Agilent Technologies 572 Tungsten Bayard-Alpert Gauge

Agilent Technologies 572 Tungsten Bayard-Alpert Gauge — Vacuum technologies.

The Agilent 572 is a dual-tungsten-filament Bayard-Alpert ionization gauge tube with a Nonex envelope, measuring nitrogen-equivalent pressure from 1×10⁻³ down to its ~2×10⁻¹⁰ Torr X-ray limit. Its externally switchable filaments and resistance or electron-bombardment degassing distinguish it within the 571/572 series. It serves as the high-vacuum sensor in systems where a hot-filament gauge is acceptable.

Agilent Technologies Vacuum Transducers Series — Model Comparison & Selection

Model Measurement rangeAtomic mass unit, gramsElectron charge, coulombMolar volume, litersMax pressure, megapascalSensitivity
572 Tungsten Bayard-Alpert Gauge Current -3...21.66160222.41-3...2—
531 Thermocouple Gauge, Rough Vacuum Pressure Measurement —1.66160222.41——
571 Bayard-Alpert Gauge Tube -3...21.66160222.411.33×10⁻⁴17
Cold Cathode Gauge Tube 525, High Vacuum (HV) Gauge —1.66160222.41——
Hot Filament Gauge MBA2-201, Ionization BA Gauge —1.66160222.41——
IMG-100 Inverted Magnetron Gauge —1.66160222.41——
IMG-300 UHV Inverted Magnetron Gauge —1.66160222.41——
UHV-24/UHV-24p 11.66160222.41—25

Full specifications

Core Performance & Functional Specifications
Max pressure -3...2 Torr
Measurement range -3...2 Torr
Electron Charge 1.602 coulomb
Internal Volume 220 cc
Maximum operating pressure 1 Torr
Molar Volume 22.41 liters
Pumping Speed, Ionic 0.06 liters/sec
X-ray Limit 2 Torr
Acceleration due to gravity mfr 9.8066
Angstrom unit mfr 10 m
Avogadro's number mfr 6.02214076
Boltzmann's constant mfr 1.38064852
Equivalent of heat mfr 4.185
Natural log base mfr 2.7183
Velocity of light mfr 2.9979
Velocity of sound mfr 343
Magnetic flux density mfr 1
Pumping speed conversion mfr 1 L/s
Throughput conversion mfr 1 cc/sec
Outgassing rate mfr 2
Envelope mfr Nonex Nonex
Collector mfr 0.010 diameter
Base leads mfr 0.060 diameter
Collector lead mfr 0.040 diameter
Shield mfr Internally connected to filament filament
Tubulation mfr 1.000 Nonex
Gauge constant S mfr 1.5
Tubulation (572-N) mfr 1.000 Nonex
Gauge constant S (Helium) mfr 1.5
Gauge constant S (Nitrogen) mfr 10
Gauge constant S (Air) mfr 10
Base Leads mfr 0.060 diameter
Collector Lead mfr 0.040 diameter
K7360301 mfr 572 tubulation
K7360302 mfr 572 tubulation
K7360303 mfr 572 tubulation
K7360304 mfr 572 tubulation
K7360305 mfr 572
K7360306 mfr 572
K7360307 mfr 572 F
Helium mfr 1.5 Torr
Nitrogen mfr 10 Torr
Air mfr 10 Torr
Tubulation OD mfr 1.000
Dimensions, Weight & Mechanical Design
Atomic Mass Unit 1.6605 grams
Shipping Weight 3 lbs.
Electrical, Power & Drive Specifications
Emission Setting For 0.01 Amp/torr Sensitivity 6.7 mA
Filament Current, Absolute Maximum 6.0 Amps
Grid mfr 180 VDC
Grid voltage mfr 180 VDC
Filament voltage mfr 4.0 VAC
Grid voltage (resistance heating) mfr 6.3 VAC
Grid current (resistance heating) mfr 8.7 Amps
Grid to filament voltage (electron bombardment) mfr 700 VDC
Grid current (electron bombardment) mfr 100 DC
Electron bombardment power, absolute maximum mfr 50 W
Grid voltage (resistance heating degas) mfr 6.3 VAC
Grid current (resistance heating degas) mfr 8.7 Amps
Grid to filament voltage (electron bombardment degas) mfr 700 VDC
Grid current (electron bombardment degas) mfr 100 DC
Electron bombardment power mfr 50 W
Filament Voltage mfr 4.0 VAC
Electron bombardment power absolute maximum mfr 50 W
Grid to filament voltage mfr 700 VDC
Grid current mfr 100 DC
Materials, Media & Process Compatibility
Filament Temperature, Absolute Maximum 1400 °C
Shield coating mfr Platinum, internally connected to filament filam
Shield Coating mfr Platinum, internally connected to filament filam
Standards, Safety & General Data
Standard pressure mfr 101325 Pa

Description

Construction and series positioning

Behavior and measurement limits

Electrical operation and degassing

Installation and control

Selection & Integration Q&A

When should I choose the 572 tungsten BA gauge rather than the 571 or a cold-cathode 525?

Pick the 572 for wide-range, linear high-vacuum measurement with dual tungsten filaments that can be switched externally without breaking vacuum; tungsten gives a more stable reading than thoria-coated iridium. Choose the 571 if that tube's configuration fits; choose the 525 cold-cathode gauge where a hot filament is unacceptable.

What must the controller and grounding arrangement provide for the 572?

The controller must supply grid +180 VDC, filament +30 VDC, collector at 0 VDC through the electrometer, and 4.0 VAC / 3.5 A filament with 1 mA grid current, without limiting range or exceeding maxima. Controller and vacuum system must be separately grounded to a common ground capable of ≥10 A; a single interconnecting ground wire is not safe.

Can the 572 be mounted in any orientation, and what environment limits apply?

It may be mounted and operated vertically or horizontally, including during degassing, for prolonged periods. Mount close to the measurement point through a large tubulation (1 in OD recommended) to avoid low-conductance error below 10⁻⁸ Torr. Do not bake above 10⁻⁵ Torr, and do not energize with combustible gases present.

What maintenance and consumables are needed beyond the gauge tube itself?

Plan degassing time: resistance degas typically one-half hour or more in the 10⁻⁹ Torr range, overnight before measurements below 10⁻⁸ Torr. Use powder-free butyl or polycarbonate gloves; Apiezon L grease (Agilent 695400004) if grease is needed. The controller, cable and flange hardware are not included with the tube and must be ordered separately.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Agilent 572 Ionization Gauge User Manual
Manual ·32 pp.
View PDF