Agilent Technologies Reliable, consistent GC introduction of liquefied gases
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies Reliable, consistent GC introduction of liquefied gases

Agilent Technologies Reliable, consistent GC introduction of liquefied gases — Gas chromatography.

The Agilent 990 Micro GC Micro Gasifier is a heated vaporizing inlet accessory that enables reliable, consistent GC introduction of liquefied gas samples onto the 990 Micro GC platform. It accepts sample pressures up to 100 kPa at the Micro GC inlet and delivers a factory-set 10 Psi outlet pressure at a 100 °C vaporizing regulator temperature. This modification is the liquefied-gas introduction path within the GC sample preparation series, distinct from autosampler and headspace siblings.

Agilent Technologies GC Sample Preparation Introduction Series — Model Comparison & Selection

Model Sample volume, milliliterAnalysis time, minBake flow, mL/minBake temperature, °CBake time, minCalibration concentration, µg/ml
Reliable, consistent GC introduction of liquefied gases Current ——————
GC autosampler, GC sample injection 0.00180———100
PAL3 autosampler, automated solid phase extraction ——————
purge and trap, volatile organic compounds analysis ——2002802—

Full specifications

Core Performance & Functional Specifications
Inlet Pressure 1000 Psi
Maximum sample pressure at Micro GC inlet mfr 100 kPa
Outlet pressure (factory set) mfr 10 Psi
Required Micro GC firmware version mfr 1.04 greater
Supported data system mfr 2.1
Required tool mfr -20 pcs
Dimensions, Weight & Mechanical Design
Mounting screws mfr 4 pcs
Standards, Safety & General Data
Manual part number mfr 3535
Filter kit part number mfr 3535
Gasifier mounting bracket part number mfr 3535
USB cable part number mfr 8121
Core ferrite part number mfr 9170

Description

Construction and series positioning

Behavior and process envelope

Installation, control and operation

Integration & Operating Notes

Process Limits
Keep sample pressure at or below 100 kPa at the Micro GC inlet

Exceeding the 100 kPa maximum sample pressure at the Micro GC inlet risks overpressurizing the gasifier inlet path and invalidating the introduction. Verify the sample delivery pressure before connection; the separate 1000 Psi inlet-pressure entry is not used because it conflicts with this kPa limit.

Factory Specification Sheet
Mounting
Use the dedicated bracket and four screws for the gasifier

The gasifier mounting bracket (part number 3535) and four mounting screws are the specified mounting hardware. Substituting other brackets can misalign the heated inlet path and the USB routing to the Micro GC.

990 Micro GC Micro Gasifier document, installation
Electrical
USB link requires the specified cable and core ferrite

Connection to the Micro GC uses USB cable part number 8121 with core ferrite part number 9170. The ferrite suppresses interference on the USB line; omitting it can degrade communication with the gasifier during runs.

990 Micro GC Micro Gasifier document, installation
Maintenance
Filter kit 3535 is the service item for the gasifier inlet

The filter kit (part number 3535) is listed as a separate service item. Plan filter replacement according to sample cleanliness; a loaded filter alters the inlet pressure seen by the gasifier and can cause inconsistent liquefied-gas introduction.

Factory Specification Sheet; Gasifier user manual G3535-90000, maintenance

Selection & Integration Q&A

When should I choose the Micro Gasifier instead of the autosampler or headspace siblings in this series?

Choose the Micro Gasifier when the sample is a liquefied gas that must be vaporized and introduced at a controlled 100 °C regulator temperature and 10 Psi outlet. Autosampler and headspace siblings handle liquid injection and headspace matrices, not liquefied-gas introduction.

What upstream and downstream conditions must the sample delivery and Micro GC meet for this gasifier?

Sample pressure at the Micro GC inlet must not exceed 100 kPa. The Micro GC must have firmware greater than 1.04 and data system 2.1. The gasifier connects via USB cable 8121 with core ferrite 9170 and mounts with bracket 3535 and four screws.

What installation and environment limits does the manual impose on the gasifier?

The gasifier is installed only in combination with the 990 Micro GC using the dedicated bracket and four screws. The vaporizing regulator is factory set to 100 °C, so the surrounding installation must tolerate that heated zone; firmware greater than 1.04 is required before operation.

What maintenance items and accessories are not included in the base delivery?

The filter kit (3535), gasifier mounting bracket (3535), four mounting screws, USB cable (8121) and core ferrite (9170) are listed separately from the base unit. Filter replacement is the routine service action; the manual G3535-90000 covers maintenance and troubleshooting.

Dimensional Drawings & Technical Documentation

Dimensional drawings

Reliable, consistent GC introduction of liquefied gases — drawing 1

Click drawing to inspect vector dimensions

Technical documentation