BINDER MKFT 115
BINDER MKFT 115 — Climate chambers with Temperature -70…180 °C.
The BINDER MKFT 115 is a 115 L alternating climate chamber with deep-temperature capability to -70 °C and humidity control from 10 to 95 % RH. It is the smallest deep-temperature model in the MKFT series, distinguished from the MKF 115 by its extended low-temperature range and from larger MKFT 240/720 by chamber volume. It serves as a programmable test platform for cyclic temperature/humidity stress screening of components and materials.
BINDER MKFT Series — Model Comparison & Selection
Full specifications
| Temperature | -70...180 °C |
| Interior volume | 115 L (MKFT115-400V) |
| Load per rack | 30 kg (MKFT115-400V) |
| Maximum heat compensation | 400 W |
| Nominal power | 5 kW (MKFT115-400V) |
| Article Number mfr | 9020 |
| Power frequency | 50 Hz (MKFT115-400V) |
| Rated voltage | 400 V |
| Unit fuse | 16.0 A (MKFT115-400V) |
| Humidity range | 95 |
| Cooling down time from 180°C to -70°C mfr | 120 min |
| Humidity fluctuation with humidity dependent on set value mfr | 2.5 % r.F. |
| Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr | 5.5 K/min |
Description
Construction and series positioning
The MKFT 115 (Art. No. 9020-0382, designation 115-400) is a 400 V 3~/N/PE 50 Hz chamber with a 115 L interior and a nominal power of 5 kW. Within the MKFT deep-temperature series it is the smallest volume; the MKFT 240 and MKFT 720 share the same -70…180 °C range but offer larger interiors. Compared with the MKF 115, the MKFT adds the deep-temperature stage, extending the lower limit from the MKF range to -70 °C. The standard model is the 400 V version; the -C suffix denotes a voltage and frequency changer variant (MKFT115-400V-C, Art. No. 9020-0362) for 480 V 3~ 60 Hz supplies.
Thermal and humidity behavior
- Temperature range: -70…180 °C; with humidity active, the range narrows to 10…90 °C.
- Humidity range: 95 to 10 % RH, with a humidity fluctuation of 2.5 % RH depending on set value.
- Temperature uniformity: 0.1…1.3 K; temperature fluctuation: 0.1…0.1 K (0.1…1 K with humidity).
- Average heating rate (IEC 60068-3-5, -40 to 180 °C): 5.5 K/min; average cooling rate (180 to -40 °C): 4.2 K/min.
- Cooling down time from 180 °C to -70 °C: 120 min.
- Maximum heat compensation: 400 W; load per rack: 30 kg.
These figures define the achievable ramp rates and the load the refrigeration system can offset during temperature transitions.
Installation, control and operation
The chamber requires a 400 V 3-phase supply with 16 A unit fuse. From size 115 on, a rear power switch is provided in addition to the main power switch; both must be on for commissioning. Water connections are required for humidification: freshwater supply and wastewater drain, with an internal freshwater can from size 115 on. The MB2 chamber controller supports fixed-value, time-program and week-program modes, with set-point ramps/steps, tolerance ranges and section repetition. Temperature safety devices include an over-temperature protective device (class 1) and a safety controller (class 2), with an optional over/under temperature safety device class 2. Zero-voltage switching outputs are available for external signaling. The controller logs events and supports a chart-recorder display.
Integration & Operating Notes
From size 115 on, the chamber has a rear power switch in addition to the main power switch. During commissioning and troubleshooting, both must be switched on; otherwise the chamber will not operate even with the main switch on. Verify this before diagnosing a no-power condition.
Although the chamber reaches -70…180 °C in dry operation, humidity control is only permitted between 10 and 90 °C. Programming humidity set points outside this window will not produce valid results and may trigger alarms. Plan test profiles so that humid phases stay within this temperature band.
The humidity sensor has resistance limits for certain substances: ammonia (NH3) maximum workplace threshold 20 ppm / 14 mg/m³ and permanent load 5500 ppm / 4000 mg/m³; ozone (O3) 0.1 ppm / 0.2 mg/m³ and 0.5 ppm / 1 mg/m³. Process atmospheres exceeding these concentrations can damage the sensor and invalidate humidity readings.
Regular maintenance is required at least once a year, including a leak test per Regulation (EU) 517/2014 (Article 4 and Article 10 (1)(b)). Document the leak test in the service log book. Additionally, an annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended.
After 8760 operating hours or two years, a service message appears on the controller. After confirmation it reappears every two weeks until reset by BINDER Service. Plan a service intervention at this interval to avoid recurring alarms and to maintain warranty compliance.
Condensers should be cleaned every 1 to 2 years by a qualified technician. With increased dust in ambient air, check the fan grid behind the left maintenance access flap every week and clean by suction after disconnecting the chamber. Neglecting this reduces cooling performance and can cause overheating.
The door gasket must be replaced only when the chamber is cold. Attempting replacement at elevated temperature can deform or damage the new gasket and compromise the seal, leading to temperature/humidity deviations and energy loss.
The humidification system needs a freshwater connection and a wastewater drain. From size 115 on, an internal freshwater can is provided for manual supply. Ensure water quality is suitable; steam humidifier maintenance intervals depend on water quality and steam output. A water circle lever for condensate recycling is available as an option for sizes 115 and above.
The chamber provides zero-voltage switching outputs that can be used to signal operating states or alarms to external equipment. These are dry contacts and require an external power source and wiring per the manual. Use them to integrate chamber status into a supervisory system.
Selection & Integration Q&A
Choose MKFT 115 when you need deep-temperature cycling to -70 °C in a compact 115 L chamber. The MKF 115 lacks the deep-temperature stage. The MKFT 240/720 offer larger volumes but same temperature range; select them only if 115 L is insufficient for your load.
The standard MKFT115-400V requires 400 V 3~/N/PE 50 Hz with a 16 A unit fuse. For 480 V 3~ 60 Hz supplies, specify the MKFT115-400V-C variant (Art. No. 9020-0362), which includes a voltage and frequency changer. Do not connect the standard model to 480 V.
Connect freshwater supply and wastewater drain. From size 115 on, an internal freshwater can is provided for manual supply. The steam humidifier requires maintenance at least once a year, with intervals depending on water quality and steam output. A water circle lever for condensate recycling is optional.
Plan annual maintenance including leak test per EU 517/2014 and electrical safety test per EN 50678/VDE 0701. Clean condensers every 1–2 years; in dusty air, check fan grid weekly. A service message appears at 8760 h or two years and requires BINDER Service reset. Door gasket replacement only when cold.