Agilent Technologies IDP-4
Agilent Technologies IDP-4 — Vacuum technologies with Motor power 0.119 kW, Pumping speed 3.6 m³/h, Ultimate pressure 2.4 mbar.
The IDP-4 is an air-cooled dry scroll vacuum pump driven by a 0.16 HP (0.12 kW) single-phase motor, available for 100–127 VAC or 200–240 VAC 50/60 Hz service. It delivers 65 L/min peak pumping speed and an ultimate pressure of 1.8 × 10⁻¹ Torr at the default 3300 rpm, with speed adjustable from 2200 to 3300 rpm via serial interface. It serves as a clean, oil-free backing or roughing stage for analytical instruments and laboratory vacuum systems.
Agilent Technologies Idp Pumps Series — Model Comparison & Selection
| Model | Motor power, kW | Ultimate pressure, mbar | Pumping speed, m³/h | Base pressure, mbar | Supply frequency, Hz | Operating temperature, °C |
|---|---|---|---|---|---|---|
| IDP-4 Current | 0.119 | 2.4 | 3.6 | 3.3 | — | 5 |
| 20.60 CFM | 0.119 | 3.3 | 3.6 | 3.3 | 50 | — |
| 26.48 CFM | 0.119 | 3.3 | 3.6 | 3.3 | 50 | — |
| 6.0 CFM | — | — | — | — | — | — |
| Scroll Pumps for Helium Recirculation & Cryogenic Applications | 0.119 | — | 3.6 | 3.3 | — | — |
Full specifications
| Motor power | 0.16 HP |
| Pumping speed | 60 L/min |
| Ultimate pressure | 1.8 Torr |
| Base Pressure | 3.3 mbar |
| Coverslip type | No corrosive, explosive or particulate forming gases |
| Operating Speed | 3300 % rpm |
| Leak rate | 1 helium |
| Peak pumping speed mfr | 65 L/min |
| Maximum recommended continuous inlet pressure mfr | 1.0 atmosphere |
| Maximum outlet pressure mfr | 0.45 bar |
| Gas ballast mfr | 1 Thread |
| Motor protection mfr | Thermal protection/over current protection/fuse |
| Installation mfr | 1 |
| Emissions and Immunity mfr | 2014 |
| Safety mfr | 2006 |
| IDP-10 mfr | 3810 |
| IDP-10 with inlet isolation valve mfr | 3810 |
| IDP-7/10 tip seal replacement kit mfr | 3807 |
| Exhaust silencer kit mfr | 110 |
| Purge kit mfr | 3807 |
| Gas ballast kit mfr | 3807 |
| Inlet trap mfr | 25 |
| Vibration isolation kit mfr | 110 VIBISOKIT |
| IDP-35 mfr | 3835 |
| IDP-35 with Gauge mfr | 3835 |
| IDP-35 with Inlet Valve mfr | 3835 |
| IDP-35 with Inlet Valve and Gauge mfr | 3835 |
| Inlet HEPA filter, NW40 mfr | 40 |
| HEPA filter, NW25 mfr | 25 |
| IDP-45 mfr | 3845 |
| IDP-45 with Gauge mfr | 3845 |
| IDP-45 with Inlet Valve mfr | 3845 |
| IDP-45 with Inlet Valve and Gauge mfr | 3845 |
| IDP-15 mfr | 3815 |
| IDP-15 with inlet isolation valve mfr | 3815 |
| IDP-15 tip seal replacement kit mfr | 3815 |
| IDP-3 mfr | Vibration isolation kit kit |
| IDP-4 mfr | Replacement filter element element |
| IDP-7 mfr | Vibration isolation kit kit |
| inlet valve closing time mfr | 20 milliseconds |
| inlet valve opening time after power restore mfr | 10 seconds |
| tip seal replacement duration mfr | half an hour hour |
| IDP-4 maintenance duration mfr | 15...30 minutes |
| integrated Pirani gauge pressure range mfr | -4...1000 mbar |
| VPI valve isolation time mfr | 20 ms |
| Maximum outlet pressure*** mfr | 0.45 bar |
| Description of the IDP-4 Dry Scroll Pump mfr | 188 |
| Technical Specifications mfr | 189 |
| IDP-4 Dry Scroll Pump Outline Drawing mfr | 191 |
| Performance Curves mfr | 193 |
| Electrical Connections mfr | 195 |
| Grounding Instructions mfr | 195 |
| Mechanical Connections mfr | 197 |
| Gas Ballast mfr | 199 |
| Accessories and Spare Parts mfr | 201 |
| Optional Accessories mfr | 201 |
