Describes hermetic design, quiet operation, reliability, oil-free construction
Overviewpp. 6-7
Presents the new compact IDP-4 and smart features
Configp. 8
Guidance on selecting the appropriate IDP dry scroll pump
Orderingpp. 9-10
Lists ordering codes, part numbers, and available configurations
Servicesp. 11
Highlights partnership, service, and support offerings
p. 10table, 33 rows
p. 9table, 38 rows
p. 8table, 7 rows
Ordering information — Agilent IDP Dry Scroll Pumps
Model / Description
Part number
IDP-10
X3810-64000
IDP-10 with inlet isolation valve
X3810-64010
Service part numbers
IDP-7/10 tip seal replacement kit
X3807-67000
Accesory part numbers
Exhaust silencer kit
X3807-68003
Purge kit
X3807-68004
Gas ballast kit
X3807-68008
Inlet trap
SCRINTRPNW25
Vibration isolation kit
SH110VIBISOKIT
IDP-35
X3835-64011
IDP-35 with Gauge
X3835-64010
IDP-35 with Inlet Valve
X3835-64001
IDP-35 with Inlet Valve and Gauge
X3835-64000
Accesory part numbers
Inlet HEPA filter, NW40
SCRINTRPNW40
HEPA filter, NW25
SCRINTRPNW25
IDP-45
X3845-64011
IDP-45 with Gauge
X3845-64010
IDP-45 with Inlet Valve
X3845-64001
IDP-45 with Inlet Valve and Gauge
X3845-64000
Accesory part numbers
Inlet HEPA filter, NW40
SCRINTRPNW40
HEPA filter, NW25
SCRINTRPNW25
IDP-15
X3815-64000
IDP-15 with inlet isolation valve
X3815-64010
Service part numbers
IDP-15 tip seal replacement kit
X3815-67000
Accesory part numbers
Exhaust silencer kit
EXSLRSH110
Purge kit
X3807-68004
Gas ballast kit
X3807-68008
Inlet trap
SCRINTRPNW25
Таблица на странице напечатана как несколько последовательных блоков (IDP-10, IDP-35, IDP-45, IDP-15) с одинаковой структурой «Description / Part number»; здесь они объединены в одну таблицу с сохранением порядка строк.
Ordering information — Agilent IDP Dry Scroll Pumps
Model
Description
Part number
IDP-3
IDP-3, 1φ, 220 V, 50/60 Hz
IDP3A01
IDP-3
IDP-3, 1φ, 115 V, 60 Hz
IDP3B01
IDP-3
IDP-3, 1φ, 100 V, 50/60 Hz
IDP3C01
IDP-3
IDP-3, 24 VDC
IDP3D01
IDP-3
With inlet isolation valve
IDP-3
IDP-3, 1φ, 220 V, 50/60 Hz
IDP3A21
IDP-3
IDP-3, 1φ, 115 V, 60 Hz
IDP3B21
IDP-3
IDP-3, 1φ, 100 V, 50/60 Hz
IDP3C21
IDP-3
IDP-3, 24 VDC
IDP3D21
IDP-3
Service part numbers
IDP-3
Tip seal replacement kit
IDP3TS
IDP-3
Accessory part numbers
IDP-3
Exhaust silencer kit
EXSLRIDP3
IDP-3
Inlet trap
SCRINTRPNW16
IDP-3
Breather vent and adapter kit
IDP3GBKIT
IDP-3
Replacement filter element
REPLSLRFILTER2
IDP-3
Vibration isolation kit
IDP3VIBISOKIT
IDP-4
IDP-4 dry scroll pump
X3802-64000
IDP-4
IDP-4 dry scroll pump, with VPI
X3802-64010
IDP-4
IDP-4 dry scroll pump, for helium use
X3802-64400
IDP-4
Service part numbers
IDP-4
Tip seal replacement kit
IDP3TS
IDP-4
Accesory part numbers
IDP-4
Exhaust silencer kit
EXSLRIDP3
IDP-4
Inlet valve kit
VPI16IDP24DC
IDP-4
Breather vent and adapter kit
IDP3GBKIT
IDP-4
Vibration isolation kit
IDP3VIBISOKIT
IDP-4
Replacement filter element
REPLSLRFILTER2
IDP-7
IDP-7
X3807-64000
IDP-7
IDP-7 with inlet isolation valve
X3807-64010
IDP-7
Service part numbers
IDP-7
IDP-7/10 tip seal replacement kit
X3807-67000
IDP-7
Accesory part numbers
IDP-7
Exhaust silencer kit
X3807-68003
IDP-7
Purge kit
X3807-68004
IDP-7
Gas ballast kit
X3807-68008
IDP-7
Inlet trap
SCRINTRPNW25
IDP-7
Vibration isolation kit
SH110VIBISOKIT
Модель (IDP-3, IDP-4, IDP-7) добавлена как отдельная колонка для группировки; в исходной таблице она указана заголовком раздела. Написание «Accesory» сохранено как в тексте.
