KB (E7) 09 2026

Technical Reference Summary

Covered Models & Document Scope

Covers BINDER KB and KB-UL (E7) Cooling Incubators with compressor technology and RD4 controller, sizes 65, 130, 260, 470, 720 and 1060. Each size has a standard (KB) and a UL variant (KB-UL). Article numbers: KB 65 9020-0471/9120-0471, KB 65-UL 9020-0491/9120-0491, KB 130 9020-0472/9120-0472, KB 130-UL 9020-0500/9120-0500, KB 260 9020-0473/9120-0473, KB 260-UL 9020-0501/9120-0501, KB 470 9020-0474/9120-0474, KB 470-UL 9020-0502/9120-0502, KB 720 9020-0475/9120-0475, KB 720-UL 9020-0503/9120-0503, KB 1060 9020-0476/9120-0476, KB 1060-UL 9020-0504/9120-0504. Options include analog temperature outputs, zero-voltage alarm contacts, door lock, internal LED light, flexible Pt 100 object temperature sensor, APT-COM 4 software, racks with U-rails or telescopic rails, and a flexible tilt protection kit for KB/KB-UL 65, 130 and 470.

Section-by-Section Technical Breakdown

pp. 18-21
Chamber description and connection panel

Describes the chamber overview, the connection panel on the chamber rear and the triangular instrument panel. The chambers are equipped with a temperature safety device and interfaces for computer communication (Ethernet) and USB; APT-COM 4 software permits networking up to 100 chambers.

  • Connection panel is located on the chamber rear
  • Triangular instrument panel houses the RD4 controller
  • Ethernet and USB interfaces for computer communication
  • APT-COM 4 Multi Management Software can network up to 100 chambers
pp. 24-27
Installation and connections

Covers location and ambient conditions, wall spacer from size 260, mounting of the flexible tilt protection kit for KB/KB-UL 65, 130 and 470, the condensate collection pan, rack installation and electrical connection. Minimum distances and ventilation requirements must be observed.

  • Spacer for wall distance required from size 260 on
  • Flexible tilt protection kit mounted on KB/KB-UL 65, 130 and 470
  • Condensate collection pan must be installed
  • Racks available with U-rails or telescopic rails (accessory)
  • Electrical connection per type plate
pp. 28-30
Functional overview of the RD4 chamber controller

Explains the menu structure and access levels of the RD4 controller and its performance during and after power failures. This is the basis for set-point entry and all special controller functions.

  • Menu structure with access levels
  • Defined behavior during and after power failures
  • Controller display with text field for messages
pp. 30-31
Start up and set-point entry

Describes initial start up and entry of the temperature set-point and fan speed set-point. These are the basic operating settings of the chamber.

  • Temperature set-point entry
  • Fan speed set-point entry
pp. 32-37
Special controller functions and automatic setpoint switchover

Covers special controller functions and automatic setpoint switchover controlled by date or by regular cyclic control, including entry of the alternative temperature set-point and activation/deactivation of the functions.

  • Automatic setpoint switchover by date
  • Regular cyclic setpoint switchover
  • Entry of alternative temperature set-point
  • Activating/deactivating special controller functions
pp. 38-40
Hot-air disinfection

Describes preparations, starting, the process itself, completion and premature termination of hot-air disinfection.

  • Preparations before hot-air disinfection
  • Starting and running the hot-air disinfection process
  • Completing or prematurely terminating hot-air disinfection
pp. 40-42
Password

Covers the password request and assignment/modification of the User password and Admin password.

  • Password request on the controller
  • Assign and modify User password
  • Assign and modify Admin password
pp. 42-47
Temperature safety devices

Describes the over temperature protective device (class 1) and the safety controller class 2 / class 3.3, including selection between temperature limiter and temperature protection, safety controller mode, limit and offset values, class 3.2 undertemperature protection, alarm state measures and function check.

  • Over temperature protective device class 1
  • Safety controller class 2 (temperature limiter) or class 3.3 (temperature protection)
  • Setting safety controller mode, limit value and offset value
  • Safety controller class 3.2 undertemperature protection
  • Message and measures in the state of alarm
  • Function check of the safety devices
pp. 48-51
General controller settings

Covers menu language, temperature unit, date and time, language selection at restart, chamber address and display brightness.

