BINDER M 115
BINDER ·Made in Germany · In Active Production

BINDER M 115

BINDER M 115 — Drying ovens with Temperature 23…300 °C.

The BINDER M 115 is a forced-convection drying and heating oven with a 113 L interior, operating from +23 to 300 °C via a 2.4 kW single-phase 230 V supply. It sits between the smaller M 56 and larger M 260/M 720 in the M series, distinguished by its 230 V single-phase connection and 105 air changes per hour. It is used for reproducible drying, heating, and thermal testing of laboratory and industrial loads up to 150 kg.

Temperature
23…300 °C

BINDER M Series — Model Comparison & Selection

Model Temperature, °CAir change rate, 1/hourDepth, mmHeating up time to 150 c, minHeight, cmInterior volume, L
M 115 Current 23–3001056361173.2113
M 260 14–300657861396.2256
M 56 23–300175596962.259
M 720 14–3004081614159735

Full specifications

Core Performance & Functional Specifications
Temperature 23...300 °C
Air change rate 105 x/h
Heating time to 150°C 11 min
Interior volume 113 L
Load per rack 30 kg
Nominal power 2.4 kW
Permitted load 150 kg
Temperature fluctuation 0.2 K
Temperature uniformity 1.0 K
Rear wall clearance 160 mm
Side wall clearance 100 mm
Article Number mfr 9010
Recovery time after door was opened for 30 s at 150°C mfr 6 min
Dimensions, Weight & Mechanical Design
Depth 636 mm
Height 732 mm
Width 710 mm
Net weight of the unit (empty) mfr 58 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 12.5 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

The M 115 (E3.1) is a forced-convection chamber with a 113 L usable volume, net dimensions of 710 mm wide × 732 mm high × 636 mm deep, and a net weight of 58 kg. Unlike the M 260 and M 720, which require a 400 V three-phase supply and have their main power switch on the rear panel, the M 115 operates on 230 V single-phase (1 N~, 50/60 Hz) with a nominal current of 10.5 A and a 12.5 A unit fuse. This makes it the largest M-series model that can be connected to a standard 230 V laboratory circuit. The M 56 shares the same +23 °C lower temperature limit, while the M 260 and M 720 extend down to +14 °C.

Thermal behavior and load capacity

The chamber uses the MB2 microprocessor controller with selectable overtemperature protection class 2 (temperature limiter) or class 3.1, per DIN 12880. The high air-change rate and forced convection make the M 115 suitable for processes requiring efficient moisture removal, but the load must not contain substances that are combustible or explosive at the working temperature; the set-point must remain below the flash point or sublimation point of the charge.

Installation, control, and operation

The M 115 requires 100 mm side clearance and 160 mm rear clearance, and must not be installed in unventilated recesses. It is rated IP 20 and must not be exposed to water or installed in damp areas. The rear panel provides a SUB-D socket for a zero-voltage relay alarm contact (standard) and an optional analog output for temperature. The MB2 controller supports fixed-value operation, timer programs, time programs, and week programs, with user management and an event list. Optional accessories include APT-COM 4 software (networking up to 100 chambers), a flexible Pt100 object temperature sensor, HEPA fresh air filter, and a mostly gas-tight version with inert gas connection.

Integration & Operating Notes

Electrical
Supply must be 230 V single-phase within ±10% at 50/60 Hz

The M 115 draws 10.5 A at 2.40 kW nominal. Verify that the branch circuit can supply this continuously and that the voltage stays within ±10% of 230 V at the indicated frequency; out-of-range supply can cause controller faults or heater damage. A 12.5 A unit fuse is fitted.

Instruction Manual M (E3.1), chap. 1.6 Type plate, p. 10

Selection & Integration Q&A

When should I choose the M 115 over the M 56 or M 260/M 720?

Choose M 115 when you need 113 L capacity and only have a 230 V single-phase supply; it is the largest M-series model on 230 V. The M 56 is smaller (56 L) but same temperature range; the M 260 and M 720 require 400 V three-phase and offer lower minimum temperature (+14 °C).

What are the requirements for connecting the M 115 to a controller or alarm system?

The rear provides a SUB-D socket for a zero-voltage relay alarm contact (standard) and an optional SUB-D analog output for temperature. For networked control, APT-COM 4 software can connect up to 100 chambers. Ensure the alarm contact is wired to a suitable monitoring device if unattended operation is required.

What maintenance is required and what accessories are not included?

Annual maintenance by BINDER-authorized personnel and annual electrical safety testing per EN 50678/VDE 0701 are required. Base delivery excludes APT-COM 4 software, Pt100 object temperature sensor, analog output, HEPA fresh air filter, and gas-tight/inert gas options. Door gasket replacement only when cold.