Agilent Technologies 572 Tungsten Bayard-Alpert Gauge
Agilent Technologies 572 Tungsten Bayard-Alpert Gauge — Vacuum technologies.
The Agilent 572 is a dual-tungsten-filament Bayard-Alpert ionization gauge tube with a Nonex envelope, measuring nitrogen-equivalent pressure from 1×10⁻³ down to its ~2×10⁻¹⁰ Torr X-ray limit. Its externally switchable filaments and resistance or electron-bombardment degassing distinguish it within the 571/572 series. It serves as the high-vacuum sensor in systems where a hot-filament gauge is acceptable.
Agilent Technologies Vacuum Transducers Series — Model Comparison & Selection
| Model | Measurement range | Atomic mass unit, grams | Electron charge, coulomb | Molar volume, liters | Max pressure, megapascal | Sensitivity |
|---|---|---|---|---|---|---|
| 572 Tungsten Bayard-Alpert Gauge Current | -3...2 | 1.66 | 1602 | 22.41 | -3...2 | — |
| 531 Thermocouple Gauge, Rough Vacuum Pressure Measurement | — | 1.66 | 1602 | 22.41 | — | — |
| 571 Bayard-Alpert Gauge Tube | -3...2 | 1.66 | 1602 | 22.41 | 1.33×10⁻⁴ | 17 |
| Cold Cathode Gauge Tube 525, High Vacuum (HV) Gauge | — | 1.66 | 1602 | 22.41 | — | — |
| Hot Filament Gauge MBA2-201, Ionization BA Gauge | — | 1.66 | 1602 | 22.41 | — | — |
| IMG-100 Inverted Magnetron Gauge | — | 1.66 | 1602 | 22.41 | — | — |
| IMG-300 UHV Inverted Magnetron Gauge | — | 1.66 | 1602 | 22.41 | — | — |
| UHV-24/UHV-24p | 1 | 1.66 | 1602 | 22.41 | — | 25 |
Full specifications
| Max pressure | -3...2 Torr |
| Measurement range | -3...2 Torr |
| Electron Charge | 1.602 coulomb |
| Internal Volume | 220 cc |
| Maximum operating pressure | 1 Torr |
| Molar Volume | 22.41 liters |
| Pumping Speed, Ionic | 0.06 liters/sec |
| X-ray Limit | 2 Torr |
| Acceleration due to gravity mfr | 9.8066 |
| Angstrom unit mfr | 10 m |
| Avogadro's number mfr | 6.02214076 |
| Boltzmann's constant mfr | 1.38064852 |
| Equivalent of heat mfr | 4.185 |
| Natural log base mfr | 2.7183 |
| Velocity of light mfr | 2.9979 |
| Velocity of sound mfr | 343 |
| Magnetic flux density mfr | 1 |
| Pumping speed conversion mfr | 1 L/s |
| Throughput conversion mfr | 1 cc/sec |
| Outgassing rate mfr | 2 |
| Envelope mfr | Nonex Nonex |
| Collector mfr | 0.010 diameter |
| Base leads mfr | 0.060 diameter |
| Collector lead mfr | 0.040 diameter |
| Shield mfr | Internally connected to filament filament |
| Tubulation mfr | 1.000 Nonex |
| Gauge constant S mfr | 1.5 |
| Tubulation (572-N) mfr | 1.000 Nonex |
| Gauge constant S (Helium) mfr | 1.5 |
| Gauge constant S (Nitrogen) mfr | 10 |
| Gauge constant S (Air) mfr | 10 |
| Base Leads mfr | 0.060 diameter |
| Collector Lead mfr | 0.040 diameter |
| K7360301 mfr | 572 tubulation |
| K7360302 mfr | 572 tubulation |
| K7360303 mfr | 572 tubulation |
| K7360304 mfr | 572 tubulation |
| K7360305 mfr | 572 |
| K7360306 mfr | 572 |
| K7360307 mfr | 572 F |
| Helium mfr | 1.5 Torr |
| Nitrogen mfr | 10 Torr |
| Air mfr | 10 Torr |
| Tubulation OD mfr | 1.000 |
| Atomic Mass Unit | 1.6605 grams |
| Shipping Weight | 3 lbs. |
| Emission Setting For 0.01 Amp/torr Sensitivity | 6.7 mA |
| Filament Current, Absolute Maximum | 6.0 Amps |
| Grid mfr | 180 VDC |
| Grid voltage mfr | 180 VDC |
| Filament voltage mfr | 4.0 VAC |
| Grid voltage (resistance heating) mfr | 6.3 VAC |
| Grid current (resistance heating) mfr | 8.7 Amps |
| Grid to filament voltage (electron bombardment) mfr | 700 VDC |
| Grid current (electron bombardment) mfr | 100 DC |
| Electron bombardment power, absolute maximum mfr | 50 W |
| Grid voltage (resistance heating degas) mfr | 6.3 VAC |
| Grid current (resistance heating degas) mfr | 8.7 Amps |
| Grid to filament voltage (electron bombardment degas) mfr | 700 VDC |
| Grid current (electron bombardment degas) mfr | 100 DC |
| Electron bombardment power mfr | 50 W |
| Filament Voltage mfr | 4.0 VAC |
| Electron bombardment power absolute maximum mfr | 50 W |
| Grid to filament voltage mfr | 700 VDC |
| Grid current mfr | 100 DC |
| Filament Temperature, Absolute Maximum | 1400 °C |
| Shield coating mfr | Platinum, internally connected to filament filam |
| Shield Coating mfr | Platinum, internally connected to filament filam |
| Standard pressure mfr | 101325 Pa |
Description
Construction and series positioning
- The 572 uses a Nonex glass envelope, a tungsten grid and collector, a helical tungsten filament (dual filaments, externally selectable), a platinum shield coating internally connected to the filament, and nickel base leads.
