Agilent Technologies Reliable, consistent GC introduction of liquefied gases
Agilent Technologies Reliable, consistent GC introduction of liquefied gases — Gas chromatography.
The Agilent 990 Micro GC Micro Gasifier is a heated vaporizing inlet accessory that enables reliable, consistent GC introduction of liquefied gas samples onto the 990 Micro GC platform. It accepts sample pressures up to 100 kPa at the Micro GC inlet and delivers a factory-set 10 Psi outlet pressure at a 100 °C vaporizing regulator temperature. This modification is the liquefied-gas introduction path within the GC sample preparation series, distinct from autosampler and headspace siblings.
Agilent Technologies GC Sample Preparation Introduction Series — Model Comparison & Selection
| Model | Sample volume, milliliter | Analysis time, min | Bake flow, mL/min | Bake temperature, °C | Bake time, min | Calibration concentration, µg/ml |
|---|---|---|---|---|---|---|
| Reliable, consistent GC introduction of liquefied gases Current | — | — | — | — | — | — |
| GC autosampler, GC sample injection | 0.001 | 80 | — | — | — | 100 |
| PAL3 autosampler, automated solid phase extraction | — | — | — | — | — | — |
| purge and trap, volatile organic compounds analysis | — | — | 200 | 280 | 2 | — |
Full specifications
| Inlet Pressure | 1000 Psi |
| Maximum sample pressure at Micro GC inlet mfr | 100 kPa |
| Outlet pressure (factory set) mfr | 10 Psi |
| Required Micro GC firmware version mfr | 1.04 greater |
| Supported data system mfr | 2.1 |
| Required tool mfr | -20 pcs |
| Mounting screws mfr | 4 pcs |
| Manual part number mfr | 3535 |
| Filter kit part number mfr | 3535 |
| Gasifier mounting bracket part number mfr | 3535 |
| USB cable part number mfr | 8121 |
| Core ferrite part number mfr | 9170 |
Description
Construction and series positioning
- The Micro Gasifier is a dedicated accessory for the 990 Micro GC, mounted via a gasifier mounting bracket (part number 3535) with four mounting screws.
- Within the GC sample preparation introduction series, this unit is the liquefied-gas vaporizing inlet; sibling models cover autosampler injection and headspace sampling, which handle liquid and headspace matrices rather than liquefied gases.
- The accessory is supplied with a filter kit (part number 3535) and requires a USB cable (part number 8121) plus a core ferrite (part number 9170) for connection.
Behavior and process envelope
- Maximum sample pressure at the Micro GC inlet is 100 kPa; the inlet-pressure spec of 1000 Psi is not used here because it conflicts with the kPa limit and is omitted.
- The vaporizing regulator is factory set to 100 °C and the outlet pressure to 10 Psi; these fixed settings define the introduction condition for liquefied gas samples.
- No digitized curve points are available, so load behavior across the pressure range cannot be characterized from the supplied facts.
Installation, control and operation
- Installation requires Micro GC firmware version greater than 1.04 and data system version 2.1.
- The manual (G3535-90000) covers installation, operation, maintenance and troubleshooting for the Gasifier on 8860 GC, 8890 GC and 990 Micro GC systems; the safety manual is G3535-90008.
- A required tool is listed with value -20, which is not a usable tool designation and is therefore not specified here.
Integration & Operating Notes
Exceeding the 100 kPa maximum sample pressure at the Micro GC inlet risks overpressurizing the gasifier inlet path and invalidating the introduction. Verify the sample delivery pressure before connection; the separate 1000 Psi inlet-pressure entry is not used because it conflicts with this kPa limit.
The gasifier mounting bracket (part number 3535) and four mounting screws are the specified mounting hardware. Substituting other brackets can misalign the heated inlet path and the USB routing to the Micro GC.
Connection to the Micro GC uses USB cable part number 8121 with core ferrite part number 9170. The ferrite suppresses interference on the USB line; omitting it can degrade communication with the gasifier during runs.
The filter kit (part number 3535) is listed as a separate service item. Plan filter replacement according to sample cleanliness; a loaded filter alters the inlet pressure seen by the gasifier and can cause inconsistent liquefied-gas introduction.
Selection & Integration Q&A
Choose the Micro Gasifier when the sample is a liquefied gas that must be vaporized and introduced at a controlled 100 °C regulator temperature and 10 Psi outlet. Autosampler and headspace siblings handle liquid injection and headspace matrices, not liquefied-gas introduction.
Sample pressure at the Micro GC inlet must not exceed 100 kPa. The Micro GC must have firmware greater than 1.04 and data system 2.1. The gasifier connects via USB cable 8121 with core ferrite 9170 and mounts with bracket 3535 and four screws.
The gasifier is installed only in combination with the 990 Micro GC using the dedicated bracket and four screws. The vaporizing regulator is factory set to 100 °C, so the surrounding installation must tolerate that heated zone; firmware greater than 1.04 is required before operation.
The filter kit (3535), gasifier mounting bracket (3535), four mounting screws, USB cable (8121) and core ferrite (9170) are listed separately from the base unit. Filter replacement is the routine service action; the manual G3535-90000 covers maintenance and troubleshooting.
Dimensional Drawings & Technical Documentation
Dimensional drawings
Click drawing to inspect vector dimensions