BINDER B 28
BINDER ·Made in Germany · In Active Production

BINDER B 28

BINDER B 28 — Incubators with Temperature 30…70 °C.

The BINDER B 28 is a 28-litre naturally ventilated incubator with mechanical thermostat and Class 1 overtemperature protection, set for 30–70 °C. It differs from the E 28 oven sibling by its low-temperature incubation range and thermostat brake. It serves microbiology and sample incubation where precise, stable warmth near 37 °C is required.

Temperature
30…70 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

Full specifications

Core Performance & Functional Specifications
Temperature 30...70 °C
Interior volume 28 L
Load per rack 10 kg
Nominal power 0.25 kW
Permitted load 25 kg
Temperature fluctuation 1.0 K
Temperature uniformity 1.5 K
Back wall clearance 160 mm
Side wall clearance 100 mm (B028-230V / B028-120V)
Article Number mfr 9010
Dimensions, Weight & Mechanical Design
Depth 425 mm (B028-230V / B028-120V)
Height 405 mm (B028-230V / B028-120V)
Interior width 400 mm (B028-230V / B028-120V)
Width 580 mm (B028-230V / B028-120V)
Net weight of the unit (empty) mfr 23 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
120 V UL variant: 60 Hz
Rated voltage 230 V
120 V UL variant: 120 V
Unit fuse 6.3 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

The B 28 shares its stainless-steel V2A (1.4301 / AISI 304) inner chamber, pre-heating chamber and interior door sides with the E 28 drying/heating oven, but is configured as an incubator: mechanical thermostat, control thermometer, main power switch, pilot lamps and a thermostat brake for setting the Class 1 overtemperature safety device. The E 28 instead carries a 0–120 minute timer and a 60–230 °C range. The B 28 is therefore the low-temperature member of the 28-litre mechanical-adjustment pair, not a drying oven.

Working range and behaviour

Installation and operation

Integration & Operating Notes

Mounting
Keep 160 mm rear and 100 mm side clearance

Natural ventilation depends on free air around the housing; the manual specifies 160 mm at the rear, 100 mm at the sides, 250 mm between adjacent chambers and at least 200 mm above. The unit must not be stacked. Violating these distances risks overheating and invalidates the temperature uniformity figures.

Operating Manual B-E 01-2025, Section 3.4 Installation Location
Process Limits
Set-point must sit at least 5 K above ambient

The control range starts 5 °C above room temperature and ends at 70 °C, valid only for ambient +18 °C to +40 °C. A 37 °C incubation therefore requires ambient no higher than about 32 °C; in a warm room the chamber cannot regulate near the lower set-points.

Operating Manual B-E 01-2025, Section 2 Technical Description
Electrical
Match fuse count and plug to the voltage variant

The 230 V variant (Art. 9010-0002) uses a CEE 7/7 plug and two chamber fuses; the 120 V UL variant (Art. 9010-0067) uses a NEMA 5-15 plug and one chamber fuse. The 6.3 A unit fuse rating and 0.25 kW nominal power apply; a repeated fuse response requires BINDER Service rather than a higher-rated replacement.

Data Sheet Model B 28; Operating Manual B-E 01-2025, Section 2.3
Control & I/O
Thermostat brake sets the Class 1 safety limit

The overtemperature protective device per DIN 12880:2007 is adjusted with the thermostat brake on the front panel, separate from the main thermostat. It must not be bypassed or altered; the brake is the only means of setting the safety limit on this mechanical model.

Operating Manual B-E 01-2025, Sections 2.1 and 6
Maintenance
Annual service and electrical repeat inspection required

Maintenance must be performed at least once a year by licensed electricians or BINDER-authorized experts, with electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 annually and after any repair. The rear panel must not be removed by operating personnel; isolate power at the main switch and unplug before service.

Operating Manual B-E 01-2025, Section 8.2 Maintenance
Cooling & Media
Stainless interior resists corrosion but not halides

The inner chamber and interior door sides are V2A (1.4301 / AISI 304). Charging materials containing acids or halides can attack the stainless steel; the chamber must not become wet during operation or cleaning, and cleaning while switched on is prohibited.

Operating Manual B-E 01-2025, Sections 1.8 and 2

Selection & Integration Q&A

When should I choose the B 28 incubator instead of the E 28 oven in the same 28-litre mechanical series?

Pick the B 28 for incubation between 30 °C and 70 °C, typically near 37 °C, where the thermostat brake sets Class 1 overtemperature protection. Choose the E 28 only for drying or heat treatment from 60 °C to 230 °C with its 0–120 minute timer; the B 28 cannot reach those temperatures.

What supply and protection must be provided for the 230 V B 28 variant?

Provide a grounded 230 V 1~ 50/60 Hz outlet within reach of the fixed 1800 mm cable, using the CEE 7/7 plug. The unit draws 0.25 kW and carries a 6.3 A unit fuse plus two chamber fuses; the plug must remain accessible for emergency disconnection.

Can the B 28 be installed in a small warm room or stacked with another chamber?

No stacking is permitted. Ambient must stay within +18 °C to +40 °C at max. 70 % r.H. non-condensing, and the set-point must be at least 5 K above ambient, so a warm room limits low-temperature incubation. Keep 160 mm rear, 100 mm side and 250 mm between chambers.

What maintenance and consumables are required beyond the base delivery?

Plan annual maintenance by BINDER-authorized personnel plus electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702. Door gaskets are replaced only when cold. No cables or accessories are included beyond the fixed 1800 mm power cable; fuses must match the variant (two for 230 V, one for 120 V).

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model B 28 en
Datasheet ·439,6 КБ ·5 pp.
View PDF
PDF
B E 01 2025 en
Manual ·6,0 МБ ·49 pp.
View PDF