Agilent Technologies GC autosampler, GC sample injection — Gas chromatography.
The 7693A is a tower-mounted automatic liquid sampler for Agilent 8890/8860 GC systems, delivering 1 µl injections from a 5 µl syringe through a split/splitless inlet at 250 °C. It is the standard ALS in the 7693 series, distinguished from the 7650A by its injector tower configuration and from the 7697A/8697 headspace samplers by handling only liquid samples.
Agilent Technologies GC Sample Preparation Introduction Series — Model Comparison & Selection
| Model | Sample volume, milliliter | Analysis time, min | Bake flow, mL/min | Bake temperature, °C | Bake time, min | Calibration concentration, µg/ml |
|---|---|---|---|---|---|---|
| GC autosampler, GC sample injection Current | 0.001 | 80 | — | — | — | 100 |
| PAL3 autosampler, automated solid phase extraction | — | — | — | — | — | — |
| purge and trap, volatile organic compounds analysis | — | — | 200 | 280 | 2 | — |
| Reliable, consistent GC introduction of liquefied gases | — | — | — | — | — | — |
Full specifications
| Analysis time | 80 min |
| Calibration Concentration | 100 µg/ml |
| Oven Temperature Ramp | 5 °C/min |
| Retention Time, 1-butanol | 11.43 min |
| Retention Time, 2-butanol | 7.88 min |
| Retention Time, Ethylacetat | 4.51 min |
| Retention Time, Ethylbutyrat | 8.09 min |
| Retention Time, Ethylhexanoat | 13.88 min |
| Retention Time, Ethyllactat | 16.93 min |
| Retention Time, Isoamylalkohol | 13.25 min |
| Retention Time, Isobutanol | 9.96 min |
| Retention Time, Methanol | 4.75 min |
| Retention Time, N-propanol | 8.26 min |
| Sample volume | 1 µl |
| Split Flow | 30 ml/min |
| Syringe Volume | 5 µl |
| GC system mfr | 8890 GC |
| Detector mfr | FID FID |
| Autosampler mfr | 7693 ALS |
| Split ratio mfr | 20 |
| Liner mfr | Split, Ultra inert, glass wool, low pressure pre |
| Carrier gas mode mfr | constant pressure pressure |
| Sample solvent mfr | hexane hexane |
| Sample concentration mfr | 50...100 ng/μl |
| FAME mix concentration range mfr | 200...400 ng/μl |
| RF RSD mfr | 4 |
| Maximum component deviation mfr | 4.0 |
| Syringe size mfr | 10 μl |
| Inlet type mfr | Split/Splitless Split/Splitless |
| Inlet mode mfr | Split Split |
| Gas saver time mfr | Aus Aus |
| High vacuum pump mfr | Turbo Turbo |
| Acquisition mode mfr | Scan Scan |
| Scan range mfr | 40 m/z |
| Tuning algorithm mfr | Etune Etune |
| GC/MS flow requirement (EPC capability) mfr | 2 ml/min |
| Component weight percent range mfr | 3...30 |
| Inlet Temperature | 250 °C |
| Quadrupole Temperature | 175 °C |
| Source Temperature | 250 °C |
| Standard Stock Solution Concentration | 1000 µg/ml |
| Number of compounds in standard mfr | 28 |
Description
Construction and series positioning
- The 7693A is a liquid-only autosampler built around a syringe tower that mounts directly on the GC inlet; it is used with the 8890 and 8860 GC platforms and the 5977 GC/MSD in the attached application notes.
- Within the sampler family it sits between the 7650A (a different ALS variant) and the 7697A/8697 headspace samplers, which handle volatile analytes from solid or liquid matrices rather than direct liquid injection. The 7693A is the correct choice when the sample is a liquid dissolved in a solvent such as hexane and is introduced through a split/splitless inlet.
- Documented configurations use a 5 µl syringe (application note) or a 10 µl syringe (FAME method), so syringe volume is a method-level selection, not a fixed property of the sampler.
Behavior under method conditions
- In the distilled-spirits application the sampler injects 1 µl at a split flow of 30 ml/min (split ratio 20) into a 250 °C S/SL inlet with a low-pressure split liner; the oven ramps at 5 °C/min and the run takes 80 min, resolving methanol at 4.75 min through ethyl lactate at 16.93 min.
- In the FAME application the same sampler type supports the GB 5009.168-2016 and GB 5009.257-2016 methods on an 8890 GC with FID, covering 37-component and 13-component FAME mixes and cis/trans C18:3 isomers; component weight-percent range 3–30 with RF RSD 4 and maximum component deviation 4.0.
- In the forensic application the sampler feeds a 5977 MSD in scan mode (m/z 40 and up) with Etune; the 28-compound standard is run at 5 µg/ml. These are method parameters, not sampler ratings.
Installation, control and operation
- The sampler is controlled as part of the GC data system; the digests give no separate sampler control interface, so integration is through the GC method (inlet temperature, split flow, carrier-gas mode such as constant pressure).
- Carrier-gas mode in the documented methods is constant pressure; gas saver is switched off, so the inlet purge flow continues for the full 80-minute run and the gas supply must be sized for that duty.
- The sampler is mounted on the inlet; the digests do not specify clearance, supply voltage or communication cabling, so those must be taken from the instrument installation manual rather than inferred here.
Integration & Operating Notes
The 7693A injects through the S/SL inlet, so the tower position and syringe guide are set for that inlet geometry; fitting it to a different inlet type or a different GC platform requires the matching mounting hardware and re-alignment before the first injection.
Documented methods use 1 µl injections from a 5 µl syringe or a 10 µl syringe; the injected volume, syringe size and inlet split ratio must be chosen together, because a volume too large for the syringe or the liner will cause carryover or inlet flooding rather than a valid result.
The documented methods use constant-pressure carrier-gas mode and leave gas saver disabled, so the inlet consumes carrier gas for the entire 80-minute run; the gas supply and purity chain must be sized for continuous flow over the full sequence, not just the injection window.
The 5 µl or 10 µl syringe and the split, ultra-inert, glass-wool low-pressure liner are method-specific consumables; they wear with each injection and must be replaced on the laboratory's preventive schedule, and neither is defined as part of the base sampler delivery in the digests.
Injection volume, split flow and inlet temperature are set in the GC method that also controls the 8890/8860 and the 5977 MSD; the sampler has no separate method editor described in the digests, so method validation covers the sampler and GC together.
Selection & Integration Q&A
Pick the 7693A for direct liquid injection through a split/splitless inlet, as in the distilled-spirits, FAME and forensic methods. Choose the 7697A or 8697 when analytes must be volatilized from the matrix by headspace; the 7650A is the alternative ALS variant in the same family.
A split/splitless inlet with a split, ultra-inert, glass-wool low-pressure liner, run at 250 °C in constant-pressure mode with gas saver off; the documented split flow is 30 ml/min at split ratio 20, so the carrier-gas supply must sustain that flow for the full 80-minute run.
The syringe (5 µl in the spirits method, 10 µl in the FAME method) and the split, ultra-inert, glass-wool low-pressure liner are method-specific items; the digests do not list them as part of base delivery, so they must be ordered and replaced on the laboratory's preventive-maintenance schedule.