BINDER FD 115
BINDER FD 115 — Drying ovens with Temperature 10…300 °C.
The FD 115 is a forced-convection drying and heating oven (Avantgarde.Line) with a 116 L interior (550 × 550 × 385 mm) and a working range of 10–300 °C. It is the mid-size 115 chamber of the FD series, distinguished from smaller FD 23/56 and larger FD 260/720 siblings by its 1.3 kW single-phase 230 V supply and 30 kg per-rack load. It serves tempering, drying and heat treatment of solid or bulk materials where uniform airflow and 1.7 K temperature uniformity are required.
BINDER FD Series — Model Comparison & Selection
Full specifications
| Temperature | 10...300 °C |
| Air change rate | 32 x/h |
| Heating time to 150°C | 18 min |
| Interior volume | 116 L |
| Load per rack | 30 kg (FD115-230V) |
| Nominal power | 1.3 kW |
| Permitted load | 150 kg (FD115-230V) |
| Temperature fluctuation | 0.3 K |
| Temperature uniformity | 1.7 K |
| Rear wall clearance | 160 mm (FD115-230V) |
| Side wall clearance | 100 mm (FD115-230V) |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 150°C mfr | 4 min |
| Depth | 605 mm (FD115-230V) |
| Height | 735 mm (FD115-230V) |
| Interior depth | 385 mm (FD115-230V) |
| Interior height | 550 mm (FD115-230V) |
| Interior width | 550 mm (FD115-230V) |
| Width | 710 mm (FD115-230V) |
| Net weight of the unit (empty) mfr | 54 kg |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 6.3 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- Forced-convection (fan) oven with stainless-steel 1.4301 (AISI 304) inner chamber, pre-heating chamber and door interiors; external housing with triangular instrument panel and outer door.
- Interior volume 116 L; usable dimensions 550 W × 550 H × 385 D mm; net exterior 710 W × 735 H × 605 D mm; empty weight 54 kg.
- Within the FD series, the 115 sits between the compact FD 23 (5–300 °C) and FD 56 and the larger FD 260 and FD 720; unlike the 720 it runs on a single 230 V 1~ supply (1.3 kW, 6.3 A external fuse) and uses an IEC connector plug rather than a fixed cable.
- Overtemperature protective device class 1 trips at 350 °C for the FD 115 (FD sizes 56/115/260/720: 330/350/330/330 °C); safety controller class 2 per DIN 12880.
Behaviour under load and in operation
- Temperature range 10–300 °C with fluctuation 0.3 K and uniformity 1.7 K; heating from ambient to 150 °C takes 18 min, and recovery after a 30 s door opening at 150 °C is 4 min.
- Air-change rate 32 x/h; the fan speed is fixed on FD (no adjustable fan set-point, unlike BF/FED), so airflow is set by construction and loading discipline.
- Permitted load 150 kg total, 30 kg per rack; usable volume must not be loaded by more than half, and samples must keep at least 10 % of interior dimensions as wall clearance to preserve uniformity.
- Controller display resolves to 1 °C (FD series); the safety controller set-point is adjustable 10–300 °C and should be set 2–5 °C above the process set-point.
Installation, connection and operation
- Indoor installation on a non-flammable surface with ventilation; ambient 18–40 °C, ≤70 % r.h. non-condensing, altitude ≤2000 m; wall clearance 160 mm at the rear and 100 mm at the sides.
- Supply 230 V 1~ 50 Hz, ±10 % tolerance; unit fuse 6.3 A; IP 20 per IEC 60529, pollution degree 2, over-voltage category II per IEC 61010-1.
- Hot-air sterilization of the empty chamber is possible at 190 °C for at least 30 min; racks are removable and autoclavable.
Integration & Operating Notes
The chamber must not be installed in unventilated recesses; the prescribed rear and side clearances plus an 18–40 °C ambient range are required for the 1.3 kW heat load to dissipate and for the specified temperature uniformity to hold. Violating them risks overtemperature trips and invalid drying results.
The FD 115 draws 1.3 kW on a single-phase 230 V 1~ 50 Hz supply with a 6.3 A time-lag external chamber fuse. Supply fluctuations beyond ±10 % are outside approved conditions, and the upstream circuit must be sized for the 6.3 A fuse rather than the nominal current alone.
Samples must remain inside the usable volume with at least 10 % of the interior dimensions as clearance to walls, and the usable volume must not be loaded by more than half. Overloading blocks the forced airflow and destroys the 1.7 K uniformity; total load is also capped at 150 kg with 30 kg per rack.
The safety controller set-point is an absolute maximum limit (adjustable 10–300 °C on FD) and must be re-checked after every temperature set-point change. If the class 1 overtemperature device trips, the chamber must be disconnected, cooled below the reset threshold, examined and restarted; the controller display resolves to 1 °C on FD.
Selection & Integration Q&A
Pick the FD 115 for a 116 L usable volume with 550 × 550 × 385 mm interior and 30 kg per rack, still on a single 230 V 1~ 1.3 kW supply. The FD 56 is smaller; the FD 260 needs 2.3 kW and a 12.5 A fuse, so choose it only when the larger batch justifies the higher circuit rating.
230 V 1~ 50 Hz with ±10 % tolerance, 1.3 kW nominal, external 6.3 A time-lag chamber fuse (5×20 mm, 250 V). It uses an IEC connector plug with shockproof plug, not the fixed cable of size 720. IP 20, pollution degree 2, over-voltage category II per IEC 61010-1.
Install indoors on a non-flammable surface at 18–40 °C and ≤70 % r.h. non-condensing, below 2000 m altitude, with 160 mm rear and 100 mm side clearance and no unventilated recess. Load at most half the usable volume with 10 % wall clearance; total load 150 kg, 30 kg per rack.
Annual maintenance by BINDER-authorized personnel plus electrical testing per EN 50678/VDE 0701 and EN 50699/VDE 0702. Consumables not included: 6.3 A time-lag fuses, door gasket (replace cold only), neutral cleaning agent art. 1002-0016 and disinfectant spray art. 1002-0022. Racks are removable and autoclavable.