BINDER FD 56
BINDER ·Made in Germany · In Active Production

BINDER FD 56

BINDER FD 56 — Drying ovens with Temperature 10…300 °C.

The FD 56 is a forced-convection drying and heating oven in BINDER's Avantgarde.Line, with a 60 L interior and a working range of 10–300 °C. It is the smallest FD size (56) and uses a fixed-speed fan, distinguishing it from adjustable-fan FED models and natural-convection ED ovens. It serves precise tempering, drying and heat treatment of solid or bulk materials in laboratory and production processes.

Temperature
10…300 °C

BINDER FD Series — Model Comparison & Selection

Model Temperature, °CAir change rate, 1/hourDepth, mmHeating up time to 150 c, minHeight, cmInterior volume, L
FD 56 Current 10–300805651562.560
FD 115 10–300326051873.5116
FD 23 5–300645202549.520
FD 260 10–30098251994259
FD 720 12–300587025159741

Full specifications

Core Performance & Functional Specifications
Temperature 10...300 °C
Air change rate 80 x/h
Heating time to 150°C 15 min
Interior volume 60 L
Load per rack 30 kg (FD056-230V / FD056UL-120V)
Nominal power 1.1 kW
Permitted load 70 kg (FD056-230V / FD056UL-120V)
Temperature fluctuation 0.3 K
Temperature uniformity 1.7 K
Rear wall clearance 160 mm (FD056-230V / FD056UL-120V)
Side wall clearance 100 mm (FD056-230V / FD056UL-120V)
Article Number mfr 9010
Recovery time after door was opened for 30 s at 150°C mfr 4 min
Dimensions, Weight & Mechanical Design
Depth 565 mm (FD056-230V / FD056UL-120V)
Height 625 mm (FD056-230V / FD056UL-120V)
Width 560 mm (FD056-230V / FD056UL-120V)
Net weight of the unit (empty) mfr 41 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 6.3 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behavior under load and in operation

Installation, connection and control

Integration & Operating Notes

Electrical
Supply must stay within ±10 % of nominal voltage

The FD 56 is rated at 230 V 1~ 50/60 Hz (or 120 V 1~ 60 Hz for the UL variant) with a nominal power of 1.1 kW. Power supply fluctuations beyond ±10 % of nominal voltage are outside the approved operating conditions and can affect heating performance and safety. Verify supply quality before connection and size the upstream circuit for the 6.3 A time-lag chamber fuse.

Quick Start Guide, Requirements for the location of installation / Electrical connection data, p.18
Mounting
Keep 160 mm rear and 100 mm side clearance

The chamber requires 160 mm wall clearance at the back and 100 mm at the sides. It must not be installed in unventilated recesses. The unit weighs 41 kg empty; lift size 56 with two people and do not lift or transport it by the door handle or door. Install indoors on a non-flammable surface.

Factory Specification Sheet; Quick Start Guide, Requirements for the location of installation, p.18; Operating Manual, Section 1.8, p.9
Process Limits
Load usable volume to no more than half

Samples must remain inside the usable volume with at least 10 % of the interior dimensions as wall clearance, and the usable volume must not be loaded by more than half. Overloading or placing samples too close to the walls prevents sufficient airflow and homogeneous temperature distribution, compromising the 1.7 K uniformity specification.

Quick Start Guide, Placing the charging material in the usable volume, p.19
Control & I/O
Set safety controller 2–5 °C above temperature set-point

The safety controller set-point is an absolute maximum limit temperature and must be set approximately 2 °C to 5 °C above the temperature set-point. The permissible range for the FD 56 is 10 °C to 300 °C with 1 °C accuracy. Re-check the safety controller set-point after every temperature set-point change. The class 1 overtemperature protective device trips at 330 °C.

Quick Start Guide, Set-point entry for temperature, p.20; Setting the safety controller set-point, p.22; Electrical connection data, p.18
Maintenance
Annual maintenance by BINDER-authorized personnel

Maintenance work must be performed at least once a year by licensed electricians or experts authorized by BINDER. Annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended, including before initial startup and after maintenance or repair. Replace the door gasket only when the chamber is cold.

Operating Manual, Section 10.1 and 10.2, p.36
Cooling & Media
Hot-air sterilization possible at 190 °C for 30 min

The FD 56 can be hot-air sterilized at 190 °C for at least 30 minutes; all inflammable goods must be removed from the interior beforehand. Racks can be removed and autoclaved. Use neutral cleaning agent art. no. 1002-0016 and disinfectant spray art. no. 1002-0022; do not use neutral cleaning agent on zinc-coated hinge parts and the rear chamber wall.

Quick Start Guide, Cleaning and decontamination, p.23

Selection & Integration Q&A

When should I choose the FD 56 over the FD 23 or FD 115?

Choose the FD 56 for 60 L loads needing 10–300 °C with forced convection and a 230 V single-phase supply. The FD 23 offers a lower 5 °C start but a smaller chamber; the FD 115 provides more volume at the same temperature range and electrical configuration.

What supply and fusing does the FD 56 require, and can I use a standard outlet?

The FD056-230V uses 230 V 1~ 50/60 Hz with a CEE 7/7 plug and an external 6.3 A time-lag fuse; the FD056UL-120V uses 120 V 1~ 60 Hz with a NEMA 5-15 plug. Supply fluctuation must stay within ±10 % of nominal voltage.

What ambient conditions and clearances must the installation site meet?

Install indoors on a non-flammable surface at +18 to +40 °C, max. 70 % r.h. non-condensing, max. 2000 m altitude, with 160 mm rear and 100 mm side clearance. Do not install in unventilated recesses or explosive areas.

What maintenance and accessories are required beyond the base delivery?

Annual maintenance by BINDER-authorized personnel and annual electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 are required. Optional accessories include the disconnectable audible over-temperature alarm and analog temperature output; cleaning agent 1002-0016 and disinfectant 1002-0022 are not included.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model FD 56 en
Datasheet ·388,4 КБ ·8 pp.
View PDF
PDF
QuickStartGuide BD BF ED FD FED E31 03 2023 dt en fr
Manual ·1,7 МБ ·35 pp.
View PDF
PDF
BD ED FD (E2) 09 2026 en
Manual ·12,9 МБ ·71 pp.
View PDF