Protector Echo and Airo Filtered Fume Hoods Manual (2020-Present)

Technical Reference Summary

Covered Models & Document Scope

Covers Labconco Protector® Echo™ and Protector® Airo™ Filtered Fume Hoods, bench-top and floor-mount, in 3, 4, 5, 6 and 8 FT widths. Model/ordering codes listed: Echo bench-top 180410xxx/180510xxx/180610xxx/180810xxx, 181410xxx/181510xxx/181610xxx/181810xxx, 182410xxx/182510xxx/182610xxx/182810xxx; Airo bench-top 184300xxx/184310xxx/184410xxx; Echo floor-mount 183417xxx/183517xxx/183617xxx/183817xxx, 183427xxx/183527xxx/183627xxx/183827xxx, 183437xxx/183537xxx/183637xxx/183837xxx. Options include Neutrodine filters, optional HEPA filter, acid (A) or formaldehyde (F) sensors, service fixtures, duplex outlets, and eGuard command module.

Section-by-Section Technical Breakdown

pp. 10
Location and Clearance Requirements

The hood must be installed on a level, solid surface away from traffic patterns, doors, windows, fans, ventilation registers and other air-handling devices that could disrupt airflow. All room windows should be closed. Overall space requirements are given in Appendix B.

  • Locate away from traffic patterns, doors, windows, fans and ventilation registers.
  • All windows in the room should be closed.
  • Area must be level and of solid construction.
  • See Appendix B: Dimensions for overall product dimensions.
pp. 11
Electrical Requirements

Table 2-1 gives airflow, makeup air, face velocity, number of modules, fan power, light power and total power for each hood width. A dedicated outlet with appropriate circuit breaker must be close to the product and provide a protective earthing ground.

  • 3 FT: up to 130 CFM, 1 module, 36 W fan, 18 W light, 54 W total.
  • 4 FT: up to 260 CFM, 2 modules, 72 W fan, 36 W light, 108 W total.
  • 5 FT: up to 390 CFM, 3 modules, 108 W fan, 54 W light, 162 W total.
  • 6 FT: up to 520 CFM, 4 modules, 144 W fan, 72 W light, 216 W total.
  • 8 FT: up to 650 CFM, 5 modules, 180 W fan, 90 W light, 270 W total.
  • Face velocity 60-100 fpm for all widths.
  • Dedicated outlet must provide protective earthing ground.
pp. 11
Service Line Requirements

Service lines must be ¼-inch outside diameter copper (brass for natural gas) with an easily accessible shut-off valve. Recommended operating pressure is 40 PSI, maximum allowable 200 PSI.

  • ¼-inch outside diameter copper tubing; brass for natural gas.
  • Equipped with easily accessible shut-off valve.
  • Recommended operating pressure 40 PSI.
  • Maximum allowable pressure 200 PSI.
  • Consider a pressure regulator to reduce line pressure to 40 PSI.
pp. 22-28
Installation

Covers unpacking, moving, setting on the work surface, electrical connection, service line connection, sealing to the work surface, filter installation, airflow sensor calibration and certification. The hood must be grounded to MAINS protective earthing; 115V phase is black and neutral white, 230V Phase1 brown and Phase2 blue.

  • Leave hood on shipping skid until close to final location; move with floor jack or furniture dolly, do not tilt onto hand truck.
  • Sash weight is individually matched and must not be exchanged between units.
  • Work surface must support combined weight and be at least as wide as the hood; align flush with back of hood.
  • Seal hood to work surface with silicone sealant.
  • Main power to control panel, fans and lights uses less than 3 amps; main power cord included.
  • Optional 115V harness 9582710 or 230V harness 9582720 for duplex outlets.
  • Ground with ring terminal sized for #10-24 machine screw to protective earthing lug.
  • 115V: Phase (Hot) black, Neutral white; 230V: Phase1 brown, Phase2 blue.
pp. 28-32
Filter Installation

Describes column-based Neutrodine filter and fan box stacking for each hood size. M1 (3 ft Airo) uses 1 column and 2 filters; M2 (4 ft Echo/Airo) 2 columns and 4 filters; M3 (5 ft Echo) 3 columns and 6 filters; M4 (6 ft Echo) 4 columns and 8 filters; M5 (8 ft Echo) 5 columns and 10 filters.

