Agilent Technologies UHV-24/UHV-24p
Agilent Technologies UHV-24/UHV-24p — Vacuum technologies.
The UHV-24/UHV-24p is a nude Bayard-Alpert ionization gauge with dual field-replaceable filaments (tungsten or yttria-coated iridium) on a 2.75 in DN 40 ConFlat flange. The UHV-24 covers 1 mTorr to 2×10⁻¹⁰ Torr (x-ray limit 2×10⁻¹¹ Torr); the UHV-24p extends to 5×10⁻¹¹ Torr (x-ray limit 5×10⁻¹² Torr). It serves as the high-vacuum sensor in UHV systems, requiring an external controller for filament, grid and degas potentials.
Agilent Technologies Vacuum Transducers Series — Model Comparison & Selection
| Model | Measurement range | Atomic mass unit, grams | Electron charge, coulomb | Molar volume, liters | Max pressure, megapascal | Sensitivity |
|---|---|---|---|---|---|---|
| UHV-24/UHV-24p Current | 1 | 1.66 | 1602 | 22.41 | — | 25 |
| 531 Thermocouple Gauge, Rough Vacuum Pressure Measurement | — | 1.66 | 1602 | 22.41 | — | — |
| 571 Bayard-Alpert Gauge Tube | -3...2 | 1.66 | 1602 | 22.41 | 1.33×10⁻⁴ | 17 |
| 572 Tungsten Bayard-Alpert Gauge | -3...2 | 1.66 | 1602 | 22.41 | -3...2 | — |
| Cold Cathode Gauge Tube 525, High Vacuum (HV) Gauge | — | 1.66 | 1602 | 22.41 | — | — |
| Hot Filament Gauge MBA2-201, Ionization BA Gauge | — | 1.66 | 1602 | 22.41 | — | — |
| IMG-100 Inverted Magnetron Gauge | — | 1.66 | 1602 | 22.41 | — | — |
| IMG-300 UHV Inverted Magnetron Gauge | — | 1.66 | 1602 | 22.41 | — | — |
Full specifications
| Measurement range | 1 Torr |
| Sensitivity | 25 Torr |
| Collector Diameter | 0.007 in |
| Electron Charge | 1.602 coulomb |
| Molar Volume | 22.41 liters |
| Operation | 16 |
| X-ray Limit | 2 Torr |
| Acceleration due to gravity mfr | 9.8066 |
| Angstrom unit mfr | 10 m |
| Avogadro's number mfr | 6.02214076 |
| Boltzmann's constant mfr | 1.38064852 |
| Equivalent of heat mfr | 4.185 |
| Natural log base mfr | 2.7183 |
| Velocity of light mfr | 2.9979 |
| Velocity of sound mfr | 343 |
| Magnetic flux density mfr | 1 |
| Pressure conversion mfr | 1 Torr |
| Pumping speed conversion mfr | 1 L/s |
| Throughput conversion mfr | 1 cc/sec |
| Outgassing rate mfr | 2 |
| Pumping speed (nude ion gauge) mfr | 0.5 litres/sec |
| Degas duration mfr | 25 minute |
| Description of the UHV-24/UHV-24p Ionization Gauge mfr | 16 |
| Installation mfr | 18 |
| Technical Specifications mfr | 19 |
| Application Notes mfr | 21 |
| Degas mfr | 23 |
| Gas Correction Factors mfr | 26 |
| Accessories and spare parts mfr | 29 |
| X3004-64403 mfr | Filament replacement kit kit |
| UHV-24 pressure range mfr | 1 Torr |
| UHV-24p x-ray limit mfr | 5 Torr |
| UHV-24p sensitivity mfr | 20 |
| degas duration mfr | 25 minute |
| nude gauge pumping speed mfr | 0.5 litres/sec |
| Atomic Mass Unit | 1.6605 grams |
| Emission Current | 4 mA |
| Filament Voltage mfr | 3...4.5 current |
| Filament Current mfr | 2.5...3.5 current |
| Filament voltage mfr | 28 VDC |
| Grid voltage mfr | 180 VDC |
| Degas grid voltage mfr | 600 VDC |
| Degas emission current (total) mfr | 67 DC |
| Degas filament voltage mfr | 7 V |
| Ion current at 10-9 Torr mfr | 10 A |
| Normal operating electron current mfr | 4 mA |
| Degas Grid Voltage mfr | 600 VDC |
| Degas Emission Current (total) mfr | 67 DC |
| Degas Filament Voltage mfr | 7 V |
| Degas Power Requirements (40 W) mfr | 600 V |
| normal operating electron current mfr | 4 mA |
| degas power mfr | 40 W |
| Filament material mfr | Tungsten or yttria-coated iridium filaments fila |
| Filament supports material mfr | SST SST |
| Grid material mfr | SST SST |
| Feedthrough material mfr | SST, Alumina, Nickel Alloy Alloy |
| Filament-to-grid distance mfr | 0.050...0.070 in |
| Filament Supports material mfr | SST SST |
| Maintenance: Filament Replacement mfr | 24 |
| Tungsten Filament Replacement mfr | 24 |
| Yttria-coated iridium filament Replacement mfr | 25 |
| filament-to-grid distance mfr | 0.050...0.070 |
| Bakeout Temperature | 450 °C |
| Standard pressure mfr | 101325 Pa |
Description
Construction and series positioning
- Nude ionization gauge: tungsten or yttria-coated iridium filaments, stainless-steel grid and filament supports, 0.002 in collector, alumina/nickel-alloy feedthrough, 2.75 in DN 40 ConFlat flange.
