Quick Start Guide (Art. no. 7001-0329, issue 03/2023) for BINDER Avantgarde.Line chambers: BD series incubators with natural convection, BF series incubators with forced convection, ED series drying and heating ovens with natural convection, FD series drying and heating ovens with forced convection, and FED series drying and heating ovens with forced convection and enhanced timer functions. Chamber sizes covered by the electrical data table are 56, 115, 260 and 720. The guide is valid only together with the full operating manual (art. no. 7001-0292).

Section-by-Section Technical Breakdown

pp. 16
Intended use

Defines the approved applications: exact tempering of harmless materials, drying and heat treatment of solid, pulverized or bulk material, drying of glassware and warm storage of liquids in containers. BD and BF incubators are intended for incubation of cultures at a standard temperature of 37 °C / 98.6 °F. Charging material must not be explosive or flammable, must not form an explosive mixture with air, and the operating temperature must lie below its flash point or sublimation point.

  • No component of the charging material may release toxic gases.
  • A solvent content must not be explosive or flammable.
  • Operating temperature must be below the flash point or sublimation point of the charging material.
  • Modifications to the device by the user are not permitted.
  • Only original BINDER accessories or BINDER-authorized third-party accessories may be used.
pp. 16
Components and front view

Identifies the main external and internal components of the chambers: triangular instrument panel, outer door, controller display and functional buttons, door handle, rack, and glass door with glass door handle on BD and BF incubators.

  • Triangular instrument panel
  • Outer door and door handle
  • Controller display and functional buttons
  • Rack
  • Glass door and glass door handle (BD and BF only)
pp. 17
Controller overview and display

Describes the controller display variants and status indicators. Chambers without fan (BD, ED) or with fixed fan speed (FD) show the normal display; chambers with adjustable fan speed (BF, FED) have a different normal display. Display accuracy is 0.1 degrees for BD and BF incubators and 1 degree for ED, FD and FED ovens. Status symbols indicate heating active, timer operation, overtemperature alarm of the safety controller, and standby mode.

  • Temperature display accuracy: BD = 0.1 degrees; ED, FD = 1 degree.
  • Temperature display accuracy: BF = 0.1 degrees; FED = 1 degree.
  • Standby mode is entered by pressing the button for 3 seconds; press again 3 seconds to activate the display.
  • Status symbols: heating active, timer operation, overtemperature alarm of the safety controller, standby mode.
pp. 18
Requirements for the location of installation

Specifies indoor installation on a non-flammable surface with sufficient ventilation, permissible ambient conditions, and supply tolerances. Chambers must not be set up in non-ventilated recesses or in potentially explosive areas.

  • Permissible ambient temperature during operation: +18 °C up to +40 °C / 64.4 °F up to 104 °F.
  • Permissible ambient humidity: 70 % r.h. max., non-condensing.
  • Installation height: max. 2000 m / 6562 ft above sea level.
  • Power supply fluctuations must not exceed +/- 10 % of the nominal voltage.
  • Installation inside a building on a non-flammable surface.
pp. 18-19
Electrical connection data and installation

Provides nominal voltage, current type, nominal power, chamber fuse and overtemperature protective device class for each series and chamber size. ED/FD/FED size 720 come with a fixed power connection cable of at least 1800 mm; other chambers use an IEC connector plug with shockproof plug. Pollution degree 2 and over-voltage category II apply according to IEC 61010-1; housing protection is IP 20 according to IEC 60529.

  • BD: 230 V 1N~, 0.30/0.35/0.85/1.65 kW for sizes 56/115/260/720; external fuses 6.3/6.3/8.0/12.5 A; overtemperature protection class 1 at 120 °C.
  • BF: 230 V 1N~, 0.40/0.40/0.90/1.75 kW; external fuses 6.3/6.3/8.0/12.5 A; overtemperature protection class 1 at 120 °C.
  • ED: 230 V 1N~ for sizes 56/115/260 and 400 V 3N~ for size 720; 1.05/1.25/2.25/4.10 kW; external fuses 6.3/6.3/12.5 A, size 720 internal miniature circuit breaker 16 A; overtemperature protection class 1 at 330 °C.
  • FD, FED: 230 V 1N~ for sizes 56/115/260 and 400 V 3N~ for size 720; 1.10/1.30/2.30/4.50 kW; external fuses 6.3/6.3/12.5 A, size 720 internal miniature circuit breaker 16 A; overtemperature protection class 1 at 330/350/330/330 °C.
  • Only original BINDER connection cables may be used.
  • The power plug must remain easily accessible for complete disconnection.
pp. 19
Placing the charging material in the usable volume

Defines how to load the chamber for sufficient airflow and homogeneous temperature. Samples must stay inside the defined usable volume with at least 10 % of the interior dimensions as wall clearance, and the usable volume must not be loaded by more than half.

