MKF+MKFT (E5) 10 2026
Technical Reference Summary
Covered Models & Document Scope
This operating manual (MKF/MKFT (E5) 10/2026, Art. No. 7001-0365) covers the BINDER MKF (E5) alternating climate chambers with program control and HIT Insulation® and the MKFT (E5) alternating climate chambers with deep temperature and program control. Covered models: MKF 56 (MKF056-230V, MKF056-240V), MKF 115 (MKF115-400V, MKF115-400V-C), MKF 240 (MKF240-400V, MKF240-400V-C), MKF 400 (MKF400-400V, MKF400-400V-C), MKF 720 (MKF720-400V, MKF720-400V-C), MKF 1020 (MKF1020-400V, MKF1020-400V-C), MKFT 115 (MKFT115-400V, MKFT115-400V-C), MKFT 240 (MKFT240-400V, MKFT240-400V-C) and MKFT 720 (MKFT720-400V, MKFT720-400V-C). The suffix -C denotes versions with a voltage and frequency changer. Art. Nos. 9020-xxxx and 9120-xxxx are listed per model version.
Section-by-Section Technical Breakdown
Describes the chamber overview, the instrument panel, the lateral control panel, the main power switch (MKF 56) and the rear power switch (MKF/MKFT from size 115 on), plus the rear chamber view. These sections identify the physical controls and connection points used during installation and start-up.
- MKF 56 uses a main power switch; MKF/MKFT from size 115 on additionally have a rear power switch (12).
- Instrument panel and lateral control panel are described for operating the MB2 chamber controller.
- Rear chamber view identifies connections and the rear power switch location.
Covers wastewater connection, freshwater supply (automatic via water pipe, manual via external freshwater can for MKF 56, manual via internal freshwater can from size 115 on), the water circle lever for condensate recycling, cooling water inlet/outlet for the water cooling option, connection kit and hose burst protection kit, installation of the voltage and frequency changer, and electrical connection including information for chambers with voltage and frequency changer.
- Freshwater can be supplied automatically via water pipe or manually via external can (MKF 56 option) or internal can (MKF/MKFT from size 115 on).
- Water circle lever for condensate recycling is an option for MKF/MKFT from size 115 on.
- Water cooling connections are an option for MKF sizes 56, 115, 240, 720, 1020 and MKFT 720.
- Electrical connection section includes specific information for chambers equipped with a voltage and frequency changer.
Describes operating functions in normal display, display views (normal, program, chart-recorder), controller icons, operating modes, menu structure (main menu, Settings, Service submenus), principle of controller entries, and controller performance during power failures and when opening the door.
- Three display views: normal display, program display and chart-recorder display.
- Menu structure includes main menu plus Settings and Service submenus.
- Controller behavior is defined for power failures and door opening.
Covers turning on the chamber, controller settings upon start up, and turning humidity control on/off. These steps follow electrical and water connection and precede set-point entry.
- Chamber is turned on at the main power switch (and rear power switch from size 115 on).
- Controller settings are made upon start up.
- Humidity control can be switched on and off.
Describes set-point entry through the “Setpoints” menu and direct setpoint entry via the normal display, plus special controller functions available in fixed-value mode.
- Setpoints can be entered via the “Setpoints” menu or directly in the normal display.
- Special controller functions are available in fixed-value operating mode.
Covers starting, cancelling and end-of-program behavior for the timer (stopwatch) program and for time programs, creating new time programs, program editor and section editor management, and value entry for program sections including section duration, set-point ramp/step, special controller functions, setpoint entry, tolerance range, section repetition and saving.
- Timer program and time programs have defined behavior during program delay time and after program end.
- Program editor manages programs; section editor adds, copies, replaces and deletes sections.
- Section values include duration, set-point ramp/step, tolerance range and repetition of sections.
Covers starting and cancelling week programs, creating new week programs, program editor and section editor management, and value entry for week program sections including set-point ramp/step modes, weekday, start time, setpoint entry and special controller functions.
- Week programs are managed via program editor and section editor.
- Section values include weekday, start time, set-point ramp/step mode and setpoint.
- Week programs can be started and cancelled.
Provides an overview of notifications and alarm messages, messages concerning the humidity system, the state of alarm, resetting an alarm and listing active alarms, and activating/deactivating the audible alarm buzzer.
- Alarm messages and humidity-system messages are listed separately.
- Active alarms can be listed and reset.
- Audible alarm (buzzer) can be activated or deactivated.
Describes the over temperature protective device (class 1), the safety controller (over temperature safety device class 2) including modes, setting, message and measures in the state of alarm, and function check, plus the optional over/under temperature safety device class 2.
