Agilent Technologies 5200 Fragment Analyzer
Agilent Technologies 5200 Fragment Analyzer — Automated electrophoresis.
The 5200 Fragment Analyzer is a 12-capillary automated electrophoresis system for sizing, quantification and integrity assessment of IVT mRNA and DNA fragments. It is the 12-capillary member of the Fragment Analyzer series, distinguished by exclusive support for the 22 cm effective / 47 cm total capillary array. It serves as the QC instrument for IVT mRNA drug substances and CRISPR/NGS fragment analysis in biopharmaceutical process analytics.
Agilent Technologies Fragment Analyzer Systems Series — Model Comparison & Selection
| Model | Measurement range | Plate capacity | Average sizing of 2 200 nt ivt mrna, nt | Lower marker size of hs rna kit, nt | Maximum ivt mrna sample size analyzed, nt | Sample volume per well, microliter |
|---|---|---|---|---|---|---|
| 5200 Fragment Analyzer Current | — | — | 2370 | 15 | 6000 | 2 |
| 5300 Fragment Analyzer | — | — | 2370 | 15 | 6000 | — |
| Fragment Analyzer Capillary Arrays | — | 12 | — | — | — | 2 |
| RNA QC, DNA QC, Fragment Analyzer | — | 12 | — | — | — | 2 |
| Ultra-high-throughput fragment analysis 5400 Fragment Analyzer | 0.5 | nine plates | 2370 | 15 | 6000 | — |
Full specifications
| Accuracy | 20 gram |
| Average Sizing Of 2,200 Nt Ivt Mrna | 2370 nt |
| Gel Volume | 10 mL |
| Intercalating Dye Volume | 1.0 µL |
| Lower Marker Size Of Hs Rna Kit | 15 nt |
| Maximum Ivt Mrna Sample Size Analyzed | 6000 nt |
| Measured Concentration | 431 pg/µL |
| Precision | 15 %CV |
| Sample Volume Per Well | 2 µL |
| Sizing range mfr | 35 bp |
| Sizing precision mfr | 5 %CV |
| Sizing accuracy mfr | 10 |
| Sample types analyzed mfr | IVT mRNA drug substances and LNP-encapsulated IV |
| Additional sample mfr | 1 |
| Instrument mfr | 5300 system |
| LNP disruption mfr | varying amounts of surfactant surfactant |
| Surfactant levels tested mfr | 20 |
| Sample concentration range mfr | 1 ng/μL |
| Extended method mfr | Method BE BE |
| Dilution ratios tested mfr | 1 |
| Recommended sample concentration range mfr | 1...100 ng/μL |
| Fragment Analyzer systems mfr | 5200 |
| DNA sizing range mfr | 100 bp |
| Separation resolution mfr | 10 |
| DNA sizing precision mfr | 2 % CV |
| DNA sizing accuracy mfr | 5 better |
| DNA fragment concentration range mfr | 0.005 DNA |
| Sizing precision (%CV) mfr | 5 %CV |
| Sizing accuracy (%error) mfr | 10 |
| IVT mRNA drug product sizes analyzed mfr | 2,200 nt |
| Surfactant concentrations tested for LNP disruption mfr | 20 |
| Sample concentration range assessed mfr | 1 ng/μL |
| Dilution ratios assessed with extended method mfr | 1 |
| Instrument model mfr | 5200 |
| DNA sizing precision1 mfr | 2 % CV |
| DNA sizing accuracy1 mfr | 5 better |
| Inlet Buffer fill volume per well mfr | 1.0 mL/well |
| Capillary Storage Solution fill volume per well mfr | 1.0 mL/well |
| Marker plate fill volume per well mfr | 30 μL/well |
| MDK DNA Ladder volume per well mfr | 24 μL |
| Library dilution concentration mfr | 500 pg/μl |
| Smear analysis region mfr | 200...1,000 bp |
| disruption time mfr | 20 minutes |
| sizing accuracy specification mfr | 15 % error |
| sizing precision specification mfr | 5 % CV |
| heat denaturation time mfr | 2 minutes |
| plate vortex duration mfr | two minutes minutes |
| pipette mixing volume mfr | 20 μL |
| diluent marker volume per well mfr | 22 μL |
| recommended sample concentration range mfr | 1...100 ng/μL |
| Sample sizes (nucleotide length) mfr | 2,200 nt |
| maximum peak height mfr | 24000 RFU |
| RNA kit (15 nt) part number mfr | -471 |
| standard sizing upper limit mfr | 6000 nt |
Description
Construction and series positioning
- The 5200 is a 12-capillary array instrument; the 22 cm effective / 47 cm total length array is available only on this 12-capillary configuration, giving 10 bp resolution at 500 bp versus 8 bp on the 33-55/55-80 arrays.