  • Take at least 10 % of the interior dimensions as wall clearance.
  • Do not load the usable volume by more than half.
  • Do not divide the usable volume with large-area samples.
  • Do not place samples too close to each other to permit circulation.
pp. 20
Set-point entry for temperature

Describes entering the temperature set-point via the controller menu and the relationship to the safety controller. The safety controller set-point is an absolute maximum limit temperature and should be set approximately 2 °C to 5 °C above the temperature set-point. On two-door chambers ED, FD and FED 720 the maximum adjustable temperature set-point depends on the ambient temperature.

  • Check and/or adjust the safety controller after any change of the temperature set-point.
  • Set the safety controller set-point by approx. 2 °C to 5 °C above the temperature set-point.
  • ED, FD, FED 720: maximum adjustable temperature set-point depends on ambient temperature.
pp. 21
Set-point entry for fan speed on chambers BF + FED

Describes entering the fan speed set-point on BF and FED chambers. The fan speed is adjusted in steps of 10 % and confirmed with OK.

  • Fan speed set-point is adjustable in steps of 10 %.
  • Access the fan speed setting by pressing the button twice (x2).
pp. 22
Setting the safety controller set-point

Describes the safety controller set-point entry menu and its permissible range. The set-point range is 10 °C up to 100 °C with BD and BF (accuracy 0.1 degrees) and 10 °C up to 300 °C with ED, FD and FED (accuracy 1 degree).

  • BD, BF: safety controller set-point 10 °C up to 100 °C, accuracy 0.1 degrees.
  • ED, FD, FED: safety controller set-point 10 °C up to 300 °C, accuracy 1 degree.
  • Access the menu with x4 button presses on chambers without fan and x5 on chambers with fan.
pp. 22
Troubleshooting in the event of overtemperature

Procedure when the overtemperature protection has turned off the heating. The chambers are equipped with an overtemperature protective device class 1 and a safety controller class 2 or 3.1. The chamber must be disconnected, cooled below the reset threshold or safety controller set-point, examined if appropriate, alarm reset on class 2, and restarted.

  • Disconnect the chamber from the power supply (pull the power plug).
  • Let the chamber cool below the pre-set reset temperature with class 1 or below the selected safety controller set-point value with class 2 and 3.1.
  • Have an expert examine and rectify the cause of the fault if appropriate.
  • With class 2: reset the alarm.
  • Restart the chamber.
pp. 23
Cleaning and decontamination

Specifies cleaning agents and methods for exterior surfaces, racks, door gaskets, instrument panel, zinc-coated hinge parts and rear chamber wall, and inner chamber. Neutral cleaning agent art. no. 1002-0016 is recommended; disinfectant spray art. no. 1002-0022 is recommended for chemical disinfection. Decontamination by hot air sterilization is possible for ED, FD and FED at 190 °C / 374 °F for at least 30 minutes, and hot-air disinfection for BD and BF at 100 °C / 212 °F. Racks can be removed and autoclaved.

