z when the method is the documented Zeeman GFAAS grain workflow with GTA 120 atomizer and PSD 120 dispenser. The supplied facts give no differentiating specifications for the 240Z, so compare the two only against verified site requirements, not assumed performance differences.

What adjacent system parts does the documented 280Z configuration require?

The verified configuration pairs the 280Z with a GTA 120 Graphite Tube Atomizer and PSD 120 Programmable Sample Dispenser, an UltrAA cadmium hollow cathode lamp run at 4 mA, an Omega graphite tube platform, and 99.99 % inert gas. Electrical and interface requirements are not in the supplied facts.

What mounting and environment limits apply to the 280Z GFAAS?

The supplied facts contain no mounting, clearance, ambient temperature or humidity specifications for this model. The digest only notes that data collection is independent of orientation and describes a 45° reflectance interface angle; confirm installation limits from the site manual before siting.

Dimensional Drawings & Technical Documentation

Dimensional drawings

performance curve performance curve

Click drawing to inspect vector dimensions

Technical documentation

PDF
Determination of Pb, Cd, Cr and Ni in
Brochure ·511,3 КБ ·6 pp.