  • M1: 1 column, 2 Neutrodine filters.
  • M2: 2 columns, 4 Neutrodine filters.
  • M3: 3 columns, 6 Neutrodine filters.
  • M4: 4 columns, 8 Neutrodine filters.
  • M5: 5 columns, 10 Neutrodine filters.
  • Primary filter labeled 'P', secondary labeled 'S' with column number.
  • Fan box motor up, Erlab logo upright; locator poles must align with locator holes.
  • Acid sensor fan box labeled 'A', formaldehyde sensor labeled 'F'; place per Appendix G.
pp. 32-34
Calibrating the Airflow Sensor

Airflow sensor calibration is required for safe operation and should be repeated frequently, with annual calibration recommended. The process takes about 5 minutes and must be done with both side panels installed and no wiring or cords blocking the sensor inlet.

  • Calibrate with both side panels installed to prevent cross drafts.
  • Set sash 2 to 3 inches above working height to avoid nuisance alarms.
  • Hold Power Button 10 seconds until maintenance and containment icons appear red.
  • Press Fan ON/OFF to start calibration; do not move sash during calibration.
  • Zero value calculated with fan off (~1 minute), then ~80 fpm (0.4 m/s) value with fan on (~1 minute).
  • Recalibrate if fan set points are changed.
  • After first installation or filter change, clear maintenance alarm by holding mute alarm button 2 seconds.
pp. 34
Certification and Re-Certification

The hood is designed for inflow velocity between 60-100 fpm (0.31-0.51 m/s). Certification must be at a sash position correlating to the required inflow velocity using the organization's Chemical Hygiene Plan; ANSI/AIHA/ASSP Z9.5 and ANSI/ASHRAE 110 are referenced. Recertification is required at least annually.

  • Inflow velocity range 60-100 fpm (0.31-0.51 m/s).
  • Lower inflow velocities by raising sash; higher by lowering sash.
  • Certification includes airflow velocity testing and smoke testing.
  • Recertify at least annually or as directed by Chemical Hygiene Plan.
pp. 35
Performance Features

The hood combines the patented Protector Hood design (U.S. Patent No. 6,461,233) with Neutrodine filtration technology for ductless operation. It provides velocities from 60 to 100 fpm and uses a by-pass feature with Neutrodine universal filter technology.

  • Ductless operation with Neutrodine filtration.
  • Designed for 60-100 fpm face velocity.
  • By-pass feature maintains safe airflow.
  • U.S. Patent No. 6,461,233.
pp. 40, 96
Command Module and eGuard

The command module has a Smart Halo Indicator and mute alarm button. eGuard homepage has INFO, LOGS and SETTINGS sections; INFO shows face velocity, fan, filtration and maintenance states with green/red dots. Alarm screens flash the halo and sound an audible alarm; mute stops sound for ten minutes.

  • INFO shows usage time and safety state of Face Velocity, Fan, Filtration and Maintenance.
  • Gray dot when ventilation off; green when running correctly; red on error.
  • Alarm: sides and front Smart Halo Indicator flash, audible alarm sounds, mute button illuminates.
  • Mute stops audible alarm for ten minutes; alarm resumes if unresolved.
  • SETTINGS requires username and password; sensor sensitivity adjustable.
  • Wi-Fi network GFH-XXXXX-XXXX, password GFH-2018, browser address http://192.168.1.1.
  • Laptop Ethernet: IP 192.168.0.201, subnet 255.255.255.0, browser http://192.168.0.205.
pp. 46-66
Maintenance and Filter Replacement

Covers front panel removal, pre-filter replacement, base frame replacement, Neutrodine filter replacement, fan replacement, optional HEPA filter installation/replacement, airflow sensor replacement and solvent/acid/formaldehyde sensor replacement. Primary filters must be replaced when embedded sensors detect chemicals between primary and secondary filters or after 36 months.

