Agilent Technologies Bond Elut HLB | SPE Sorbent
Agilent Technologies Bond Elut HLB | SPE Sorbent — Sample preparation with Pumping speed 1.8×10⁻⁵ m³/h.
Bond Elut HLB is a 6 mL, 500 mg polymeric SPE cartridge for extracting sulfonamide drugs from water. It is the HLB modification in the Bond Elut series, distinguished by its hydrophilic-lipophilic balance sorbent. The cartridge serves as the sample preparation front-end for LC/MS/MS quantification, achieving method detection limits of 0.28–1.88 ng/L.
Agilent Technologies Bond Elut Series — Model Comparison & Selection
| Model | Pumping speed, m³/h | Sample volume, milliliter | Drying gas temperature, °C | Average recovery all compounds, % | Cartridge sorbent mass, mg | Centrifugation speed, rpm |
|---|---|---|---|---|---|---|
| Bond Elut HLB | SPE Sorbent Current | 1.8×10⁻⁵ | 0.002 | 40 | 73.4...118 | 500 | — |
| Bond Elut Carbon S & Carbon S/NH2 | 0.024 | 0.005 | 230 | 99.2 | 150 | 1900 |
| Bond Elut Lipid Extraction | 3...5 | — | 30 | — | — | — |
| Bond Elut LMS | Large Molecule Size | — | 0.005 | — | — | — | 2800 |
Full specifications
| Pumping speed | 0.3 mL/min |
| Average Recovery (all Compounds) | 73.4...118 |
| Cartridge Volume | 6 mL |
| Gas Flow | 6 L/min |
| Reconstitution Volume | 1 mL |
| Rsd (all Compounds) | 14 % |
| Sample volume | 2 µL |
| Sheath Gas Flow | 11 L/min |
| Spe Cartridge Volume | 6 mL |
| Syringe Filter Pore Size | 0.2 µm |
| Syringe filter diameter mfr | 15 mm |
| LC flow rate mfr | 0.3 mL/min |
| Method detection limit range mfr | 0.28...1.88 ng/L |
| Method quantitation limit range mfr | 1.12...7.52 ng/L |
| Linearity (R2) mfr | 0.996...0.999 |
| Linearity (R2) mfr | 0.996...0.999 |
| Water sample volume mfr | 500 mL |
| EDTA final concentration mfr | 0.5 g/L |
| Sample loading flow rate mfr | 5 mL/min |
| Elution volume per aliquot mfr | 4 mL |
| Cartridge Sorbent Mass | 500 mg |
| Spe Cartridge Sorbent Mass | 500 mg |
| Capillary Voltage | 3500 V |
| Nozzle Voltage | 500 V |
| Drying Gas Temperature | 40 °C |
| Gas Temperature | 325 °C |
| Sheath gas heater mfr | 350 °C |
Description
Construction and Series Positioning
- Bond Elut HLB is a 6 mL SPE cartridge packed with 500 mg of hydrophilic-lipophilic balance sorbent (part number 5610-2147).
- Within the Bond Elut series, it is the HLB modification; siblings include C18, ENV, Carbon-Carbon-NH2, and Carbon-S-Carbon-S-NH2. The HLB sorbent is selected for balanced retention of polar and nonpolar analytes, whereas C18 is a classical reversed-phase silica and ENV is a styrene-divinylbenzene polymer.
- The cartridge format is a single-use, positive-pressure or vacuum-driven SPE column; no electrical or moving parts are integral to the cartridge itself.
Behavior Under Load and Method Performance
- The validated application targets 19 sulfonamide drugs in water. Method detection limits range from 0.28 to 1.88 ng/L and quantitation limits from 1.12 to 7.52 ng/L.
- Average recovery across the method ranges from 73.4% to 102%, with a recovery RSD of 14%. Linearity (R²) is 0.996–0.999.
- The LC/MS/MS detection step uses a 2 µL sample injection, 0.3 mL/min LC flow, capillary voltage 3500 V, nozzle voltage 500 V, sheath gas flow 11 L/min at 350 °C heater, MS gas flow 6 L/min at 325 °C, and eluent drying temperature 40 °C.
- A 0.2 µm syringe filter (15 mm diameter) is specified for sample filtration prior to injection.
Installation, Connection and Operation
- The cartridge is installed in a standard SPE vacuum manifold or positive-pressure manifold; connection is by Luer tip or compatible adapter. No electrical supply is required for the cartridge.
- The method uses the cartridge in a single-use mode: condition, load, wash, and elute steps are performed sequentially; the eluate is then dried and reconstituted for LC/MS/MS injection.
- The application note does not specify re-use; the cartridge is intended for one extraction cycle.
Integration & Operating Notes
The HLB polymeric sorbent operates across a wide pH range, but the application note specifies conditioning and elution solvents. Deviating from the validated solvent sequence can shift recovery outside the 73.4–102% range and compromise the 0.28–1.88 ng/L MDL.
The cartridge is a single-use consumable. Reusing a cartridge after elution risks carryover and variable recovery. The application note describes a single condition-load-wash-elute sequence per cartridge; no regeneration procedure is provided.
Capillary voltage 3500 V, nozzle voltage 500 V, sheath gas 11 L/min at 350 °C, MS gas 6 L/min at 325 °C, and LC flow 0.3 mL/min are detection-method settings from the application note. They must be optimized on the actual instrument and are not supplied by the cartridge.
Selection & Integration Q&A
Choose HLB for balanced retention of polar and nonpolar analytes such as sulfonamides; C18 is a silica-based reversed phase with limited polar retention, and ENV is a styrene-divinylbenzene polymer. HLB is the validated choice for the 19-sulfonamide method with MDL 0.28–1.88 ng/L.
The method specifies a 0.2 µm syringe filter (15 mm diameter) for sample filtration and a 2 µL injection volume. LC flow is 0.3 mL/min. These are method requirements; the cartridge itself does not include filters or syringes.