BINDER FD 56
BINDER FD 56 — Drying ovens with Temperature 10…300 °C.
The FD 56 is a forced-convection drying and heating oven in BINDER's Avantgarde.Line, with a 60 L interior and a working range of 10–300 °C. It is the smallest FD size (56) and uses a fixed-speed fan, distinguishing it from adjustable-fan FED models and natural-convection ED ovens. It serves precise tempering, drying and heat treatment of solid or bulk materials in laboratory and production processes.
BINDER FD Series — Model Comparison & Selection
Full specifications
| Temperature | 10...300 °C |
| Air change rate | 80 x/h |
| Heating time to 150°C | 15 min |
| Interior volume | 60 L |
| Load per rack | 30 kg (FD056-230V / FD056UL-120V) |
| Nominal power | 1.1 kW |
| Permitted load | 70 kg (FD056-230V / FD056UL-120V) |
| Temperature fluctuation | 0.3 K |
| Temperature uniformity | 1.7 K |
| Rear wall clearance | 160 mm (FD056-230V / FD056UL-120V) |
| Side wall clearance | 100 mm (FD056-230V / FD056UL-120V) |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 150°C mfr | 4 min |
| Depth | 565 mm (FD056-230V / FD056UL-120V) |
| Height | 625 mm (FD056-230V / FD056UL-120V) |
| Width | 560 mm (FD056-230V / FD056UL-120V) |
| Net weight of the unit (empty) mfr | 41 kg |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 6.3 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- The FD 56 is a single-door, forced-convection oven with a stainless steel 1.4301 (AISI 304) inner chamber, pre-heating chamber and door interior. The fan runs at a fixed speed; unlike BF incubators and FED ovens, fan speed is not adjustable.
- Within the FD series, the 56 is the smallest chamber (60 L) and shares the 10–300 °C range with the FD 115 and FD 260; the FD 23 starts at 5 °C and the FD 720 at 12 °C. The FD 56 is the only FD size with a 230 V single-phase supply and an external 6.3 A time-lag fuse; the FD 720 requires 400 V 3N~ and has an internal 16 A miniature circuit breaker.
- The controller displays temperature with 1 °C resolution, consistent with the ED and FED oven lines, whereas BD and BF incubators resolve to 0.1 °C.
Behavior under load and in operation
- Temperature fluctuation is 0.3 K and temperature uniformity is 1.7 K, measured under the factory conditions for this model. Heating from ambient to 150 °C takes 15 min, and recovery after a 30 s door opening at 150 °C takes 4 min.
- The air-change rate is 80 x/h. The usable volume must not be loaded by more than half, and samples must keep at least 10 % of the interior dimensions as wall clearance to maintain airflow and homogeneity.
- The overtemperature protective device class 1 trips at 330 °C for the FD 56. The safety controller (class 2 per DIN 12880) set-point range is 10–300 °C with 1 °C accuracy and should be set 2–5 °C above the temperature set-point.
Installation, connection and control
- The FD 56 variant FD056-230V uses 230 V 1~ 50/60 Hz with a CEE 7/7 plug and an external 6.3 A time-lag fuse; the UL variant FD056UL-120V uses 120 V 1~ 60 Hz with a NEMA 5-15 plug. Nominal power is 1.1 kW.
- Required clearances are 160 mm at the rear and 100 mm at the sides. The chamber must be installed indoors on a non-flammable surface with sufficient ventilation, not in unventilated recesses or explosive areas.
- The controller supports temperature set-point entry with or without ramp function and continuous or timer operation. The RS 422 interface and APT-COM 4 software are listed as options for BD and ED, not for the FD series; the analog temperature output and disconnectable audible over-temperature alarm are available as options.
Integration & Operating Notes
The FD 56 is rated at 230 V 1~ 50/60 Hz (or 120 V 1~ 60 Hz for the UL variant) with a nominal power of 1.1 kW. Power supply fluctuations beyond ±10 % of nominal voltage are outside the approved operating conditions and can affect heating performance and safety. Verify supply quality before connection and size the upstream circuit for the 6.3 A time-lag chamber fuse.
The chamber requires 160 mm wall clearance at the back and 100 mm at the sides. It must not be installed in unventilated recesses. The unit weighs 41 kg empty; lift size 56 with two people and do not lift or transport it by the door handle or door. Install indoors on a non-flammable surface.
Samples must remain inside the usable volume with at least 10 % of the interior dimensions as wall clearance, and the usable volume must not be loaded by more than half. Overloading or placing samples too close to the walls prevents sufficient airflow and homogeneous temperature distribution, compromising the 1.7 K uniformity specification.
The safety controller set-point is an absolute maximum limit temperature and must be set approximately 2 °C to 5 °C above the temperature set-point. The permissible range for the FD 56 is 10 °C to 300 °C with 1 °C accuracy. Re-check the safety controller set-point after every temperature set-point change. The class 1 overtemperature protective device trips at 330 °C.
Maintenance work must be performed at least once a year by licensed electricians or experts authorized by BINDER. Annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended, including before initial startup and after maintenance or repair. Replace the door gasket only when the chamber is cold.
The FD 56 can be hot-air sterilized at 190 °C for at least 30 minutes; all inflammable goods must be removed from the interior beforehand. Racks can be removed and autoclaved. Use neutral cleaning agent art. no. 1002-0016 and disinfectant spray art. no. 1002-0022; do not use neutral cleaning agent on zinc-coated hinge parts and the rear chamber wall.
Selection & Integration Q&A
Choose the FD 56 for 60 L loads needing 10–300 °C with forced convection and a 230 V single-phase supply. The FD 23 offers a lower 5 °C start but a smaller chamber; the FD 115 provides more volume at the same temperature range and electrical configuration.
The FD056-230V uses 230 V 1~ 50/60 Hz with a CEE 7/7 plug and an external 6.3 A time-lag fuse; the FD056UL-120V uses 120 V 1~ 60 Hz with a NEMA 5-15 plug. Supply fluctuation must stay within ±10 % of nominal voltage.
Install indoors on a non-flammable surface at +18 to +40 °C, max. 70 % r.h. non-condensing, max. 2000 m altitude, with 160 mm rear and 100 mm side clearance. Do not install in unventilated recesses or explosive areas.
Annual maintenance by BINDER-authorized personnel and annual electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 are required. Optional accessories include the disconnectable audible over-temperature alarm and analog temperature output; cleaning agent 1002-0016 and disinfectant 1002-0022 are not included.