  • Primary filters replaced if embedded sensor (molecode S, A or F) detects chemicals between primary and secondary filters.
  • Primary filters also replaced after 36 months of use.
  • All primary filters must be replaced at once.
  • Old secondary filter becomes new primary filter; new filter installed as secondary.
  • After replacement, clear 'replaced' alarm by pressing mute button.
  • Do not leave hood running while troubleshooting.
pp. 69-75
Accessories

Lists filters, work surfaces, base stands, accessory shelves, seismic supports, hydraulic lift base stands, storage cabinets, service fixture kits (standard, cold water gooseneck, hot/cold mixing, deionized/distilled water), electrical duplex outlet kits, sash stop kit, distillation grid kits and sash reduction wing kits.

  • Filters and work surfaces available.
  • Base stands, accessory shelves, seismic supports, hydraulic lift base stands.
  • Standard service fixture kits; cold water gooseneck; hot and cold water mixing gooseneck; deionized/distilled water gooseneck.
  • Electrical receptacle kits and duplex outlet installation.
  • Sash stop kit, distillation grid kits, sash reduction wing kits.
pp. 77-83
Troubleshooting

Section 10 details common troubleshooting: service valves no longer operate, fans and lights not working, vertical sash no longer operates smoothly, electrical duplex outlets no longer have power, fan operates but lights dim or not working, contaminates outside the filtered hood, lights operate but fans will not, and Smart Command alarms.

  • Service valves no longer operate.
  • Fans and lights not working.
  • Vertical sash no longer operates smoothly.
  • Electrical duplex outlets no longer have power.
  • Fan operates but lights dim or not working.
  • Contaminates outside the filtered hood.
  • Lights operate but fans will not.
  • Smart Command alarms.
pp. 93-94
Specifications and Wiring Diagram

Appendix C provides environmental conditions and the wiring diagram. The main power to control panel, fans and lights uses less than 3 amps; the single point internal junction box is used for duplex outlet connections.

