Operating Instructions — VSI 300 OXY vacuum pump (BA 832-43-EN, EN)

Technical Reference Summary

Covered Models & Document Scope

This document (BA 832-43-EN, edition 1.4.2016) covers the screw vacuum pump S-VSI 300 (43) OXY, a blue-varnished OXY variant for handling gases with raised oxygen share (>21 vol%) and other oxidants. It describes the standard version with continuous flow cooling and the options: control unit, thermostatic water valve (fresh water cooling control), and circulation cooling system. Motor variants referenced are 5.5 kW and 7.5 kW.

Section-by-Section Technical Breakdown

pp. 12-14
Setup, Description and Data Plate

Describes the mechanical layout and operating principle of the S-VSI 300 (43) OXY double-shaft screw vacuum pump, including component identification figures and the data plate contents. The pump uses two parallel non-contacting screw rotors synchronised by an oil-lubricated gear pair; the gearbox and compression chamber are separated by special seals with an atmospherically ventilated intermediate area. A gas ballast valve is fitted as standard.

  • Double-shaft screw vacuum pump with dry, non-contacting rotors
  • Gearbox and compression chamber separated by special seals; intermediate area can be loaded with sealing gas (special version)
  • Gas ballast valve fitted as standard to reduce water vapour condensation
  • OXY pumps varnished blue for differentiation from standard versions
  • Driven by standard flanged three-phase motor via elastomer coupling
  • Data plate shows type/size, serial number, year, item no., suction capacity, ultimate vacuum, speed, motor output, operating mode
pp. 15
Areas of Application and Operating Limits

Defines the intake pressure range, maximum pumping capacity and special operating restrictions for the OXY variant. The pump is suitable for closed-system evacuation or continuous vacuum between 0.1 and 1000 mbar (abs.) and for very damp gases. Specific limits apply to the 5.5 kW motor version and to evacuation of closed systems.

  • Intake pressure range 0.1 to 1000 mbar (abs.)
  • Maximum pumping capacity with free suction 320 m3/h at 50 Hz
  • 5.5 kW motor version only for continuous operation S1
  • Closed-system evacuation volume max. 80 l
  • Frequent switching (>10 times/hour) or high ambient/intake temperatures may exceed motor winding and bearing temperature limits
  • Outdoor installation requires protection from environmental influences (e.g. protective roof)
pp. 16-17
Cooling the Machine

Covers the standard continuous flow cooling through the double-walled compressor housing and the optional control unit, thermostatic water valve and circulation cooling system. Safety devices include solenoid valve, temperature control and safety valve. Factory settings and control voltages are specified.

  • Standard continuous flow cooling through cavity in double-walled compressor housing
  • Temperature control factory setting Tmax = 60 °C
  • Safety valve protects against cooling water pressure > 6 bar
  • Solenoid valve control voltage 24 V DC
  • Thermostatic water valve (option) factory setting Tmax = 55 °C
  • Circulation cooling option uses thermostat-controlled three-way valves, temperature switch and flow switch; weight 101/131 kg, tank capacity 30 l, dimensions 580 x 580 x 650 mm
pp. 18-23
Installation

Provides requirements for preparing the installation site, mounting, connecting suction and discharge pipes, connecting cooling water, filling lubricating oil and connecting the motor. Includes clearance requirements, back-pressure limits, cooling water quality and electrical connection data.

  • Cooling air inlets/outlets must be at least 30 cm from adjacent walls
  • Machine must be set up horizontally on a level, straight foundation able to take its weight
  • Suction pipe must not be too narrow or too long; only screw pipes in by hand
  • Exhaust air outlet must not be closed or restricted; prevent liquid accumulation in exhaust line
  • Cooling water: pH-neutral, clean, filtered; operating pressure max. 6 bar; temperature 15–50 °C
  • Anti-freeze at least 10 % to prevent frost damage
  • Lubricating oil OXY-GEAR-LUBE 160 filled to middle of sight glasses
  • Motor protection class IP 55, insulation class F; electrical installation per EN 60204
  • Permissible without derating: ±5 % voltage deviation, ±2 % frequency deviation
pp. 24-27
Commissioning and Decommissioning

Describes commissioning checks including rotation direction verification and post-run procedure, plus decommissioning, storage and re-commissioning steps. Emphasises the exclusive use of OXY-GEAR-LUBE 160 and the explosion hazard from mixing organic substances with oxygen.

