Agilent Technologies PAL3 autosampler, automated solid phase extraction
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies PAL3 autosampler, automated solid phase extraction

Agilent Technologies PAL3 autosampler, automated solid phase extraction — Gas chromatography.

The PAL3 autosampler is a robotic front-end for GC and GC/MS that automates solid-phase extraction and SPME sample preparation. It is distinguished within the series by its RTC (robotic tool change) configuration, which enables automated SPE and SPME workflows. It serves as the sample-introduction and preparation stage for trace analysis, including pesticides in durian and aldehydes in beer.

Agilent Technologies GC Sample Preparation Introduction Series — Model Comparison & Selection

Model Sample volume, milliliterAnalysis time, minBake flow, mL/minBake temperature, °CBake time, minCalibration concentration, µg/ml
PAL3 autosampler, automated solid phase extraction Current ——————
GC autosampler, GC sample injection 0.00180———100
purge and trap, volatile organic compounds analysis ——2002802—
Reliable, consistent GC introduction of liquefied gases ——————

Full specifications

Core Performance & Functional Specifications
Carrier gas Helium
Carrier Gas Flow 1.2 mL/min
Number of target pesticides mfr 197
Analytical system mfr 3 GC/MSD
Sample matrix mfr 1 ratio
Recovery criterion mfr 70...120
Precision criterion (RSD) mfr 20
Quantification limit range mfr 3...100 ppb
MRM database size mfr 1100 compounds
MRMs per compound mfr 10
Flavor threshold, hexanal mfr 350 μg/L
Flavor threshold, furfural mfr 15.157 mg/L
Flavor threshold, phenylacetaldehyde mfr 105 μg/L
Flavor threshold, trans-2-nonenal mfr 0.1 μg/L
Calibration range, hexanal mfr 0.05...10 μg/L
Calibration range, furfural mfr 5...1,000 μg/L
Calibration range, phenylacetaldehyde mfr 0.1...50 μg/L
Calibration range, trans-2-nonenal mfr 0.025...5 μg/L
Coefficient of determination R2, hexanal mfr 0.999
Coefficient of determination R2, furfural mfr 0.998
Coefficient of determination R2, phenylacetaldehyde mfr 0.996
Coefficient of determination R2, trans-2-nonenal mfr 0.998
Repeatability %RSD (n = 3), hexanal in beer mfr 1.3...4.9
Repeatability %RSD (n = 3), furfural in beer mfr 1.3...2.8
Repeatability %RSD (n = 3), phenylacetaldehyde in beer mfr 0.9...3.7
Repeatability %RSD (n = 3), trans-2-nonenal in beer mfr 1.7...2.8
Method detection limit, hexanal mfr 0.0009 μg/L
Method detection limit, furfural mfr 0.52 μg/L
Method detection limit, phenylacetaldehyde mfr 0.015 μg/L
Method detection limit, trans-2-nonenal mfr 0.003 μg/L
Inlet liner mfr 5190 Ultra
Ion source drawout lens diameter mfr 9 mm
Autotune algorithm mfr etune etune
Software version mfr 10.1
Calibration range mfr 0.01...0.5 μg/mL
LLOQ mfr 0.02 μg/mL
Number of analytes tested mfr 100
Standards, Safety & General Data
Internal standard mfr Chlorpyrifos-
Ion source drawout lens part number mfr 3870

Description

Construction and positioning

Behavior and analytical performance

Installation and operation

Integration & Operating Notes

Control & I/O
Autosampler must be synchronized with the GC/MS run sequence

The PAL3 executes sample preparation steps (SPE, SPME) before injection. The GC/MS must be started only after the autosampler signals that the sample is ready; otherwise the injection may occur before extraction is complete, leading to low or erratic responses. Configure the start-signal handshake in the chromatography data system method.

Application notes: automated sample preparation workflows
Process Limits
Recovery and precision criteria define method validity

For the pesticide method, recovery must fall within 70–120% and precision RSD must not exceed 20%. Results outside these limits indicate a problem in the sample preparation or analytical sequence and should trigger investigation before reporting data.

Application note: 197 pesticides in durian

Selection & Integration Q&A

When should I choose the PAL3 autosampler with automated SPE over a basic GC autosampler or a headspace sampler in this series?

Choose this PAL3 configuration when the method requires automated solid-phase extraction or SPME sample preparation before GC injection, as in trace aldehyde or pesticide analysis. For simple liquid injection without preparation, a basic GC autosampler suffices; for volatile-only matrices, a headspace sampler is more appropriate.

Are there any environmental or duty-cycle limits for the PAL3 autosampler?

The verified facts do not include environmental or duty-cycle specifications. Consult the instrument manual for operating temperature, humidity, and duty-cycle limits before installing the system in a laboratory environment.

What consumables or accessories are required for the aldehyde-in-beer method that are not part of the base autosampler delivery?

The method specifies an Ultra inlet liner (part number beginning 5190) for the GC inlet. SPME fibers and SPE cartridges are also required for the preparation steps but are not listed in the verified facts; order them separately according to the application note.