  • Selecting menu language and temperature unit
  • Setting current date and time
  • Language selection at restart
  • Setting the chamber address
  • Display brightness
pp. 51-52
Tolerance range settings

Covers the delay time for the tolerance range alarm and the temperature tolerance range settings.

  • Setting delay time for tolerance range alarm
  • Setting temperature tolerance range
pp. 52-54
Notification and alarm functions

Describes information messages, alarm messages and activation/deactivation of the audible alarm (alarm buzzer).

  • Information messages
  • Alarm messages
  • Activating/deactivating the audible alarm buzzer
pp. 55-56
Program start / stop function

Covers the start time program, start week program, stopping the program and performance after power failures.

  • Start time program
  • Start week program
  • Stopping the program
  • Performance after power failures
pp. 56-61
Ethernet network settings

Covers showing and changing network settings: MAC address, IP address, subnet mask, standard gateway and DNS server address, plus selection of automatic or manual IP/DNS assignment.

  • Showing MAC address, IP address, subnet mask, standard gateway and DNS server address
  • Selecting automatic or manual IP address assignment
  • Selecting automatic or manual DNS server address assignment
  • Assigning IP address, subnet mask, standard gateway and DNS server address
pp. 62-63
Data recorder

Describes recorded data, storage capacity, setting the storage rate for the DL1 recorder data and deleting the data recorder.

  • Recorded data and storage capacity
  • Setting storage rate for DL1 recorder data
  • Deleting the data recorder
pp. 63-66
USB menu: data transfer via USB interface

Covers connecting a USB stick, import/export functions, ongoing data transfer, errors during data transmission and removing the USB stick.

  • Connecting the USB stick
  • Import/export functions
  • Ongoing data transfer
  • Error during data transmission
  • Removing the USB stick
pp. 66-67
Self-test function

Describes activating and deactivating the self-test function of the chamber.

  • Activating the self-test function
  • Deactivating the self-test mode
pp. 68
Defrosting at refrigerating operation and anti-condensation protection

Covers defrosting during refrigerating operation and the anti-condensation protection function.

  • Defrosting at refrigerating operation
  • Anti-condensation protection
pp. 69-73
Options and accessories

Describes APT-COM 4 Multi Management Software, analog outputs for temperature, zero-voltage alarm contacts for collective alarm output, door lock, internal LED light and object temperature control with flexible Pt 100 temperature sensor including sensitivity and maximum deviation settings.

  • APT-COM 4 Multi Management Software (accessory)
  • Analog outputs for temperature (option)
  • Zero-voltage alarm contacts for collective alarm output (option)
  • Door lock (option)
  • Internal LED light (option)
  • Object temperature control with flexible Pt 100 sensor (option)
  • Setting sensitivity and maximum deviation of object temperature control
pp. 74-76
Cleaning and decontamination

Covers cleaning, decontamination/chemical disinfection and hot-air disinfection procedures.

  • Cleaning
  • Decontamination / chemical disinfection
  • Hot-air disinfection
pp. 77-80
Maintenance and service, troubleshooting, repair, testing

Covers personnel qualification, maintenance intervals and service, the service reminder, simple troubleshooting, and returning the chamber to BINDER GmbH. Annual maintenance and annual electrical repeat inspection are required.

  • Maintenance at least once a year
  • Annual electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702
  • Service reminder shown on controller after one year of operation
  • Simple troubleshooting by operating personnel without disassembly
  • Detailed troubleshooting and repair by BINDER Service or qualified partners
  • Chamber must be tested before resuming operation after maintenance or repair
pp. 84-94
Technical description

Covers factory calibration and adjustment, over current protection, definition of usable volume, KB/KB-UL technical data, equipment and options, accessories and spare parts, and dimensions.

