Agilent Technologies FTIR Spectroscopy Equipment
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies FTIR Spectroscopy Equipment

Agilent Technologies FTIR Spectroscopy Equipment — Molecular spectroscopy.

The Agilent 4200 FlexScan is a handheld FTIR spectrometer for onsite molecular analysis, covering 5100–600 cm⁻¹ with a single-reflection ATR interface. Its 2.6 µm effective pathlength and 2.5 mm ATR surface suit direct contact sampling of liquids and solids with minimal preparation. At 3.8 kg, it is positioned as the field-portable entry in the FlexScan/4300/4500/5500 series, distinguished by its handheld format rather than benchtop or compact configurations.

Agilent Technologies Molecular Spectroscopy Series — Model Comparison & Selection

Model Wavelength rangeSample volume, milliliterAbsorbance error, %Atr path length, µmAtr surface diameter, mmCertified absorbance value of filter
FTIR Spectroscopy Equipment Current 5100...6000.02—2.62.5—
Fluorescence Spectroscopy Instrumentation, Spectrofluorometers 900—————
LDIR Chemical Imaging, IR & Imaging Spectroscopy ——————
UV-Vis & UV-Vis-NIR Spectroscopy, Cary Instruments ——0.189——0.369

Full specifications

Core Performance & Functional Specifications
Wavelength range 5100...600
Atr Surface Diameter 2.5 mm
Effective Pathlength 2.6 μm
Optical Aperture 25 mm
Sample volume 20 µL
Bench space required mfr 20 cm
ATR spectral range mfr 5100...600
Spectrometer Body mfr Glass-fiber reinforced polyamide for durability
Detector mfr High-sensitivity thermoelectrically cooled deute
Interferometer mfr 45 aperture
Bench space mfr 20 cm
Spectral range (KBr optics) mfr 7000...350
Spectral range (ZnSe optics) mfr 5100...600
DialPath path lengths mfr 30...1,000 microns
Dimensions, Weight & Mechanical Design
Atr Path Length 2.6 μm
Weight 3.8 kg
Weight range 3.5 kg
ATR effective path length mfr 36.5
Footprint mfr 26.4 module
Materials, Media & Process Compatibility
ATR effective pH range mfr 1...14
ATR pH range mfr 5...9

Description

Construction and series positioning

Spectral behavior and sample compatibility

Installation and operation

Integration & Operating Notes

Process Limits
Keep sample pH within 5–9 at the ATR crystal

The ATR element is rated for pH 5–9. Samples outside this window can attack the crystal and degrade the effective pathlength; neutralize or dilute aggressive samples before contact, and clean the crystal immediately after each measurement.

Factory Specification Sheet, extra specs: ATR effective pH range
Process Limits
Minimum sample volume of 20 µL for reliable contact

The specified 20 µL sample volume reflects the amount needed to fully cover the 2.5 mm ATR surface. Underfilling leaves part of the evanescent field sampling air, which weakens absorbance and skews library matching.

Factory Specification Sheet, sample_volume
Maintenance
Clean the ATR crystal after every sample contact

Because sampling is direct-contact ATR with a fixed 2.6 µm pathlength, residue on the 2.5 mm crystal surface directly biases subsequent spectra. Wipe with a solvent compatible with the crystal and the pH 5–9 window before the next measurement.

Factory Specification Sheet, ATR surface diameter and effective pathlength

Selection & Integration Q&A

When should I choose the 4200 FlexScan over the 4300, 4500, or 5500 models in the same series?

Pick the 4200 when true handheld field use and direct-contact ATR sampling are the priority. The 4300 is also handheld; the 4500 and 5500 are portable/compact benchtop units, better suited when bench stability or additional sampling modules are needed. The supplied facts do not give performance differences beyond format.

What sample interface does the 4200 present to adjacent system parts, and what must be compatible with it?

The 4200 uses a built-in single-reflection ATR element with a 2.5 mm surface diameter and 2.6 µm effective pathlength, so no external flow cell or transmission accessory is required. Samples must be compatible with the pH 5–9 crystal rating and provide at least 20 µL for full surface coverage.

What environmental and duty-cycle limits apply when operating the 4200 FlexScan in the field?

The verified facts give weight (3.8 kg) and a 20 cm bench footprint but no operating temperature, humidity, or duty-cycle ratings. The only process limit stated is the ATR pH range of 5–9. Confirm environmental ratings with Agilent before committing to outdoor or high-humidity deployment.

What maintenance and consumables are required, and what is not included in base delivery?

Routine maintenance is cleaning the ATR crystal after each contact, keeping samples within pH 5–9. The supplied facts list no cables, calibration standards, or replacement ATR crystals as included; the attached digests cover Cary 635 benchtop accessories, not 4200 FlexScan consumables, so confirm the delivery scope with Agilent.