- Internal volume is 220 cc excluding tubulation; shipping weight is 3 lbs.
- Within the series, the 571 is the other Bayard-Alpert tube; the 531 and 536 are thermocouple (rough-vacuum) gauges and the 525 is a cold-cathode high-vacuum gauge. The 572 is the wide-range, linear hot-filament BA tube with dual tungsten filaments.
- Ordering variants differ only in tubulation and flange: 1 in or 0.75 in Nonex or Kovar tubulation, and NW25, NW40 or DN 35 CF-F (2.75 in CFF) flanges. Model 572-N is part number 0572K7360301 (1 in Nonex tubulation).
Behavior and measurement limits
- Useful range is 1×10⁻³ to 2×10⁻¹⁰ Torr; maximum operating pressure is 1×10⁻³ Torr. Below about 2×10⁻¹⁰ Torr (N₂) readings are not unequivocal because of the X-ray limit, which varies with collector surface condition.
- Readings are composition-sensitive and expressed as nitrogen-equivalent pressure. Gauge constants: helium 1.5 per Torr at 6.7 mA emission; nitrogen and air 10 per Torr at 1.0 mA emission.
- The gauge acts as a small ion pump: ionic pumping speed is 0.06 L/s for N₂ at 1 mA grid current, which matters in low-conductance systems.
- Tungsten filaments run hotter than thoria-coated iridium, giving a more stable reading; the second filament allows external switching without breaking vacuum.
Electrical operation and degassing
- Normal electrode potentials: grid +180 VDC, filament +30 VDC, collector at 0 VDC (ground) through the electrometer. Normal filament supply is 4.0 VAC at 3.5 A with 1 mA grid current; absolute maxima are 6.0 VAC and 6.0 A, and filament temperature must not exceed 1400 °C.
- Resistance degassing: 6.3 VAC grid voltage, 8.7 A grid current, 1200 °C grid temperature (absolute maxima). Typically takes one-half hour or more in the 10⁻⁹ Torr range; overnight degassing is usually needed before measurements below 10⁻⁸ Torr.
- Electron-bombardment degassing: +700 VDC grid-to-filament, 100 mA DC grid current, 1200 °C max grid temperature, 50 W absolute maximum. It cleans faster but requires careful watching because of the high voltage.
Installation and control
- The gauge may be mounted and operated vertically or horizontally, including during degassing, for prolonged periods.
- Mount as close as possible to the measurement location through a large tubulation (1 in OD recommended); low conductance makes the gauge read lower than system pressure, particularly below 10⁻⁸ Torr.
- The controller must provide the electrode potentials and filament supply above and must not limit the gauge range or exceed maximum ratings. Pin arrangement (572-N): A grid; D(1) & C filament 1; D(2) & E filament 2; H grid.
- Controller and vacuum system must be separately grounded to a common ground; a single interconnecting ground wire between chamber and controller chassis is not safe.
Selection & Integration Q&A
Pick the 572 for wide-range, linear high-vacuum measurement with dual tungsten filaments that can be switched externally without breaking vacuum; tungsten gives a more stable reading than thoria-coated iridium. Choose the 571 if that tube's configuration fits; choose the 525 cold-cathode gauge where a hot filament is unacceptable.
The controller must supply grid +180 VDC, filament +30 VDC, collector at 0 VDC through the electrometer, and 4.0 VAC / 3.5 A filament with 1 mA grid current, without limiting range or exceeding maxima. Controller and vacuum system must be separately grounded to a common ground capable of ≥10 A; a single interconnecting ground wire is not safe.
It may be mounted and operated vertically or horizontally, including during degassing, for prolonged periods. Mount close to the measurement point through a large tubulation (1 in OD recommended) to avoid low-conductance error below 10⁻⁸ Torr. Do not bake above 10⁻⁵ Torr, and do not energize with combustible gases present.
Plan degassing time: resistance degas typically one-half hour or more in the 10⁻⁹ Torr range, overnight before measurements below 10⁻⁸ Torr. Use powder-free butyl or polycarbonate gloves; Apiezon L grease (Agilent 695400004) if grease is needed. The controller, cable and flange hardware are not included with the tube and must be ordered separately.