BINDER FP 115
BINDER FP 115 — Drying ovens with Temperature 10…300 °C.
The BINDER FP 115 is a forced-convection drying and heating oven with a 116 L interior and a working range of +10 to 300 °C. It is the mid-size 230 V single-phase model in the FP series, positioned between the FP 56 and FP 260, and delivers 1.4 kW with 32 air changes per hour. It is intended for reproducible drying, heating, and thermal testing of non-explosive, non-corrosive loads in laboratory and industrial processes.
BINDER FP Series — Model Comparison & Selection
Full specifications
| Temperature | 10...300 °C |
| Air change rate | 32 x/h |
| Heating time to 150°C | 19 min |
| Interior volume | 116 L |
| Load per rack | 30 kg (FP115-230V / FP115UL-120V) |
| Nominal power | 1.4 kW |
| Permitted load | 150 kg (FP115-230V / FP115UL-120V) |
| Temperature fluctuation | 0.3 K |
| Temperature uniformity | 1.4 K |
| Rear wall clearance | 160 mm (FP115-230V / FP115UL-120V) |
| Side wall clearance | 100 mm (FP115-230V / FP115UL-120V) |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 150°C mfr | 5 min |
| Depth | 636 mm (FP115-230V / FP115UL-120V) |
| Height | 732 mm (FP115-230V / FP115UL-120V) |
| Width | 714 mm (FP115-230V / FP115UL-120V) |
| Net weight of the unit (empty) mfr | 55 kg |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 10.0 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- The FP 115 uses a stainless-steel interior chamber with forced horizontal air circulation and a RD4 microprocessor controller. The interior measures 714 mm wide × 732 mm high × 636 mm deep (net), giving a usable volume of 116 L.
- Within the FP series, the FP 115 sits between the FP 56 (smaller) and FP 260 (larger). All share the +10 to 300 °C range, but the FP 115 is the smallest model available in a 230 V single-phase configuration with a 1.4 kW nominal power draw; the FP 720 requires 400 V three-phase.
- The unit is supplied with a CEE 7/7 plug for the 230 V variant (FP115-230V, Art. 9010-0364) and a NEMA 5-20 plug for the UL variant (FP115UL-120V, Art. 9010-0368).
Thermal behavior and load characteristics
- The chamber reaches 150 °C in approximately 19 minutes from ambient. Temperature fluctuation is 0.3 K and spatial uniformity is 1.4 K, enabling stable steady-state conditions for testing.
- After a 30-second door opening at 150 °C, the chamber recovers to setpoint within about 5 minutes, minimizing process interruption during loading or sampling.
- The air-change rate of 32 h⁻¹ supports drying processes that require removal of moisture or solvents. The air flap position can be adjusted via the controller to modulate fresh-air intake and exhaust.
- The maximum permitted load is 150 kg, with a maximum of 30 kg per rack. This limits the number and weight of shelves used in operation.
Installation, control, and operation
- The RD4 controller provides setpoint entry for temperature, fan speed, and air flap position. It includes a real-time clock, program functions (start/stop, week programs), and a data recorder with adjustable storage rate.
- A zero-voltage switching contact (via SUB-D socket 'ALARM / SWITCH CONTACT') allows external alarm or monitoring systems to be connected. An optional analog output for temperature is available via a separate SUB-D socket.
- Ethernet and USB interfaces enable network integration and data export. Only external devices compliant with EN 61010-1:2010 or EN 60950-1:2006 may be connected.
- The chamber is protected to IP 20 and must be installed indoors with minimum wall clearances: 160 mm at the rear and 100 mm at the sides. It must not be installed in unventilated recesses.
- The overtemperature protection can be configured as class 2 (temperature limiter) or class 3.1 (safety controller) per DIN 12880. The failsafe function and alarm buzzer are configurable via the controller.
Integration & Operating Notes
The chamber requires a minimum clearance of 160 mm at the rear and 100 mm at the sides for heat dissipation. Installing it in a recess or closer to walls can cause the unit to overheat, potentially damaging components and affecting temperature control. Ensure the installation location is indoors and well-ventilated.
The FP 115 is rated for 230 V ±10% at 50/60 Hz. Out-of-tolerance supply can cause the controller to malfunction or the heating to underperform, and may compromise safety. Verify the site voltage before connection and use a dedicated circuit rated for at least 10 A (unit fuse 10.0 A).
The chamber is not explosion-proof. Operating temperature must remain below the flash point or sublimation point of the charge, and no explosive mixture with air may form. Acids, halides, and other corrosive ingredients can damage the stainless steel, aluminum, and copper components; such damage is excluded from warranty.
Only connect IT equipment that meets EN 61010-1:2010 or EN 60950-1:2006 to the Ethernet and USB interfaces. Non-compliant devices can compromise the electrical safety of the system. Use the zero-voltage alarm contact for external monitoring rather than directly switching mains loads.
Schedule maintenance at least once a year; warranty is void if performed by non-authorized personnel. The controller displays 'Service due!' after one year of operation and repeats weekly until confirmed. Also perform annual electrical safety tests per EN 50678/VDE 0701 and EN 50699/VDE 0702.
Attempting to replace the door gasket while the chamber is hot can damage the gasket material and cause leaks. Allow the chamber to cool to ambient before service. Also, never remove the rear panel or allow water to penetrate the chamber during cleaning.
Selection & Integration Q&A
Choose the FP 115 when your load fits within 116 L and requires up to 300 °C with 1.4 kW single-phase power. The FP 56 offers smaller capacity (56 L) for lower throughput, while the FP 260 provides larger volume but requires a higher-power circuit. The FP 115 balances capacity and electrical demand for typical lab-scale drying.
Use the zero-voltage switching contact via the SUB-D socket 'ALARM / SWITCH CONTACT' on the rear. The contact is rated for low-voltage signals and can be wired to a BMS or alarm panel. For analog temperature output, the optional ANALOG OUTPUT socket is required. Ensure any connected device complies with EN 61010-1.
No. The chamber must not be installed in unventilated recesses. Maintain at least 160 mm clearance at the rear and 100 mm at the sides for heat dissipation. The installation site must be indoors and well-ventilated to prevent overheating and ensure temperature uniformity.
Annual maintenance by BINDER-authorized personnel is mandatory to preserve warranty, including electrical safety tests per EN 50678/50699. The base delivery does not include a Pt100 object temperature sensor, HEPA fresh air filter, or APT-COM 4 software; these must be ordered separately if needed. The service reminder appears after one year and must be confirmed.