| Optional Inlet Valve Kit mfr | 202 |
| Cleaning and Preventive Maintenance mfr | 205 |
| External cleaning mfr | 205 |
| Internal cleaning as part of a Preventive Maintenance plan mfr | 206 |
| Verifying Performance mfr | 206 |
| Tip Seal Replacement mfr | 207 |
| Tip Seal Replacement Instructions mfr | 208 |
| User Maintenance and Repair mfr | 216 |
| Factory-Only Service mfr | 217 |
| Service Options mfr | 218 |
| Troubleshooting mfr | 219 |
| Window Description mfr | 223 |
| Europe mfr | 10 m |
| Denmark mfr | 10 m |
| Switzerland mfr | 10 m |
| UK/Ireland mfr | 13 m |
| India mfr | 10 m |
| Israel mfr | 10 m |
| Japan mfr | 12 m |
| North America mfr | 13 m |
| Body mfr | Aluminum Aluminum |
| Seal mfr | Viton Viton |
| Closing time (after solenoid valve power loss) mfr | 25 ms |
| Closing burst mfr | 0.01 liter |
| IDP3GBKIT mfr | Breather Vent and Adapter Adapter |
| VPI16IDP24DC mfr | -24 |
| EXSLRIDP3 mfr | Exhaust Silencer Kit Kit |
| X3802-68001 mfr | Serial communication cable kit kit |
| Clearance on all sides mfr | 20 cm |
| Lifting device threshold mfr | 20 kg |
| Gas ballast duration mfr | 30 hours |
| Weight | 7.7 kg |
| Operating voltage mfr | 100 VAC |
| Maximum inlet pressure at nominal voltage mfr | 1.0 atmosphere |
| Operating voltages mfr | 200 VAC |
| Fuse type mfr | 5 mm |
| IDP-35/IDP-45 supply frequency mfr | 50 Hz |
| Inlet connection | 16 |
| Interface dimensions mfr | 237 mm |
| Serial Connector & Pin allocation mfr | 221 |
| Pump Serial Connector mfr | 221 |
| Serial communication settings mfr | 222 |
| Window Description and Protocol mfr | 223 |
| Windows Protocol mfr | 226 |
| Cooling System | Air-cooled |
| Operating temperature | 5 °C |
| Storage temperature | 20 °C |
| Ambient Relative Humidity mfr | 5 condensing |
| Sound Pressure Levels mfr | 50...55 dba |
| Altitude mfr | 2000 m |
| Relative humidity mfr | 5...95 |
| Ultimate Pressure* (ISO 21360-1: 2020) (ISO 21360-2: 2020) mfr | 1.8 rpm |
| Motor rating: EN 60034-1-2:2007 EN 60034-30 (GB 18613-2020) mfr | 0.16 HP |
| Sound Pressure Levels (EN ISO 2151:2008) (ISO 4871:1996) (ISO 11201:2010) mfr | 50...55 configuration |
Description
Construction and series positioning
- The IDP-4 is a dry scroll pump in the Agilent IDP family (IDP-3, IDP-4, IDP-7, IDP-10, IDP-15, IDP-35, IDP-45). Its mounting holes are backward compatible with the IDP-3, allowing drop-in replacement of the smaller sibling without reworking the baseplate.
- Unlike the 24 VDC-driven IDP-3, the IDP-4 uses a 0.16 HP single-phase AC motor with a 5 A slow-blow 250 V 5×20 mm fuse and thermal/overcurrent protection. It is air-cooled and weighs 7.7 kg.
- The standard pump ships with the gas ballast port sealed; the IDP3GBKIT is required for water-vapor or solvent duty. An optional VPI inlet valve is available and opens approximately 10 seconds after power-on.
Performance behavior
- Peak pumping speed is 65 L/min (4 m³/h, 2.35 cfm) at full speed, averaged for N₂ and He without an inlet valve. Ultimate pressure is 1.8 ± 0.2 × 10⁻¹ Torr (2.4 ± 0.27 × 10⁻¹ mbar) at the default 3300 rpm; published ultimate pressure assumes dry conditions and default speed.
- Speed is variable from 2200 to 3300 rpm (default 3300 rpm ±10%) and can be set via serial window WIN#127. Maximum recommended continuous inlet pressure is 1.0 atmosphere; maximum outlet pressure is 0.45 bar (6.5 psig), above which the pump automatically shuts off to protect the head.