  • Check rotation direction by brief start (max. 2 s); motor fan must rotate clockwise
  • Use phase sequence indicator to check for anti-clockwise rotating field
  • Post-run at 50–100 mbar (abs.) with dry air for minimum 10 minutes before non-operating periods >2 hours
  • Decommissioning: switch off, close cut-off devices, disconnect power, depressurise, remove pipes, seal connections, drain cooling water and condensate
  • Only OXY-GEAR-LUBE 160 may be used; other lubricants or greases prohibited (explosion danger)
  • If oil is suspected of mixing with organic substances, change oil immediately and have pump cleaned by authorised Elmo Rietschle personnel
pp. 28-34
Maintenance and Repair

Defines maintenance intervals and procedures for oil changes, air filtering, coupling inspection and cooling system maintenance. Specifies the required lubricant, cleaning agents and safety precautions including prohibition of oily compressed air.

  • Monthly: check pipes/screws for leaks, terminal box and cable inlets, clean cooling ribs, check oil level, check cooling water system
  • Oil change every 7,500 operating hours with OXY-GEAR-LUBE 160
  • Clean intake air mesh filter and gas ballast valve filter depending on contamination
  • Coupling sprocket checked at least once a year; wear reduces with frequent starting and high ambient temperature
  • Clean dirt trap in cooling circuit depending on cooling liquid quality
  • Only washing agent Antikor 30-57 with wetting agent 200-4 (Kluthe) permitted for filter cleaning
  • Do not use oily compressed air for cleaning (explosion danger)
pp. 36
Spare Parts

Specifies how to order spare parts using the spare parts list E 832/6 for S-VSI 300 (43) OXY and notes that wear parts and gaskets are listed separately. Only original or manufacturer-approved parts may be used.

  • Spare parts list E 832/6 for S-VSI 300 (43) OXY
  • Download from manufacturer website under Downloads > Product Documents > S-Series > Spare Parts
  • Wear parts and gaskets indicated separately
  • Only original spare parts or manufacturer-approved parts permitted

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Ambient and suction temperature must be between 5 and 40 °C. p.7 ↗ Operation outside this range requires consultation with the manufacturer; selection must account for site temperature. S-VSI 300 (43) OXY
Back pressure on outlet side: S-VSI (43) 7.5 kW < +0.2 bar; S-VSI (43) 5.5 kW < +30 mbar. p.7 ↗ Discharge piping and silencer must be sized so that these limits are not exceeded. S-VSI 300 (43) OXY motor variants
Intake pressure range for continuous vacuum: 0.1 to 1000 mbar (abs.). p.15 ↗ Pump is suitable for closed-system evacuation and continuous vacuum within this range. VSI OXY
Maximum pumping capacity with free suction is 320 m3/h at 50 Hz. p.15 ↗ Use data sheet D 832/43 for pumping capacity dependence on intake pressure. S-VSI 300 (43) OXY
The 5.5 kW motor version may only be operated in continuous operation S1. p.15 ↗ Do not apply intermittent duty to the 5.5 kW variant. S-VSI 300 (43) OXY with 5.5 kW motor
When evacuating closed systems the volume to be evacuated must be max. 80 l. p.15 ↗ Larger closed volumes require alternative evacuation strategy or consultation. S-VSI 300 (43) OXY
Cooling air inlets and outlets must be at least 30 cm away from adjacent walls. p.18 ↗ Installation layout must provide this clearance to avoid recirculation of cooling air. S-VSI 300 (43) OXY
Cooling water operating pressure must not exceed 6 bar; cooling water temperature must be between 15 and 50 °C. p.21 ↗ Cooling water supply must be regulated and monitored within these limits. S-VSI 300 (43) OXY
Cooling water must be pH-neutral, clean and filtered; anti-freeze at least 10 % to prevent frost damage. p.21 ↗ Water quality per info sheet I 832-EN; anti-freeze concentration adjusted to ambient conditions. S-VSI 300 (43) OXY
Specified lubricating oil is OXY-GEAR-LUBE 160; no other lubricants or greases may be used. p.22 ↗ Explosion hazard if non-compliant lubricant is used; oil must be filled to middle of sight glasses. VSI OXY
Motor protection class IP 55, insulation class F; electrical installation per EN 60204. p.23 ↗ Main switch must be provided by operating company; motor protection switch with delayed switch-off recommended. S-VSI 300 (43) OXY
Without derating, permissible voltage deviation ±5 % and frequency deviation ±2 %. p.23 ↗ Mains conditions at installation site must match motor data plate within these tolerances. S-VSI 300 (43) OXY
Rotation direction check: brief start max. 2 seconds; motor fan must rotate clockwise. p.26 ↗ Incorrect rotation direction damages the machine; use phase sequence indicator. S-VSI 300 (43) OXY
Post-run at 50–100 mbar (abs.) with dry air for minimum 10 minutes before non-operating periods >2 hours. p.26 ↗ Removes dampness and impurities to prevent deposits and starting problems. S-VSI 300 (43) OXY
Oil change interval: every 7,500 operating hours with OXY-GEAR-LUBE 160. p.29 ↗ Maintenance planning must include oil changes at this interval; check oil level monthly. S-VSI 300 (43) OXY
Coupling sprocket must be checked at least once a year; frequent starting and high ambient temperature reduce its service life. p.34 ↗ Defective sprocket may lead to rotor shaft breakage. S-VSI 300 (43) OXY
Temperature control factory setting Tmax = 60 °C; thermostatic water valve factory setting Tmax = 55 °C; solenoid valve control voltage 24 V DC. p.16 ↗ Cooling control settings must be verified during commissioning. S-VSI 300 (43) OXY with control unit / thermostatic water valve
Circulation cooling option: weight 101/131 kg, tank capacity 30 l, dimensions 580 x 580 x 650 mm. p.17 ↗ Space and floor loading must accommodate these dimensions and weight. S-VSI 300 (43) OXY with circulation cooling option
Storage conditions: relative humidity 0–80 %, storage temperature -10 to +60 °C, max. 6 months. p.11 ↗ Long-term storage beyond 6 months requires consultation of storage guidelines I 150-EN. S-VSI 300 (43) OXY
Installation at more than 1000 m above sea level causes noticeable reduction in power. p.19 ↗ Contact manufacturer for high-altitude applications. S-VSI 300 (43) OXY

