BINDER FD 260
BINDER FD 260 — Drying ovens with Temperature 10…300 °C.
The FD 260 is a forced-convection drying and heating oven in BINDER's Avantgarde.Line, with a 259 L interior (650 × 780 × 515 mm) and a working range of +10 to 300 °C. It is the mid-large single-door 230 V variant of the FD series, distinguished from the 56/115 sizes by its 2.3 kW heating and 12.5 A fuse, and from the 720 by its single-phase supply. It serves drying, heat treatment and warm storage of solid, pulverized or bulk materials where airflow must be uniform.
BINDER FD Series — Model Comparison & Selection
Full specifications
| Temperature | 10...300 °C |
| Air change rate | 9 x/h |
| Heating time to 150°C | 19 min |
| Interior volume | 259 L |
| Load per rack | 40 kg (FD260-230V) |
| Nominal power | 2.3 kW |
| Permitted load | 270 kg (FD260-230V) |
| Temperature fluctuation | 0.4 K |
| Temperature uniformity | 1.9 K |
| Rear wall clearance | 160 mm (FD260-230V) |
| Side wall clearance | 100 mm (FD260-230V) |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 150°C mfr | 5 min |
| Depth | 825 mm (FD260-230V) |
| Height | 940 mm (FD260-230V) |
| Interior depth | 515 mm (FD260-230V) |
| Interior height | 780 mm (FD260-230V) |
| Interior width | 650 mm (FD260-230V) |
| Width | 810 mm (FD260-230V) |
| Net weight of the unit (empty) mfr | 85 kg |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 12.5 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- Forced-convection (fan) oven with stainless steel 1.4301 (AISI 304) inner chamber, pre-heating chamber and door interiors; exterior width 810 mm, depth 825 mm, height 940 mm, empty weight 85 kg.
- Within the FD series, the 260 sits between the 115 and the two-door 720: it shares the +10 to 300 °C range with the 115 and 56, while the 23 starts at +5 °C and the 720 at +12 °C. Unlike the 720, which requires 400 V 3N~ and an internal 16 A breaker, the FD 260 runs on 230 V 1~ 50 Hz with a 12.5 A external chamber fuse.
- The FD 260 is the forced-convection counterpart of the ED 260 (natural convection); the fan gives faster heat-up and tighter uniformity than the ED line.
Behavior under load and in operation
- Heating-up time to 150 °C is 19 min; temperature fluctuation 0.4 K and uniformity 1.9 K. Recovery after a 30 s door opening at 150 °C is 5 min.
- Air-change rate is 9 x/h. The controller displays temperature with 1-degree resolution (FD series), unlike the 0.1-degree BD incubators.
- The usable volume must not be loaded by more than half, and samples must keep at least 10 % of the interior dimensions as wall clearance; each rack is rated 40 kg and the total permitted load is 270 kg.
- Above 150 °C, metallic surfaces may discolor by oxidation; this is harmless and does not impair function.
Installation, connection and control
- Single-phase 230 V 1~ 50 Hz supply, nominal power 2.3 kW, unit fuse 12.5 A; supply fluctuation must stay within ±10 %. The chamber uses an IEC connector plug with shockproof plug (CEE 7/7).
- Overtemperature protection: class 1 device with 330 °C trip for the FD 260, plus safety controller class 2 (DIN 12880). The safety controller set-point range is 10–300 °C with 1-degree accuracy and should be set 2–5 °C above the process set-point.
- Wall clearances: 160 mm at the rear and 100 mm at the sides. Install indoors on a non-flammable surface, not in unventilated recesses or explosive areas.
- Hot-air sterilization of the interior is possible at 190 °C for at least 30 minutes; racks are removable and autoclavable.
Integration & Operating Notes
The FD 260 draws 2.3 kW on a single 230 V 1~ 50 Hz phase and relies on a 12.5 A external chamber fuse. Verify the supply quality and upstream circuit rating before connection; out-of-tolerance voltage or an undersized circuit can trip the fuse or prevent the chamber from reaching 300 °C.
The chamber must not be set up in unventilated recesses. The prescribed rear and side clearances are needed for heat dissipation and airflow; violating them risks overheating the housing and degrading temperature uniformity. Install indoors on a non-flammable surface.
Samples must stay inside the usable volume with at least 10 % of the interior dimensions as wall clearance, and the usable volume must not be loaded by more than half. Overloading blocks airflow and destroys the 1.9 K uniformity; each rack carries max 40 kg and the total permitted load is 270 kg.
The safety controller set-point is an absolute maximum limit temperature, adjustable from 10 to 300 °C with 1-degree accuracy on the FD series. Re-check it after every temperature set-point change; if the class 1 protection trips at 330 °C, the chamber must be disconnected, cooled below the reset threshold, examined and restarted.
Maintenance work must be performed at least once a year by licensed electricians or experts authorized by BINDER, and an annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended. Replace the door gasket only when the chamber is cold, otherwise the gasket may be damaged.
The FD 260 interior can be sterilized with hot air at 190 °C for at least 30 minutes; all inflammable goods must be removed first. Use neutral cleaning agent art. no. 1002-0016 and disinfectant spray art. no. 1002-0022; do not apply neutral cleaner to zinc-coated hinge parts or the rear chamber wall.
Selection & Integration Q&A
Pick the FD 260 for 259 L of usable volume on a single 230 V supply with 2.3 kW and a 12.5 A fuse. The FD 115 is smaller; the FD 720 is a two-door unit requiring 400 V 3N~ and an internal 16 A breaker, and its maximum set-point depends on ambient temperature.
A 230 V 1~ 50 Hz circuit within ±10 % fluctuation, sized for 2.3 kW, with a 12.5 A time-lag external chamber fuse. The chamber connects via an IEC connector plug with shockproof plug (CEE 7/7); no three-phase supply is needed.
Indoor installation on a non-flammable surface at +18 to +40 °C and max 70 % r.h. non-condensing, at altitudes up to 2000 m. Keep 160 mm at the rear and 100 mm at the sides; do not install in unventilated recesses or explosive areas.
Annual maintenance by BINDER-authorized personnel plus annual electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702. Consumables include neutral cleaning agent art. no. 1002-0016 and disinfectant spray art. no. 1002-0022; racks are removable and autoclavable.