  • Environmental conditions listed in Appendix C.
  • Wiring diagram in Appendix C.
  • Main power to control panel, fans and lights <3 amps.
  • Single point internal junction box for duplex outlets.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Face velocity range is 60-100 fpm (0.31-0.51 m/s) for all hood widths. p.11 ↗ Certify and operate the hood within this inflow velocity range; lower velocity by raising sash, higher by lowering sash. Whole series
3 FT hood: up to 130 CFM, 1 module, 36 W fan, 18 W light, 54 W total. p.11 ↗ Size electrical supply and circuit breaker for at least 54 W total. 3 FT models
4 FT hood: up to 260 CFM, 2 modules, 72 W fan, 36 W light, 108 W total. p.11 ↗ Size electrical supply for at least 108 W total. 4 FT models
5 FT hood: up to 390 CFM, 3 modules, 108 W fan, 54 W light, 162 W total. p.11 ↗ Size electrical supply for at least 162 W total. 5 FT models
6 FT hood: up to 520 CFM, 4 modules, 144 W fan, 72 W light, 216 W total. p.11 ↗ Size electrical supply for at least 216 W total. 6 FT models
8 FT hood: up to 650 CFM, 5 modules, 180 W fan, 90 W light, 270 W total. p.11 ↗ Size electrical supply for at least 270 W total. 8 FT models
Service lines: ¼-inch outside diameter copper (brass for natural gas), recommended operating pressure 40 PSI, maximum allowable 200 PSI. p.11 ↗ Use a pressure regulator to reduce line pressure to 40 PSI; provide easily accessible shut-off valve. Whole series
Main power to control panel, fans and lights uses less than 3 amps; main power cord included for wall outlet. p.26 ↗ A dedicated outlet with appropriate circuit breaker and protective earthing ground is required. Whole series
115V wiring: Phase (Hot) black, Neutral white; 230V wiring: Phase1 brown, Phase2 blue. p.26 ↗ Connect MAINS supply conductors per these color codes; ground only MAINS ground conductor to protective earthing lug. Whole series
Optional 115V harness 9582710 or 230V harness 9582720 for converting duplex outlets to wall outlet attachment. p.26 ↗ Order correct harness for instant attachment of individual duplexes by qualified electrician. Models with duplex outlets
Neutrodine filter count per hood: M1 (3 ft Airo) 2 filters/1 column; M2 (4 ft Echo/Airo) 4 filters/2 columns; M3 (5 ft Echo) 6 filters/3 columns; M4 (6 ft Echo) 8 filters/4 columns; M5 (8 ft Echo) 10 filters/5 columns. p.28 ↗ Order correct number of filters and fan boxes for the hood size. Whole series by width
Primary filters must be replaced if embedded sensor (molecode S, A or F) detects chemicals between primary and secondary filters, or after 36 months of use. p.55 ↗ Plan filter replacement at 36 months or upon filter replacement alarm; replace all primary filters at once. Whole series
Airflow sensor calibration takes about 5 minutes; zero value with fan off (~1 minute), then ~80 fpm (0.4 m/s) value with fan on (~1 minute). p.32 ↗ Perform calibration with both side panels installed, no obstructions at sensor inlet, and sash fixed 2-3 inches above working height. Whole series
After first installation or filter change, maintenance alarm (wrench icon) appears; clear by holding mute alarm button (speaker icon) for two seconds. p.11 ↗ Expected alarm after new equipment recognition; clear manually. Whole series
Heat-generating equipment up to 1058 watts or 300°C surface temperature is tolerated; above this a ducted hood is required. p.17 ↗ Do not use filtered hood if heat exceeds 1058 W or 300°C surface temperature; do not use when internal temperature rises greater than 5°C above ambient. Whole series
eGuard Wi-Fi: network GFH-XXXXX-XXXX (XXXX = serial number), password GFH-2018, browser address http://192.168.1.1. p.96 ↗ Connect to eGuard via Wi-Fi using these credentials. Models with eGuard command module
eGuard laptop Ethernet: IP address 192.168.0.201, subnet mask 255.255.255.0, browser address http://192.168.0.205. p.96 ↗ Configure laptop TCP/IPv4 accordingly for direct Ethernet connection. Models with eGuard command module
Alarm mute stops audible alarm for ten minutes; alarm resumes if error unresolved. p.40 ↗ Mute does not fix problem; investigate root cause via eGuard INFO red dot. Whole series
Sash weight is individually matched to each hood and must not be exchanged with another unit. p.24 ↗ Keep original sash weight with its hood during installation or relocation. Whole series
Certification required before initial use and recertification at least annually or whenever relocated or serviced. p.34 ↗ Use ANSI/ASHRAE 110 test methodology and ANSI/AIHA/ASSP Z9.5 risk assessment if no Chemical Hygiene Plan. Whole series

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Low Face Velocity p.40 ↗ Air being pulled from surroundings into the filtered hood and through the filters is lower than recommended; most commonly seen when sash is fully open. Usually fixed by simply lowering the sash height. Section 10: Troubleshooting
Anemometer out of order p.96 ↗ Airflow sensor malfunction. Replace airflow sensor; see Airflow Sensor Replacement. Appendix D / Section 8
Service valves no longer operate p.77 ↗ Not specified in extracted text. See Section 10 troubleshooting. Section 10
Fans and Lights not working p.78 ↗ Not specified in extracted text. See Section 10 troubleshooting. Section 10
Vertical sash no longer operates smoothly p.79 ↗ Not specified in extracted text. See Section 10 troubleshooting. Section 10
Electrical duplex outlets no longer have power p.80 ↗ Not specified in extracted text. See Section 10 troubleshooting. Section 10
Fan operates but lights dim or not working p.80 ↗ Not specified in extracted text. See Section 10 troubleshooting. Section 10
Contaminates outside the filtered hood p.81 ↗ Not specified in extracted text. See Section 10 troubleshooting. Section 10
Lights operate but fans will not p.82 ↗ Not specified in extracted text. See Section 10 troubleshooting. Section 10
Smart Command Alarms p.83 ↗ Various alarm conditions indicated by flashing Smart Halo Indicator and audible alarm. Press Mute Alarm Button to stop audible alarm; open eGuard homepage INFO to view error code and root cause. Section 6 Alarm Screens / Section 10