Elmo Rietschle VSI 100
Elmo Rietschle ·Made in Germany · In Active Production

Elmo Rietschle VSI 100

Elmo Rietschle VSI 100 — Screw Vacuum Pumps with Motor power 3.6 kW, Pumping speed 138 m³/h, Ultimate pressure 0.1 mbar.

The VSI 100 is a dry-running, non-contact screw vacuum pump with oil-lubricated gearbox and bearings, separating the compression chamber from the drive. It delivers 110 m³/h at 50 Hz (138 m³/h at 60 Hz) and an ultimate pressure of 0.1 mbar, positioning it as the smallest model in the VSI series, distinct from the higher-capacity VSI 300 and VSI 301. It is intended for continuous evacuation or vacuum holding in closed systems up to 80 liters, with water cooling as standard.

Motor power
3.6 kW
Pumping speed
138 m³/h
Ultimate pressure
0.1 mbar

Elmo Rietschle VSI Series — Model Comparison & Selection

Model Motor power, kWPumping speed, m³/hUltimate pressure, mbarHeight, mmLength, mmNoise level, dB
VSI 100 Current 31380.1404108275
VSI 300 5.53200.1580129774
VSI 301 93000.08124983975

Full specifications

Core Performance & Functional Specifications
Motor power 3.0 kW (50 Hz)
60 Hz: 3.6 kW
Pumping speed 110 m³/h (50 Hz)
60 Hz: 138 m³/h
Ultimate pressure 0.1 mbar
Rotational speed 2880 rpm (50 Hz)
60 Hz: 3550 rpm
Dimensions, Weight & Mechanical Design
Height 404 mm
Length 1082 mm
Weight 190 kg
Width 543 mm
Interfaces, Connections & Control
Inlet flange G 1 1/2
Materials, Media & Process Compatibility
Oil capacity 1.65 l
Operating Conditions, Environment & Protection
Noise level 72 dB (50 Hz)
60 Hz: 75 dB

Description

Construction and series positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Process Limits
Closed-system volume must not exceed 80 liters

When evacuating a closed system, the volume to be evacuated must be max. 80 l. Larger volumes require a different evacuation strategy or consultation with the manufacturer, as exceeding this limit may lead to overheating or inadequate evacuation performance.

Operating Instructions BA 832-EN, Section 4.3 / p.15
Process Limits
Exhaust counter pressure must stay below +0.2 bar

The permissible outlet counter pressure for the S-VSI 100 (01) is < +0.2 bar. Exhaust piping, silencers, and any downstream components must be sized so that this limit is not exceeded during operation, otherwise pumping capacity and motor load may be adversely affected.

Operating Instructions BA 832-EN, Section 2.4 / p.7 and Section 5.3 / p.19
Cooling & Media
Cooling water pressure must not exceed 6 bar

The cooling water operating pressure must not exceed 6 bar; if supply pressure can exceed this value, install a pressure regulation valve or safety valve. Cooling water temperature must be between 15 and 50 °C, and consumption is 1.8 l/min at a maximum cooling-water temperature of 50 °C. Mix with at least 10% anti-freeze where freezing conditions can occur.

Operating Instructions BA 832-EN, Section 5.4 / p.20-21
Mounting
Ensure adequate clearance for cooling and service access

Provide sufficient clearance around the pump for cooling-water connections and maintenance access. When installed at more than 1000 m above sea level, a reduction in power is noticeable; contact the manufacturer for high-altitude applications.

Operating Instructions BA 832-EN, Section 5.1 / p.18 and Section 5.2 / p.18
Electrical
Verify supply voltage and frequency tolerances before connection

Without derating, ±5% voltage deviation and ±2% frequency deviation are permissible. Verify that the mains supply matches the motor data plate before connection. The motor is available for 230/400 V ±10% (50 Hz) and 220/380 V (60 Hz) or 380/660 V (60 Hz).

Operating Instructions BA 832-EN, Section 5.6 / p.22
Maintenance
Change oil every 7,500 operating hours and check level monthly

Oil change interval is every 7,500 operating hours with clean operations. Check oil level monthly. Use only ISO VG 150 CLP HC 150 oil (GEAR-LUBE 150 or equivalent); if changing oil type, empty the oil tank completely. Correct oil quantity for S-VSI 100 is 1.65 l (1.0 l at H + 0.65 l at H1); fill to the middle of the inspection glasses.

Operating Instructions BA 832-EN, Section 7.2 / p.26-27 and Section 9 / p.35
Process Limits
Ambient and suction temperature must be 5–40 °C

Operation outside this range requires consultation with the manufacturer. Ensure that the installation location maintains ambient temperature within this range and that the suction gas temperature does not exceed 40 °C.

Operating Instructions BA 832-EN, Section 2.3 / p.7
Mounting
Install non-return valve for long connection pipes

With connection pipes of the same cross-section as the machine connection and more than 3 m long, a non-return valve (ZRK) must be installed. This prevents reverse operation when the machine has stopped.

Operating Instructions BA 832-EN, Section 5.3 / p.19

Selection & Integration Q&A

When should I choose the VSI 100 over the VSI 300 or VSI 301?

Choose the VSI 100 for lower pumping speeds up to 110 m³/h (50 Hz) and smaller systems with G 1 1/2 connections. The VSI 300 offers 320 m³/h and G 2 1/2, while the VSI 301 provides 300 m³/h and a lower ultimate pressure of 0.08 mbar. Select the VSI 100 when its capacity and ultimate pressure of 0.1 mbar suffice and space or cooling water is limited.

What are the requirements for the cooling water supply and connections?

Cooling water operating pressure must not exceed 6 bar; install a pressure regulation valve if necessary. Water temperature must be 15–50 °C, and consumption is 1.8 l/min at a maximum cooling-water temperature of 50 °C. Mix with at least 10% anti-freeze where freezing conditions can occur. The optional circulation cooling uses G 3/4 connections; the standard continuous-flow cooling connection size is not specified in the provided documentation—consult the manufacturer or data sheet D 832 for exact connection dimensions.

What are the mounting clearance and ambient temperature limits?

Provide sufficient clearance around the pump for cooling-water connections and maintenance access. Ambient and suction temperature must be between 5 and 40 °C; operation outside this range requires manufacturer consultation. At altitudes above 1000 m, a reduction in power is noticeable; contact the manufacturer for high-altitude applications.

What maintenance items and accessories are required but not included in the base delivery?

The base delivery does not include oil (ISO VG 150 CLP HC 150, 1.65 l required), cooling water fittings, or a non-return valve (ZRK) for long pipes. Maintenance includes oil changes every 7,500 hours, monthly oil level checks, coupling inspection, and cooling system maintenance. Air filter and gas ballast filter cleaning are not mentioned in the provided engineering overview or notes; consult the operating instructions for their existence and service interval. Spare parts are listed in the Spare Parts List E 832.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Spare Parts List — VSI 100 (E 832, EN)
Spare parts ·367,7 КБ ·2 pp.
View PDF
PDF
Operating Instructions — VSI 100 (BA 832-EN, EN)
Manual ·1,3 МБ ·38 pp.
View PDF
PDF
Technical Data Sheet — VSI 100 vacuum pump (EN)
Datasheet ·195,4 КБ ·2 pp.
View PDF
PDF
Spare Parts List — VSI 100 (E 832, EN)