  • Factory calibration and adjustment
  • Over current protection
  • Definition of usable volume
  • KB / KB-UL technical data
  • Equipment and options (extract)
  • Accessories and spare parts (extract)
  • Dimensions

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Nominal temperature of the chambers is 100 °C / 212 °F. p.10 ↗ Maximum rated operating temperature for selection and application planning. KB / KB-UL series
IP protection is IP 20 according to EN 60529. p.10 ↗ Chamber must be installed splash-proof and must not become wet during operation, cleaning or maintenance. KB / KB-UL series
Temperature safety device according to DIN 12880:2007, class 2/3.3 (adjustable). p.10 ↗ Safety controller can be configured as class 2 temperature limiter or class 3.3 temperature protection. KB / KB-UL series
KB 260 (example) nominal power 1.30 kW, nominal current 5.4 A at 220-240 V, 50/60 Hz, 1 N~. p.10 ↗ Electrical supply and fusing must be sized for the nominal current of the specific model. KB 260 (9020-0473)
KB 260-UL (example) nominal power 1.30 kW, nominal current 10.8 A at 120 V, 50/60 Hz, 1 N~. p.10 ↗ UL variant requires a 120 V supply sized for 10.8 A. KB 260-UL (9020-0501)
Maximum permissible operating overpressure in the refrigerating system is 11 bar. p.10 ↗ Refrigerant circuit must not be exposed to higher pressure; relevant for service work. KB / KB-UL series
Refrigerant R600A with filling weight 0.095 kg (example KB 260); contains hydrocarbon gases. p.10 ↗ Flammable refrigerant; service work on the refrigeration system only by personnel trained per EN 13313:2010. KB / KB-UL series
Intended temperature application range 4 °C / 39.2 °F to 37 °C / 98.6 °F for cultivation of microorganisms. p.1 ↗ Typical applications: long-term storage at 4 °C, refrigerated incubation 20-25 °C, incubation at 37 °C, alternating temperatures. KB / KB-UL series
Maintenance work at the refrigeration system only by qualified personnel trained according to EN 13313:2010 (e.g. refrigeration technician per Regulation (EC) no. 303/2008). p.25 ↗ Service personnel qualification must be verified before any refrigeration work. KB / KB-UL series
Regular maintenance work at least once a year; all work on the refrigeration system must be documented. p.25 ↗ Plan annual maintenance and keep documentation. KB / KB-UL series
Annual repeat electrical inspection and test before initial startup and before resuming operation after maintenance or repair, recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702. p.25 ↗ Electrical safety testing is mandatory before first use and after service. KB / KB-UL series
With increased dust in ambient air, clean the condenser of the refrigeration machine by suction several times a year; condenser is behind the tool-free removable machine room cover. p.25 ↗ Dusty environments require more frequent condenser cleaning to maintain performance. KB / KB-UL series
Replace the door gasket only when cold. p.25 ↗ Replacing a warm gasket may damage it. KB / KB-UL series
Service reminder: keep OK button pressed for 5 seconds to display remaining days until maintenance is due; after one year of operation the message 'Service due!' appears and reappears after one week if not confirmed. p.25 ↗ Controller provides a maintenance countdown and recurring due message. KB / KB-UL series
APT-COM 4 Multi Management Software permits networking up to 100 chambers and connecting them to a PC. p.2 ↗ Network planning must account for up to 100 chambers per APT-COM 4 installation. KB / KB-UL series
Spacer for wall distance required from size 260 on. p.4 ↗ Sizes 260 and larger need the wall spacer installed to maintain ventilation clearances. KB / KB-UL 260, 470, 720, 1060
Flexible tilt protection kit is mounted on KB/KB-UL 65, 130 and 470. p.4 ↗ These sizes require the tilt protection kit during installation. KB/KB-UL 65, 130, 470
Chamber must not be installed in unventilated recesses; sufficient ventilation for heat dispersal and prescribed minimum distances must be observed. p.1 ↗ Prevents overheating and chamber damage. KB / KB-UL series
Do not remove the rear panel of the chamber; disconnect power before maintenance by turning off the On/Off switch (H) and pulling the power plug. p.25 ↗ Live maintenance work is prohibited; rear panel removal is not permitted. KB / KB-UL series
Loading material must not contain corrosive ingredients such as acids and halides that may damage stainless steel, aluminum and copper components. p.1 ↗ Corrosive damage from such ingredients is excluded from BINDER liability. KB / KB-UL series

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Service due! p.78 ↗ Recommended maintenance interval (one year of operation) has elapsed. Press the OK button to confirm the message. The message reappears after one week of operation. chap. 25.3