- Class 1 over temperature protective device and class 2 safety controller are described.
- Safety controller has modes and a defined alarm state with messages and measures.
- A function check of the safety controller is specified.
- Over/under temperature safety device class 2 is available as an option.
Covers authorization levels and password protection, log in/log out, user change, password assignment and change, deleting a password for an authorization level, new password assignment when the password function was deactivated, and activation code entry.
- Authorization levels are password-protected.
- Passwords can be changed or deleted per authorization level.
- An activation code is used for controller functions.
Covers selecting menu language, setting date and time, selecting temperature unit, display configuration (display parameters and touchscreen calibration), network and communication (serial interfaces, Ethernet, e-mail), USB data transfer, and automatic turn-off of interior lighting.
- Controller language, date/time and temperature unit are configurable.
- Touchscreen can be calibrated.
- Serial interfaces, Ethernet and e-mail communication are configurable.
- Data transfer via USB interface is available.
Covers service contact page, current operating parameters, event list, technical chamber information, self-test function (MK 56), and the chart recorder display including views, legend handling, history display and parameter setting.
- Event list records controller events.
- Self-test function is described for MK 56.
- Chart recorder display supports legend pages, selective indications and history display.
Describes the function of the humidifying and dehumidifying system, including freshwater and wastewater paths. This section supports water connection and humidity control operation.
- Humidification and dehumidification system function is described.
- Freshwater and wastewater paths are covered.
Covers defrosting during refrigerating operation and the anti-condensation protection function.
- Defrosting behavior during refrigerating operation is described.
- Anti-condensation protection is described.
Describes the zero-voltage switching outputs available for external switching or signaling.
- Zero-voltage switching outputs are described for external use.
Covers APT-COM™ 4 Multi Management Software (including Basic Edition), RS485 interface, analog outputs for temperature and humidity, object temperature display with flexible Pt 100 sensor, compressed air connection, controlled compressed air dryer, water cooling, external freshwater and wastewater cans (MKF 56), BINDER Pure Aqua Service and water circle option.
- APT-COM™ 4 software and Basic Edition are described as accessories.
- RS485 interface and analog outputs for temperature and humidity are options.
- Object temperature display uses a flexible Pt 100 temperature sensor.
- Compressed air connection and controlled compressed air dryer are options.
- Water cooling is an option for MKF 56, 115, 240, 720 and MKFT 720.
- External freshwater and wastewater cans are an option for MKF 56.
Covers cleaning and decontamination/chemical disinfection of the inner chamber, including recommended disinfectant spray Art. No. 1002-0022 and procedures for biologically or chemically hazardous contamination.
- Use standard commercial surface disinfectants free from acid or halides, or alcohol-based solutions.
- Recommended disinfectant spray Art. No. 1002-0022.
- After disinfection, chamber must be absolutely dry and ventilated before start-up.
- Strongly contaminated inner chamber parts can be removed and sterilized in a sterilizer or autoclave.
Covers personnel qualification, maintenance intervals and service, simple troubleshooting by operating personnel, and returning the chamber to BINDER GmbH. Includes annual maintenance and leak test requirements, humidifier maintenance, condenser cleaning, fan grid checks, and the 8760-hour/two-year service message.
- Regular maintenance at least once a year; leak test per Regulation (EU) 517/2014.
- Steam humidifier maintenance at least once a year.
- Condensers recommended to be cleaned every 1 to 2 years by a qualified technician.
- Fan grid behind left maintenance access flap should be checked weekly in dusty ambient air.
- After 8760 operating hours or two years a service message appears and repeats every two weeks until reset by BINDER Service.
- Annual repeat electrical inspection recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702.
Covers factory calibration and adjustment, over current protection, definition of usable volume, MKF (E5) and MKFT (E5) technical data, equipment and options, accessories and spare parts, heating-up/cooling-down graphs, heat compensation graphs and dimensions.
- Technical data are provided separately for MKF (E5) and MKFT (E5).
- Usable volume definition and over current protection are specified.
- Heating-up, cooling-down and heat compensation graphs are provided.