- With the gas ballast kit fitted, water handling capability is 4 g/hr under typical conditions and up to 10 g/hr in higher pressure ranges. A dry nitrogen purge at approximately 5 L/min can be applied at the gas ballast port.
Installation, control and operation
- Inlet and exhaust connections are NW16 (16 mm); use clean NW16 or larger hardware, short lines, and bellows for vibration isolation. Exhaust flow must be unrestricted to avoid backpressure trips.
- Electrical service requires a power cord rated at least 15 A for 100–127 V and at least 10 A for 200–240 V; extension cords are not recommended (if used, minimum 16-gage, maximum 7.6 m, 3-wire). Mains fluctuation must stay within ±10% of nominal.
- Serial communication defaults to 9600 bit/s (WIN#108=4; options 4800/9600/19200/38400). WIN#205 reports status (0=Stop, 2=Start, 5=Normal, 6=Fail); WIN#206 is an error bitfield covering locked rotor, bus overvoltage, Hall error, motor overtemperature, heatsink overtemperature and system fault. Remote I/O or serial cables must not exceed 30 m.
- Install with at least 20 cm free space on all sides for ventilation, maintenance access and power-switch access. Ambient operating range is 5–40 °C, 5–95% RH non-condensing, altitude up to 2000 m, indoor pollution degree 2. Pump surface temperature above 65 °C is potentially damaging; stop and cool if observed.
Integration & Operating Notes
The IDP-4 is offered in 100–127 VAC and 200–240 VAC single-phase variants. The power cord must be rated at least 15 A for 100–127 V service and at least 10 A for 200–240 V service; undersized cords cause voltage drop and overheating. Mains fluctuation must remain within ±10% of nominal.
Maximum outlet pressure is 0.45 bar (6.5 psig). The pump automatically shuts off if discharge pressure exceeds this limit or if head temperature rises due to excessive exhaust back pressure. Size exhaust lines, mufflers and silencers so they do not restrict flow.
The standard pump ships with the gas ballast port sealed. For water vapor or solvent applications, install the IDP3GBKIT; water handling is 4 g/hr typical and up to 10 g/hr in higher pressure ranges. Do not exceed 25 in-lb torque when installing the gas ballast adapter into the manifold.
Configure the host serial port to match WIN#108 (default 4 = 9600 bit/s; options 4800/9600/19200/38400). Remote I/O or serial cables must not exceed 30 m; longer runs compromise EMC and communication reliability. Use shielded, properly grounded cable.
The motor protection circuit uses a 5 A slow-blow 250 V 5×20 mm fuse. Replace only with an identical type and rating to maintain protection. Persistent fuse blowing or overcurrent stops should be diagnosed via error code WIN#206 before replacement.
Pump surface temperature in excess of 65 °C (150 °F) is potentially damaging. Turn off and allow to cool if observed. Ensure cooling vents are not blocked and ambient temperature stays within 5–40 °C at 5–95% RH non-condensing.
Selection & Integration Q&A
Choose the IDP-4 when you need 65 L/min peak speed and 1.8 × 10⁻¹ Torr ultimate pressure from a 0.16 HP AC-powered pump with IDP-3-compatible mounting. The IDP-3 is 24 VDC and smaller; the IDP-7/IDP-10 offer higher speed for larger chambers. The IDP-4 fits analytical backing duty where AC mains are available.
Use clean NW16 or larger inlet and exhaust hardware with short lines and bellows for vibration isolation; exhaust must be unrestricted to stay below 0.45 bar (6.5 psig). Supply single-phase 100–127 VAC or 200–240 VAC 50/60 Hz within ±10%, with a cord rated 15 A (100–127 V) or 10 A (200–240 V).
Allow at least 20 cm free space on all sides for ventilation and service access; operate at 5–40 °C, 5–95% RH non-condensing, up to 2000 m altitude. If discharge pressure exceeds 0.45 bar (6.5 psig) or head temperature rises from restricted exhaust, the pump automatically shuts off to protect itself.
The base pump ships with the gas ballast port sealed; the IDP3GBKIT, VPI inlet valve, exhaust silencer and serial communication cable kit are optional accessories. The 5 A slow-blow 250 V 5×20 mm fuse is a service item. Faults are read via serial WIN#205 status and WIN#206 error bitfield (locked rotor, bus overvoltage, Hall error, motor/heatsink overtemperature, system fault).