UF V 700 drawing
BINDER ·Made in Germany · In Active Production

BINDER UF V 700

BINDER UF V 700 — Ultra-low temperature freezers with Temperature -90…-40 °C.

The BINDER UF V 700 is a 736 L ultra-low temperature freezer using a two-stage cascade refrigeration system with R290/R170 hydrocarbons, rated for -90 to -40 °C. As the largest standard 230 V model in the UF V series, it is distinguished from the UF V 350 and UF V 500 by its 1,110 mm interior width and 1.6 kW nominal power. It serves as a high-capacity storage unit for temperature-sensitive samples in laboratory and process environments.

Temperature
-90…-40 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER UF V Series — Model Comparison & Selection

Model Temperature, °CCooling down time, minInterior volume, LInterior width, mmTemperature fluctuation at 80 c, KTemperature uniformity at 80 c, K
UF V 700 Current -90–-4045073611101.52.5
UF V 350 -90–-4021040270643.5
UF V 500 -90–-403605018261.52.5

Full specifications

Core Performance & Functional Specifications
Temperature -90...-40 °C
Interior volume 736 L (all variants)
Nominal power 1.6 kW
Temperature fluctuation (-80 °C) 1.5 K
Temperature uniformity (-80 °C) 2.5 K
Article Number mfr 9020
Dimensions, Weight & Mechanical Design
Depth 938 mm (all variants)
Height 1966 mm (all variants)
Interior width 1110 mm (all variants)
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Operating Conditions, Environment & Protection
Cool-down time 450 min

Description

Construction and Series Positioning

The UF V 700 (Article 9020) is the largest chamber in the UF V series, offering 736 L of interior volume with an interior width of 1,110 mm. Unlike the UF V 350, which uses support feet for leveling, the UF V 700 is installed with rear-wall spacers and adjustable shelves. The standard 230 V version (UFV700-230V, Art.-No. 9020-0348) is one of four electrical variants; water-cooled (UFV700-230V-W), 115 V UL, and 208–230 V UL versions exist but are distinct models. The chamber uses a two-stage cascade refrigeration system with R290 (0.150 kg) in stage 1 and R170 (0.150 kg) in stage 2, both flammable hydrocarbons. The RD4 controller provides set-point entry, alarm functions, and optional Ethernet/USB data interfaces.

Thermal Behavior and Control

Installation and Operation

Integration & Operating Notes

Mounting
Rear-wall spacers required for adequate ventilation

The chamber must be installed with spacers to maintain distance from the rear wall, ensuring airflow over the condenser. Insufficient clearance leads to reduced cooling capacity and potential overheating. The installation location must be dry (IP 20) and free of ignition sources due to flammable refrigerants R290 and R170.

UF V (E3) 09-2026_en.pdf, Section 4.3 Installation and Connections
Control & I/O
Zero-voltage relay can trigger external alarm on fault

The RD4 controller provides a volt-free relay output that changes state on alarm conditions (e.g., temperature deviation, power failure, sensor failure). This contact must be wired to the site monitoring system or alarm panel according to the normally-open/normally-closed logic defined in the manual. Failure to connect it leaves the chamber without remote alarm notification.

UF V (E3) 09-2026_en.pdf, Section 14.5 Alarm Functions
Maintenance
Annual lubrication of door lock and electrical safety test

Door handle, hinges, and locking counterpart must be cleaned and lubricated with medical vaseline at least once a year. Additionally, an electrical safety inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required prior to initial startup and after any maintenance or repair, and annually thereafter. Neglecting these tasks can lead to mechanical wear or electrical hazards.

UF V (E3) 09-2026_en.pdf, Section 25.1 and 25.2.4
Process Limits
Maximum refrigerant pressure is 28 bar

The refrigerating system operates at a maximum pressure of 28 bar. Any service or connection work on the refrigeration circuit must not exceed this limit. The use of flammable hydrocarbons (R290/R170) requires that only qualified personnel perform service, and the area must be ventilated and free of ignition sources.

Type plate / p.10-11

Selection & Integration Q&A

When should I choose the UF V 700 over the UF V 500 or UF V 350?

Choose the UF V 700 when you need the largest capacity: 736 L interior volume and 1,110 mm interior width. The UF V 500 and UF V 350 offer smaller volumes. The UF V 700 also supports options like CO2 emergency cooling and door access system, which are not available on the UF V 350. If space or power is limited, the smaller models may suffice.

What maintenance tasks and accessories are required for the UF V 700, and what is not included in the base delivery?

Annual maintenance includes cleaning and lubricating the door lock with medical vaseline, cleaning the condenser air filter, and an electrical safety test per EN 50678/VDE 0701. The base delivery does not include storage racks, cryo boxes, CO2 emergency cooling, door access system, battery-buffered alarm, analog output, or advanced voltage booster. These must be ordered as options. A connection kit and hose burst protection are available for water-cooled variants.

Dimensional Drawings & Technical Documentation

Dimensional drawings

UF V 700 — drawing 1
performance curve

Click drawing to inspect vector dimensions

Technical documentation

PDF
Data Sheet Model UF V 700 en
Datasheet ·595,9 КБ ·9 pp.
View PDF
PDF
UF V (E3) 09 2026 en
Manual ·10,2 МБ ·127 pp.
View PDF