- Dimensions are listed.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| MKF 56 is available as MKF056-230V (Art. No. 9020-0378, 9120-0378) and MKF056-240V (Art. No. 9020-0389, 9120-0389). p.1 ↗ | Select the correct voltage variant and order code for MKF 56 installations. | MKF 56 |
| MKF 115, 240, 400, 720 and 1020 are 400 V models; -C versions include a voltage and frequency changer (e.g., MKF115-400V-C Art. No. 9020-0357). p.1 ↗ | Use -C variants where a voltage and frequency changer is required; otherwise specify the standard 400 V model. | MKF 115/240/400/720/1020 |
| MKFT 115, 240 and 720 are 400 V deep-temperature models; -C versions include a voltage and frequency changer (e.g., MKFT115-400V-C Art. No. 9020-0362). p.1 ↗ | Select MKFT for deep-temperature applications and -C variants when a voltage and frequency changer is needed. | MKFT 115/240/720 |
| MKF/MKFT from size 115 on have a rear power switch (12) in addition to the main power switch; MKF 56 uses the main power switch. p.24 ↗ | During commissioning and troubleshooting, both the main power switch and the rear power switch must be on for sizes 115 and above. | MKF/MKFT from size 115 on; MKF 56 |
| Water cooling option is available for MKF sizes 56, 115, 240, 720, 1020 and MKFT 720. p.34 ↗ | Plan cooling water inlet and outlet connections only for these models when ordering the water cooling option. | MKF 56/115/240/720/1020 and MKFT 720 |
| External freshwater and wastewater cans are an option for MKF 56. p.32 ↗ | For MKF 56 without a fixed water pipe, use the external can option and follow the connection procedure. | MKF 56 |
| Water circle lever for condensate recycling is an option for MKF/MKFT from size 115 on. p.33 ↗ | Condensate recycling can be configured only on sizes 115 and above. | MKF/MKFT from size 115 on |
| Controlled compressed air dryer option is available for MKF 56, 115, 240, 720 and MKFT. p.125 ↗ | Specify the compressed air dryer option for these models when low-humidity or dry operation is required. | MKF 56/115/240/720 and MKFT |
| Recommended disinfectant spray Art. No. 1002-0022 for chemical disinfection of the inner chamber. p.133 ↗ | Use the recommended disinfectant to avoid corrosive damage; other disinfectants are excluded from BINDER liability. | whole series |
| After disinfection, the chamber must be absolutely dry and ventilated before start-up because explosive gases may form during decontamination. p.133 ↗ | Allow sufficient drying and ventilation time before returning the chamber to service. | whole series |
| Regular maintenance at least once a year; leak test per Regulation (EU) 517/2014 (Article 4 and Article 10 (1)(b)). p.134 ↗ | Schedule annual maintenance and document the leak test in the service log book. | whole series |
| Steam humidifier maintenance at least once a year; intervals depend on water quality and amount of steam produced. p.134 ↗ | Plan humidifier maintenance according to local water quality and usage. | whole series |
| Condensers recommended to be cleaned every 1 to 2 years by a qualified technician. p.134 ↗ | Include condenser cleaning in the maintenance plan. | whole series |
| With increased dust in ambient air, clean the condenser fan several times a year; check fan grid behind the left maintenance access flap every week and clean by suction after disconnecting the chamber. p.135 ↗ | In dusty environments, increase fan grid inspection frequency to weekly. | whole series |
| After 8760 operating hours or two years a service message appears; after confirmation it reappears every two weeks until reset by BINDER Service. p.135 ↗ | Plan BINDER Service intervention to reset the service message at the two-year/8760-hour mark. | whole series |
| Annual repeat electrical inspection and test before initial startup and before resuming operation after maintenance or repair, recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702. p.134 ↗ | Include electrical safety testing in commissioning and annual maintenance schedules. | whole series |
| Replace the door gasket only when cold. p.134 ↗ | Allow the chamber to cool before door gasket replacement to avoid damaging the gasket. | whole series |
| Humidity sensor resistance limits include, for example, ammonia (NH3) maximum workplace threshold 20 ppm / 14 mg/m3 and permanent load 5500 ppm / 4000 mg/m3; ozone (O3) 0.1 ppm / 0.2 mg/m3 and 0.5 ppm / 1 mg/m3. p.20 ↗ | Verify that process atmospheres do not exceed the listed concentration limits for the humidity sensor. | whole series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Chamber without function. p.135 ↗ | No power supply. | Check connection to power supply. Check whether the chamber is turned on at the main power switch. | Chap. 23.3 |
| Chamber without function. p.135 ↗ | Wrong voltage. | Check power supply for correct voltage (chap. 4.8). | Chap. 23.3 |
| Chamber without function. p.135 ↗ | Main power switch (1) not turned on. | Turn on the main power switch (1). | Chap. 23.3 |
| Chamber without function. p.135 ↗ | MKF/MKFT from size 115 on: Rear power switch (12) not turned on. | Turn on the rear power switch (12). | Chap. 23.3 |