Shimadzu Ispect DIA 10
Shimadzu Ispect DIA 10 — Particle size analyzers.
The Ispect DIA 10 is a dynamic image analysis particle size analyzer from Shimadzu, configured with a syringe pump and a flow-through measurement cell. It measures particles from 5 to 100 µm using area circle equivalent diameter and other shape parameters, with a required sample volume of 50 to 1000 µL. This model is intended for laboratories needing image-based particle characterization with statistical outputs such as CV and median.
Direct drop-in replacements
Full specifications
Description
Construction and positioning
- The Ispect DIA 10 is a benchtop dynamic image analysis system comprising a measurement unit and a separate pump unit. The measurement unit houses the optics and flow cell, while the syringe pump delivers sample at a fixed flow rate of 0.1 mL/min. - Wetted parts in the measurement unit are PEEK resin, fluorine resin, quartz, and fluorine rubber; the pump unit uses fluorine resin and glass. This material selection supports compatibility with a range of aqueous and chemical samples. - The instrument measures 223 mm in width, 205 mm in height, and 465 mm in depth, and requires a 115 or 230 V AC, 50/60 Hz supply at 100 VA. ### Measurement behavior
- Particles are imaged in flow, and size is reported as area circle equivalent diameter, with additional parameters including perimeter equivalent diameter, maximum length, maximum perpendicular length, Feret diameters, and particle perimeter. - The specified range is 5 to 100 µm. Number concentration reproducibility is expressed as a coefficient of variation (CV) of 5% or better. - Statistical analysis includes average, standard deviation, CV, median (50% value), mode, and user-defined percentiles. Display options include particle images, histograms, scattergrams, cumulative distributions, and frequency/integration tables. ### Installation and operation
- The instrument requires a sample volume of 50 to 1000 µL per measurement. The syringe pump flow rate is fixed at 0.1 mL/min, so measurement time depends on the sample volume and tubing dead volume.
Integration & Operating Notes
Operating outside this range may prevent proper flow through the measurement cell or cause particle overlap, leading to invalid size distributions. Ensure the syringe pump is loaded with a volume that accounts for tubing dead volume and replicates.
The measurement unit uses PEEK, fluorine resin, quartz, and fluorine rubber; the pump unit uses fluorine resin and glass. Avoid samples containing solvents or particles that attack these materials, as this can cause leaks or contamination.
Residual sample can dry and clog the narrow flow cell or syringe pump, leading to inaccurate results. Follow the cleaning procedure in the instruction manual, using a compatible solvent, and perform regular maintenance as specified.
Selection & Integration Q&A
The pump unit delivers sample at 0.1 mL/min through tubing to the measurement cell. Ensure tubing is compatible with the wetted materials (fluorine resin, glass) and that connections are leak-free. The sample volume must be 50–1000 µL.