Shimadzu Axima Confidence
Shimadzu ·Made in Japan · In Active Production

Shimadzu Axima Confidence

Shimadzu Axima Confidence — Mass spectrometers.

The Axima Confidence is a MALDI time-of-flight mass spectrometer with a fixed-focus 337 nm nitrogen laser (3 ns pulse, 100 µJ, 50 Hz) and near-normal on-axis irradiation. It covers linear mode from 1 Da to 500 kDa and reflectron mode with >15,000 FWHM resolution, using 96/384-well microtitre or FlexiMass slide-format targets. It serves as a benchtop workhorse for routine protein and peptide mass analysis in life-science laboratories.

Direct drop-in replacements

0%
match
⚖️ Compare
0%
match
⚖️ Compare

Full specifications

Core Performance & Functional Specifications
Laser repetition rate (N2 laser) mfr 50 Hz (N2 laser)
Laser type mfr N2
Resolution (ACTH fragment 7-38) mfr >15,000 FWHM (ACTH fragment 7-38)
Ion source (near-normal (on-axis) laser irradiation) mfr MALDI (near-normal (on-axis) laser irradiation)
Sample target format (FlexiMass microscope slide-format) mfr plain or 48 well (FlexiMass microscope slide-format)
Sensitivity (linear mode, bovine serum albumin) mfr 250 fmol (linear mode, bovine serum albumin)
Target thickness (plain, 96 and 384 sample targets) mfr 2 mm (plain, 96 and 384 sample targets)
Maximum target thickness (with optional adaptor) mfr 10 mm (with optional adaptor)
Turbomolecular pump pumping speed (nominal, sample introduction, rotary backing) mfr 250 l/s (nominal, sample introduction, rotary backing)
Turbomolecular pump pumping speed (nominal, analyzer, two pumps with rotary backing) mfr 250 l/s (nominal, analyzer, two pumps with rotary backing)
Camera magnification (monochrome CCD camera) mfr 25 x (monochrome CCD camera)
Laser wavelength (nitrogen laser, fixed focus) mfr 337 nm (nitrogen laser, fixed focus)
Laser energy (nominal, per laser shot) mfr 100 μJ (nominal, per laser shot)
Maximum laser pulse rate (50 laser shots per second) mfr 50 Hz (50 laser shots per second)
Transient recorder sampling rate (8 bit, 16 bit accumulator) mfr 2 GHz (8 bit, 16 bit accumulator)
Second transient recorder sampling rate (neutral detection, 8 bit, 64 kB RAM) mfr 125 MHz (neutral detection, 8 bit, 64 kB RAM)
RAM (control PC) mfr 16 GB (control PC)
HDD storage (control PC) mfr 256 GB (control PC)
SSD storage (control PC) mfr 1 TB (control PC)
Monitor size (control PC) mfr 24 in (control PC)
Operating system (control PC and LAUNCHPADTM) mfr Microsoft® Windows® 10 64-bit (control PC and LAUNCHPADTM)
Dimensions, Weight & Mechanical Design
Mass range (linear mode, bovine serum albumin) mfr 55000-80000 m/z (linear mode, bovine serum albumin)
Detector types (optional removal of unwanted low mass ions) mfr patented beam blanker (optional removal of unwanted low mass ions)
Mass range (linear mode) mfr 1 Da to 500 kDa (linear mode)
Mass accuracy (linear mode, with internal calibration) mfr <30 ppm (linear mode, with internal calibration)
XY stage step (microtiter plate footprint MALDI target) mfr 10 μm (microtiter plate footprint MALDI target)
XY stage repeatability (microtiter plate footprint MALDI target) mfr 10 μm (microtiter plate footprint MALDI target)
Laser pulse width mfr 3 ns
Linear flight tube drift length (linear mode) mfr 1.2 m (linear mode)
Reflectron effective drift length (reflectron mode) mfr 2.0 m (reflectron mode)
Electrical, Power & Drive Specifications
Ion extraction energy (linear mode, variable, software controlled) mfr +25 kV/-20 kV kV (linear mode, variable, software controlled)
Ion extraction energy (reflectron mode, variable, software controlled) mfr +20 kV/-20 kV kV (reflectron mode, variable, software controlled)
Standards, Safety & General Data
Sample target format (standard microtitre plate-format) mfr 96 or 384 well (standard microtitre plate-format)

Description

Construction and positioning

Behavior under load and in different modes

Installation, control and operation

Integration & Operating Notes

Mounting
Target thickness above 2 mm needs the optional adaptor

Standard 96- and 384-well microtitre-format targets are 2 mm thick; the sample introduction stage accepts only this thickness unless the optional adaptor is fitted, which raises the maximum to 10 mm. Loading an unadapted thick target risks jamming the vacuum lock and misaligning the laser focus.

Technical data leaflet MO302; document digest mo305
Process Limits
Linear-mode mass accuracy depends on internal calibration

The <30 ppm figure applies only with internal calibration in linear mode; without calibrant peaks in the same spectrum, mass accuracy will be worse. Include a calibrant spot or standard in the target layout for every batch where accurate masses are required.

Technical data leaflet MO302
Maintenance
Vacuum pumps require rotary backing and periodic service

The sample-introduction and analyzer turbomolecular pumps (250 l/s nominal) rely on rotary backing pumps; loss of backing vacuum or overdue pump service will degrade sensitivity and can trip the system. Schedule rotary-pump oil checks and turbopump service per the manual.

Technical data leaflet MO302
Cooling & Media
Nitrogen laser consumes N2 gas and has a finite shot life

The fixed-focus 337 nm nitrogen laser operates at 50 Hz with 3 ns pulses and 100 µJ nominal energy; it requires a nitrogen supply and its gas fill has a limited lifetime, so plan laser service into long-term operating budgets.

Technical data leaflet MO302

Selection & Integration Q&A

When should I choose the Axima Confidence rather than another variant in the series?

No sibling variants are defined in the supplied facts, so the Axima Confidence is the only configured model here. Choose it when you need both linear mode to 500 kDa and reflectron resolution >15,000 FWHM with a fixed 337 nm N2 laser at 50 Hz.

What target formats and thicknesses are compatible with the sample introduction system?

Standard 96- or 384-well microtitre-format targets (2 mm thick) and FlexiMass microscope-slide-format targets (plain or 48-well) fit directly; the optional adaptor extends the maximum target thickness to 10 mm for unconventional layouts.

What maintenance and consumables are required beyond the base delivery?

The nitrogen laser requires a nitrogen gas supply and periodic laser service due to finite gas-fill life; rotary backing pumps need oil checks and turbopumps need periodic service. Targets and the optional thick-target adaptor are ordered separately, and no cables or accessories are listed as included.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
mo305
Datasheet ·906,6 КБ ·4 pp.
View PDF
PDF
MO302
Datasheet ·455,8 КБ ·2 pp.
View PDF