BINDER UF V 700
BINDER UF V 700 — Ultra-low temperature freezers with Temperature -90…-40 °C.
The BINDER UF V 700 is a 736 L ultra-low temperature freezer using a two-stage cascade refrigeration system with R290/R170 hydrocarbons, rated for -90 to -40 °C. As the largest standard 230 V model in the UF V series, it is distinguished from the UF V 350 and UF V 500 by its 1,110 mm interior width and 1.6 kW nominal power. It serves as a high-capacity storage unit for temperature-sensitive samples in laboratory and process environments.
Direct drop-in replacements
BINDER UF V Series — Model Comparison & Selection
Full specifications
| Temperature | -90...-40 °C |
| Interior volume | 736 L (all variants) |
| Nominal power | 1.6 kW |
| Temperature fluctuation (-80 °C) | 1.5 K |
| Temperature uniformity (-80 °C) | 2.5 K |
| Article Number mfr | 9020 |
| Depth | 938 mm (all variants) |
| Height | 1966 mm (all variants) |
| Interior width | 1110 mm (all variants) |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Cool-down time | 450 min |
Description
Construction and Series Positioning
The UF V 700 (Article 9020) is the largest chamber in the UF V series, offering 736 L of interior volume with an interior width of 1,110 mm. Unlike the UF V 350, which uses support feet for leveling, the UF V 700 is installed with rear-wall spacers and adjustable shelves. The standard 230 V version (UFV700-230V, Art.-No. 9020-0348) is one of four electrical variants; water-cooled (UFV700-230V-W), 115 V UL, and 208–230 V UL versions exist but are distinct models. The chamber uses a two-stage cascade refrigeration system with R290 (0.150 kg) in stage 1 and R170 (0.150 kg) in stage 2, both flammable hydrocarbons. The RD4 controller provides set-point entry, alarm functions, and optional Ethernet/USB data interfaces.
Thermal Behavior and Control
- The nominal temperature range is -90 to -40 °C, with a factory-stated cooling-down time of 450 min to reach -90 °C.
- At -80 °C, temperature fluctuation is 1.5 K and uniformity is 2.5 K, defining the achievable stability for stored goods.
- The safety controller is a temperature safety device per DIN 12880, class 3.1, with adjustable modes and alarm thresholds.
- The RD4 controller supports tolerance-range alarms, door-open alarms, and a zero-voltage relay output for external monitoring.
Installation and Operation
- Electrical supply must be 230 V / 50 Hz, 1 N~, with a nominal power of 1.6 kW. Voltage tolerance is ±10%.
- The chamber requires rear-wall spacing per the manual and must not be installed in damp areas (IP 20).
- Optional CO2 emergency cooling and door access system are available for the UF V 700, but these options affect certification marks (GS mark not applicable with CO2 or door access; cUL not applicable with CO2).
- The Ethernet interface allows network configuration (MAC, IP, subnet, gateway, DNS) for remote monitoring; the USB interface supports data transfer via stick.
Integration & Operating Notes
The chamber must be installed with spacers to maintain distance from the rear wall, ensuring airflow over the condenser. Insufficient clearance leads to reduced cooling capacity and potential overheating. The installation location must be dry (IP 20) and free of ignition sources due to flammable refrigerants R290 and R170.
The RD4 controller provides a volt-free relay output that changes state on alarm conditions (e.g., temperature deviation, power failure, sensor failure). This contact must be wired to the site monitoring system or alarm panel according to the normally-open/normally-closed logic defined in the manual. Failure to connect it leaves the chamber without remote alarm notification.
Door handle, hinges, and locking counterpart must be cleaned and lubricated with medical vaseline at least once a year. Additionally, an electrical safety inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required prior to initial startup and after any maintenance or repair, and annually thereafter. Neglecting these tasks can lead to mechanical wear or electrical hazards.
The refrigerating system operates at a maximum pressure of 28 bar. Any service or connection work on the refrigeration circuit must not exceed this limit. The use of flammable hydrocarbons (R290/R170) requires that only qualified personnel perform service, and the area must be ventilated and free of ignition sources.
Selection & Integration Q&A
Choose the UF V 700 when you need the largest capacity: 736 L interior volume and 1,110 mm interior width. The UF V 500 and UF V 350 offer smaller volumes. The UF V 700 also supports options like CO2 emergency cooling and door access system, which are not available on the UF V 350. If space or power is limited, the smaller models may suffice.
Annual maintenance includes cleaning and lubricating the door lock with medical vaseline, cleaning the condenser air filter, and an electrical safety test per EN 50678/VDE 0701. The base delivery does not include storage racks, cryo boxes, CO2 emergency cooling, door access system, battery-buffered alarm, analog output, or advanced voltage booster. These must be ordered as options. A connection kit and hose burst protection are available for water-cooled variants.