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Machine is switched off by the motor protection switch p.37 ↗ Mains voltage/ Frequency does not correspond with the motor data Check by qualified electrician Section 5.5
Machine is switched off by the motor protection switch p.37 ↗ Connection to motor terminal board is not correct Check by qualified electrician Section 5.5
Machine is switched off by the motor protection switch p.37 ↗ Motor protection switch is not set correctly Check by qualified electrician Section 5.5
Machine is switched off by the motor protection switch p.37 ↗ Motor protection switch is triggered too quickly Use a motor protection switch with an overload-dependent delayed switch off that takes into consideration the short term excess current at start up (version with short circuit and overload trigger as per VDE 0660 Part 2 or IEC 947-4)
Pumping capacity is insufficient p.37 ↗ The mesh filter is dirty Clean or replace the mesh filter Section 7.2.2, Section 7.4
Pumping capacity is insufficient p.37 ↗ The suction pipe is too long or too narrow Check the hose or the pipe Section 5.3
Pumping capacity is insufficient p.37 ↗ Machine or system leaking Check the pipework and screw connections for leaks and to ensure that they are firmly seated Section 7.2
Final pressure (max. vacuum) is not reached p.38 ↗ Machine or system leaking Check the pipework and screw connections for leaks and to ensure that they are firmly seated Section 7.2
Final pressure (max. vacuum) is not reached p.38 ↗ Too little cooling water Note cooling water consumption Section 9
Final pressure (max. vacuum) is not reached p.38 ↗ The mesh filter is dirty Clean or replace the mesh filter Section 7.2.2, Section 7.4
Machine gets too hot p.38 ↗ Ambient or intake temperature is too high Ensure it is being used properly Section 2.3
Machine gets too hot p.38 ↗ Cooling air supply is obstructed Check environmental conditions Section 5.1
Machine gets too hot p.38 ↗ Cooling air supply is obstructed Clean the cooling ribs Section 7.2
Machine gets too hot p.38 ↗ The cooling water system is obstructed Check the cooling water system and the pipes Section 7.2
Machine gets too hot p.38 ↗ Too little cooling water Note cooling water consumption Section 9
Machine gets too hot p.38 ↗ Cooling water inflow is too hot Note the max. intake temperature Section 9
The machine makes a abnormal noise p.38 ↗ Deposits on the rotors Clean the working space and the rotors Elmo Rietschle Service