- Sibling models in the series: 5300 (5 pg/µL sensitivity, used for LNP-encapsulated IVT mRNA CQA work), 5400 (ultra-high-throughput, 0.5 ng/µL measurement range), and shared capillary arrays and RNA/DNA QC kits.
- The 5200 is the entry-throughput 12-capillary platform; the 5300 and 5400 extend sensitivity and throughput respectively.
Analytical behavior
- DNA sizing range 100–6,000 bp; IVT mRNA sizing up to 6,000 nt with the HS RNA kit (15 nt lower marker).
- Sizing precision 5 %CV, sizing accuracy 10 %; quantification precision 15 %CV, quantification accuracy 20 %.
- DNA fragment concentration range 0.005–2 ng/µL input DNA, adjustable by dilution; recommended IVT mRNA sample concentration 1–100 ng/µL.
- Gel/dye mixture scales with sample count: 1.0 µL intercalating dye in 10 mL gel for 12 samples; 4.5 µL dye in 45 mL gel for 96 samples.
Installation, control and operation
- Runs are queued in the Fragment Analyzer software: select Tray and Row for 12-Cap before entering Sample ID and Tray ID, then choose the method matching the installed array (CRP-910-22/33/55 for CRISPR Discovery).
- Row A of the 1x 930 dsDNA Inlet Buffer tray is filled at 1.0 mL/well and replaced daily; Capillary Storage Solution in row H at 1.0 mL/well is replaced every 2–4 weeks.
- Marker plate row A: 30 µL/well plus mineral oil overlay; MDK DNA Ladder is placed undiluted at 24 µL in well 12 of each analyzed row.
Integration & Operating Notes
Input DNA above 2 ng/µL saturates the specified concentration range of the CRISPR Discovery Gel Kit; dilute into range with the supplied dilution buffer (0.1x TE) so that sizing and quantification stay within the stated precision and accuracy.
In the Fragment Analyzer software, select the CRP-910-(22, 33 or 55) CRISPR Discovery method corresponding to the physically installed array; a mismatched method invalidates resolution and sizing results. For 12-capillary runs, select Tray and Row before entering Sample ID and Tray ID.
The stated 2 %CV sizing precision and ±5% or better sizing accuracy apply only when DNA ladder or DNA fragment standards are initially prepared in 1x TE Buffer; other diluents shift migration and void the accuracy claim.
Place 24 µL of ready-to-use MDK DNA Ladder undiluted into well 12 of each row of samples; plan plate layout so that well 12 is reserved, and fill marker plate row A at 30 µL/well with a mineral oil overlay.
Selection & Integration Q&A
Pick the 5200 for 12-capillary workloads needing the 22 cm effective / 47 cm array, which is exclusive to the 12-capillary system and gives 10 bp resolution at 500 bp. Choose the 5300 for higher sensitivity (5 pg/µL) or LNP-encapsulated IVT mRNA CQA, and the 5400 for ultra-high throughput.
Order a 12-capillary array in the 22, 33 or 55 variant plus the CRISPR Discovery Gel Kit (DNF-810 gel, DNF-600-U030 dye, DNF-355 inlet buffer, DNF-475 conditioning solution, markers, MDK ladder, dilution buffer, mineral oil). Capillary Storage Solution (GP-440-0100) is sold separately.
Fill row A of the 1x 930 dsDNA Inlet Buffer tray at 1.0 mL/well and replace daily; place Capillary Storage Solution in row H at 1.0 mL/well and replace every 2–4 weeks. Fill marker plate row A at 30 µL/well with mineral oil overlay and load 24 µL undiluted MDK ladder in well 12 of each row.
The Capillary Storage Solution (GP-440-0100) is sold separately from the CRISPR Discovery Gel Kit. Gel, intercalating dye, inlet buffer, capillary conditioning solution, markers, MDK DNA Ladder, dilution buffer and mineral oil are kit components; capillary arrays and RNA/DNA QC kits are ordered per application.