  • Do not use a neutral cleaning agent on zinc-coated surfaces.
  • Recommended neutral cleaning agent: art. no. 1002-0016.
  • Recommended chemical disinfectant spray: art. no. 1002-0022.
  • ED, FD, FED: hot air sterilization at 190 °C / 374 °F for at least 30 minutes; remove all inflammable goods.
  • BD, BF: hot-air disinfection at 100 °C / 212 °F; remove all inflammable goods.
  • Racks can be removed and autoclaved.
  • Turn off the chamber at the main power switch (ED, FD, FED size 720) and disconnect the power plug before cleaning; completely dry before switching on again.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Permissible ambient temperature during operation is +18 °C up to +40 °C / 64.4 °F up to 104 °F. p.18 ↗ Install the chamber in a room whose temperature stays within this range; otherwise performance and safety limits may not be met. BD, BF, ED, FD, FED
Permissible ambient humidity is 70 % r.h. max., non-condensing. p.18 ↗ Avoid humid or condensing environments to prevent electrical and corrosion hazards. BD, BF, ED, FD, FED
Installation height is max. 2000 m / 6562 ft above sea level. p.18 ↗ Sites above this altitude are outside the approved installation conditions. BD, BF, ED, FD, FED
Power supply fluctuations must not exceed +/- 10 % of the nominal voltage. p.18 ↗ Verify supply quality before connection; nominal voltage is 230 V for sizes 56/115/260 and 400 V 3N~ for ED/FD/FED size 720. BD, BF, ED, FD, FED
BD and BF sizes 56/115/260/720: 230 V 1N~, nominal power 0.30/0.35/0.85/1.65 kW (BD) and 0.40/0.40/0.90/1.75 kW (BF); external chamber fuse 6.3/6.3/8.0/12.5 A time-lag (T) 5x20 mm / 250 V. p.18 ↗ Size the upstream circuit and fuse according to the chamber size and series; the chamber fuse is external. BD, BF
ED sizes 56/115/260/720: 230 V 1N~ for 56/115/260 and 400 V 3N~ for 720; nominal power 1.05/1.25/2.25/4.10 kW; external fuses 6.3/6.3/12.5 A, size 720 internal miniature circuit breaker 16 A. p.18 ↗ Size 720 requires a three-phase 400 V supply and has an internal 16 A miniature circuit breaker instead of an external chamber fuse. ED
FD and FED sizes 56/115/260/720: 230 V 1N~ for 56/115/260 and 400 V 3N~ for 720; nominal power 1.10/1.30/2.30/4.50 kW; external fuses 6.3/6.3/12.5 A, size 720 internal miniature circuit breaker 16 A. p.18 ↗ Size 720 requires a three-phase 400 V supply and has an internal 16 A miniature circuit breaker. FD, FED
Overtemperature protective device class 1 trip temperatures: 120 °C for BD and BF; 330 °C for ED; 330/350/330/330 °C for FD/FED sizes 56/115/260/720. p.18 ↗ The class 1 protection limits the maximum chamber temperature and defines the cooling threshold for reset after an overtemperature trip. BD, BF, ED, FD, FED
Pollution degree 2 and over-voltage category II according to IEC 61010-1; housing protection IP 20 according to IEC 60529. p.18 ↗ The chamber is protected against solid objects larger than 12.5 mm but not against water; do not spill water or cleaning agents over surfaces. BD, BF, ED, FD, FED
ED/FD/FED size 720 come with a fixed power connection cable of at least 1800 mm / 70.87 in; other chambers use an IEC connector plug with shockproof plug. p.18 ↗ Plan the installation position so the fixed cable reaches the supply; use only original BINDER connection cables. ED, FD, FED size 720; all other sizes
Take at least 10 % of the interior dimensions as wall clearance and do not load the usable volume by more than half. p.19 ↗ Overloading or placing samples too close to walls prevents sufficient airflow and homogeneous temperature distribution. BD, BF, ED, FD, FED
Set the safety controller set-point by approx. 2 °C to 5 °C above the temperature set-point. p.20 ↗ The safety controller set-point is an absolute maximum limit temperature; re-check it after every temperature set-point change. BD, BF, ED, FD, FED
Safety controller set-point range: 10 °C up to 100 °C with BD and BF (accuracy 0.1 degrees); 10 °C up to 300 °C with ED, FD and FED (accuracy 1 degree). p.22 ↗ The permissible safety limit differs by series; do not attempt to set values outside these ranges. BD, BF, ED, FD, FED
Fan speed set-point on BF and FED is adjustable in steps of 10 %. p.21 ↗ Fan speed can be tuned in 10 % increments to balance airflow and temperature uniformity. BF, FED
Temperature display accuracy: BD = 0.1 degrees, BF = 0.1 degrees; ED, FD, FED = 1 degree. p.17 ↗ Incubator readings resolve to 0.1 degrees while oven readings resolve to 1 degree; interpret set-points and readings accordingly. BD, BF, ED, FD, FED
ED, FD and FED can be hot air sterilized at 190 °C / 374 °F for at least 30 minutes; BD and BF can be hot-air disinfected at 100 °C / 212 °F. p.23 ↗ All inflammable goods must be removed from the interior before hot air sterilization or disinfection. ED, FD, FED; BD, BF
Recommended cleaning agents: neutral cleaning agent art. no. 1002-0016; disinfectant spray art. no. 1002-0022. p.23 ↗ Use these BINDER-recommended agents; do not use a neutral cleaning agent on zinc-coated surfaces such as hinge parts and rear chamber wall. BD, BF, ED, FD, FED
Lift size 56 and 115 with 2 people, size 260 with 4 people at its four lower corners, and size 720 with a fork lifter applied only from the rear in the middle of the chamber. p.17 ↗ Do not lift or transport the chamber by the door handle or door; transport only in original packaging with transport straps. BD, BF, ED, FD, FED
On two-door chambers ED, FD and FED 720 the maximum adjustable temperature set-point depends on the ambient temperature. p.20 ↗ Account for room temperature when selecting the maximum set-point on size 720 ovens. ED, FD, FED size 720

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Overtemperature alarm of the safety controller p.22 ↗ Overtemperature protection has turned off the heating; chamber equipped with overtemperature protective device class 1 and safety controller class 2 or 3.1. Disconnect the chamber from the power supply (pull the power plug). Let the chamber cool down below the pre-set reset temperature with class 1 or below the selected safety controller set-point value (class 2 and 3.1). If appropriate, have an expert examine and rectify the cause of the fault. With class 2: reset the alarm. Restart the chamber. Detailed information in the operating manual, chapter 7