How to select your Agilent IDP dry scroll vacuum pump — Agilent IDP Dry Scroll Pumps
60 Hz/24 VDC
3.3 x 10-¹ mbar
2.5 x 10-¹ Torr
50 Hz
3.3 x 10-¹ mbar
2.5 x 10-¹ Torr
0.16 HP (120 W)/
100-120; 200-240 VAC
- Leak detection
- GC/MS
- Helium recirculation
- Small and light primary pump
- 24 V drive
IDP-4
55 Hz
65 L/min
4 m3/h
at full speed setting
55 Hz
2.4 x 10-¹ mbar
1.8 x 10-¹ Torr
at full speed setting
0.16 HP (120 W)/100-127; 200-240 VAC
- Leak detection
- GC/MS
- Helium recirculation
- Lightest, most compact primary pump
- Compact size for integrating the pump into your instrument
- Universal pumping performance at all global frequencies and input voltages
- Worldwide voltages compatibility
IDP-7
60 Hz
152 L/min
9.1 m³/h
50 Hz
120 L/min
7.2 m³/h
60 Hz
2.6 x 10-² mbar 2.0 x 10-² Torr
50 Hz
4.0 x 10-² mbar 3.0 x 10-² Torr
0.38 HP (300 W)/
100-120; 200-240 VAC
- Vacuum ovens
- Helium recirculation
- Backing high vacuum pump
- General laboratory vacuum
- More pumping speed in a compact, tabletop pump
- Handles water vapor
IDP-10
50 Hz or 60 Hz at full rotational speed (factory setting)
170 L/min
10.2 m³/h
50 Hz or 60 Hz at full rotational speed (factory setting)
2.0 x 10-² mbar
1.5 x 10-² Torr
0.5 HP (350 W)/
100-127; 200-240 VAC
- Microscopy instruments
- Surface analysis instruments
- Mass Spectrometry
- Helium recirculation
- Thin film deposition
- Universal pumping performance at all input frequencies
- Remote speed control or on/off capability
60 Hz
1.3 x 10-² mbar
1.0 x 10-² Torr
50 Hz
1.3 x 10-² mbar
1.0 x 10-² Torr
0.75 HP (560 W)/
100-115; 220-230 VAC
- General laboratory vacuum
- Thin film deposition
- LED, bulb manufacturing
With inlet isolation valve
- GC/MS
- Beam and particle physics
- Backing a turbo pump
- Quietest pump available
- Lowest vibration at inlet available
- Good speed to evacuate a large chamber
- Safegaurds your system and high vacuum pump in case of power loss
- Isolates the pump inlet during turbulent stop and start
- Hermetically sealed (magnetic torque coupling)
- Precise vacuum control
- Smart features- visibility and control
- Communications/Remote access and control (analog and digital)
- Inverter for universal voltage, flexible and easy
IDP-45
60 Hz
750 L/min
45 m³/h
60 Hz
1.07 × 10-² mbar
8 × 10-³ Torr
1.5 HP (1.1 KW)/
100-127; 200-240 VAC
Значения приведены дословно, как напечатано на странице; для IDP-45 колонки Common Applications и Advantages на странице не заполнены.
Engineering Facts
14 facts · every claim linked to a page
Maintenance
Tip seal or pumping module replacement is a DIY task needed only every two to three years and takes about half an hour, so plan for minimal downtime and no service contract.