- Two variants: UHV-24 (sensitivity 25/Torr, x-ray limit 2×10⁻¹¹ Torr, usable to 2×10⁻¹⁰ Torr) and UHV-24p (sensitivity 20/Torr, x-ray limit 5×10⁻¹² Torr, usable to 5×10⁻¹¹ Torr).
- Siblings in the series are rough-vacuum thermocouple gauges (531, 536) and Bayard-Alpert tubes (571, 572); this model is the UHV nude gauge, not a rough-vacuum sensor.
Behavior under load and in different modes
- Emission current is switched with pressure: 4 mA normal, 10 mA below 5×10⁻¹⁰ Torr, ≤0.1 mA above 1×10⁻⁴ Torr. Higher emission increases gauge pumping (~0.5 L/s typical), which can lower readings in small chambers.
- Degas is electron bombardment of the grid: 40 W, grid up to +600 VDC, total emission 67 mA DC, filament 7 V. Factory degassed; needed mainly below 1×10⁻⁸ Torr, ~25 min with XGS controllers.
- At 5×10⁻¹¹ Torr with UHV-24p at 10 mA, ion current is only 10 pA; cable leakage, capacitance changes and noise dominate. PTFE-insulated, short cables and ferrite beads are advised.
Installation, connection and operation
- Mount in any position; open location; filaments horizontal. Remove protective metal sleeve from collector wire before installation.
- Controller must supply filament 3–4.5 V (AC or DC) at 2.5–3.5 A, collector 0 V, filament 28 VDC, grid +180 VDC, and a separate degas mode.
- Maximum bakeout +450 °C, ramped, no thermal shock. Handle only by ceramic base with nylon gloves; no silicone grease (use Apiezon L sparingly); no Alconox on aluminum.
- Not intrinsically safe: do not turn on with combustible gases or mixtures. Controller and vacuum system separately grounded to common ground rated ≥10 A; no ground wire between chamber and controller chassis.
Integration & Operating Notes
Use 4 mA normally, 10 mA below 5×10⁻¹⁰ Torr, and ≤0.1 mA above 1×10⁻⁴ Torr. Higher emission also increases gauge pumping, which can bias readings in small chambers.
Repeated or extended bakeout must not exceed +450 °C. Thermal shock can damage the gauge; ramp temperature changes gradually.
The ground must support at least 10 A. Do not place a ground wire between the vacuum chamber and the controller chassis; improper grounding can expose personnel to 160–900 V.
The gauge is not intrinsically safe; the hot filament and high-voltage discharge can be ignition sources. Do not turn on the gauge while pumping or backfilling combustible gases or mixtures.
Place the long bracket so the filament sits 0.050–0.070 in from the grid, then push short brackets fully in before tightening. Incorrect spacing or tension changes sensitivity and can damage the filament. Handle only by the ceramic base with nylon gloves.
Yttria-coated iridium kit X3004-64403; UHV-24 tungsten kit 9710018. Each kit contains two filament assemblies and an Allen wrench. Order the kit matching the installed filament type.
Yttria-coated iridium runs cooler and outgases less but generates more particles and has shorter life in hydrogen and halogens. Prefer tungsten where hydrogen or halogens are present.
Selection & Integration Q&A
Choose UHV-24p when measurements below 2×10⁻¹⁰ Torr are required: its x-ray limit is 5×10⁻¹² Torr and usable range extends to 5×10⁻¹¹ Torr, versus 2×10⁻¹¹ Torr and 2×10⁻¹⁰ Torr for UHV-24. The 571/572 are separate Bayard-Alpert tubes; this nude gauge is for UHV, not rough vacuum.
An external controller must supply filament 3–4.5 V at 2.5–3.5 A, collector 0 V, filament 28 VDC, grid +180 VDC, and a degas mode (40 W, grid up to +600 VDC, 67 mA DC, 7 V filament). Use short PTFE-insulated cables; at 10 pA signals, leakage and noise dominate, so ferrite beads may be needed.
Mount in any position in an open location with filaments horizontal; remove the collector protective sleeve first. Maximum bakeout is +450 °C, ramped, no thermal shock. Ground controller and vacuum system separately to a common ground rated ≥10 A; never run a ground wire between chamber and controller chassis.
Filaments are field-replaceable: yttria-iridium kit X3004-64403 or UHV-24 tungsten kit 9710018 (each contains two assemblies and an Allen wrench). Degas is factory-performed; post-install degas (~25 min with XGS controllers) is only needed below 1×10⁻⁸ Torr. Controller, cables and filament kits are not included with the gauge.