Determines maintenance schedule, spare parts stocking, and total cost of ownership when selecting a pump.
Source
“This can be performed approximately every two to three years and requires about half an hour.”
Installation
The pump uses standard IEC power cords that are not hard-wired, simplifying electrical integration and replacement.
Affects cabinet wiring, connector selection, and serviceability during installation.
Source
“Agilent IDP dry scroll pumps also use standard IEC power cords that are not hard-wired to the motor.”
Safety
The optional integral isolation valve closes the inlet in about 20 ms on power loss and reopens in 10–30 s after power restore, protecting the vacuum system from backstreaming and sudden venting.
Critical for backing turbo pumps and contamination-sensitive systems; dictates whether the valve option must be ordered.
Source
“Upon power loss, the solenoid valve vents a small chamber beneath the valve, which closes the inlet valve (approximately 20 milliseconds).”
Installation
Integral vibration isolation feet or wheels separate motor and scroll vibration from the mounting surface, reducing transmitted vibration.
Matters for sensitive instruments (microscopy, surface analysis) and for floor/cabinet mounting decisions.
Source
“Integral vibration isolation feet or wheels, depending on the model, reduce vibrations by separating motor and scroll vibrations from the mounting brackets and support surface.”
Selection
The IDP-4 is the smallest and lightest pump in the family and is recommended for helium applications with proven long-term He tightness.
Guides selection for space-constrained OEM instruments and helium recirculation where footprint and He loss matter.
Source
“The Agilent IDP-4 dry scroll pump delivers clean, oil-free vacuum in the most compact footprint of the IDP family”
Operation
The IDP-4 offers adjustable pumping speed via inverter drive, allowing speed reduction once base pressure is reached to lower noise, energy use, and wear.
Enables optimization of power consumption and noise in continuous lab operation.
Source
“Adjustable pumping speed to match application requirements”
Operation
IDP-35 and IDP-45 have a tool-free adjustable gas ballast knob with Hi, Low, or No Flow settings to handle vapor loads.
Necessary when pumping condensable vapors (ovens, wet processes) to avoid condensation inside the pump.
Source
“The IDP-35 and IDP-45 models come standard with a simple (no tools), user-adjustable gas ballast knob.”
Installation
IDP-35 and IDP-45 use a Variable Frequency Drive giving consistent performance from 105 VAC to 240 VAC at 50 or 60 Hz.
Simplifies global deployment and OEM qualification without voltage-specific variants.
Source
“Get consistent performance worldwide from 105 VAC to 240 VAC (50 Hz or 60 Hz).”
Operation
IDP-35 and IDP-45 can be fitted with an integrated Pirani gauge head covering 1x10-4 to 1000 mbar for system monitoring.
Affects whether a separate gauge is needed and how pressure monitoring is integrated.
Source
“The IDP-35 and IDP-45 pumps can be fitted with an integrated Pirani gauge head to measure pressures in the range of 1x10-4 to 1000 mbar”
Installation
IDP-35 and IDP-45 support RS-232, RS-485, or analog interfaces for PC or PLC integration and remote control.
Determines control architecture and cabling for automated vacuum systems.
Source
“The IDP-35 and IDP-45 dry scroll pumps can be used with RS-232 or RS-485 or with an analog interface.”
Selection
The IDP-15 is described as the quietest pump available with the lowest vibration at inlet, and is offered with an inlet isolation valve.
Key selection criterion for noise-sensitive labs and vibration-sensitive instruments.
Source
“Quietest pump available”
Selection
The IDP-7 is noted for handling water vapor, making it suitable for vacuum ovens and general lab vacuum.
Important when selecting a pump for wet or condensable-vapor processes.
Source
“Handles water vapor”
Performance
IDP-35 and IDP-45 are hermetically sealed with magnetic torque coupling, ensuring complete isolation of pumped gases from the environment.
Required for rare gas and helium recirculation where leakage or contamination is unacceptable.
Source
“Hermetically sealed (magnetic torque coupling)”
Compliance
The IDP-4 undergoes rigorous helium permeation testing aligned with worldwide standards.
Provides confidence for helium-critical applications and regulatory documentation.
Source
“Rigorous He permeation testing aligned with worldwide standards”