No photo available yet

Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies UHV-24/UHV-24p

Agilent Technologies UHV-24/UHV-24p — Vacuum technologies.

The UHV-24/UHV-24p is a nude Bayard-Alpert ionization gauge with dual field-replaceable filaments (tungsten or yttria-coated iridium) on a 2.75 in DN 40 ConFlat flange. The UHV-24 covers 1 mTorr to 2×10⁻¹⁰ Torr (x-ray limit 2×10⁻¹¹ Torr); the UHV-24p extends to 5×10⁻¹¹ Torr (x-ray limit 5×10⁻¹² Torr). It serves as the high-vacuum sensor in UHV systems, requiring an external controller for filament, grid and degas potentials.

Agilent Technologies Vacuum Transducers Series — Model Comparison & Selection

Model Measurement rangeAtomic mass unit, gramsElectron charge, coulombMolar volume, litersMax pressure, megapascalSensitivity
UHV-24/UHV-24p Current 11.66160222.41—25
531 Thermocouple Gauge, Rough Vacuum Pressure Measurement —1.66160222.41——
571 Bayard-Alpert Gauge Tube -3...21.66160222.411.33×10⁻⁴17
572 Tungsten Bayard-Alpert Gauge -3...21.66160222.41-3...2—
Cold Cathode Gauge Tube 525, High Vacuum (HV) Gauge —1.66160222.41——
Hot Filament Gauge MBA2-201, Ionization BA Gauge —1.66160222.41——
IMG-100 Inverted Magnetron Gauge —1.66160222.41——
IMG-300 UHV Inverted Magnetron Gauge —1.66160222.41——

Full specifications

Core Performance & Functional Specifications
Measurement range 1 Torr
Sensitivity 25 Torr
Collector Diameter 0.007 in
Electron Charge 1.602 coulomb
Molar Volume 22.41 liters
Operation 16
X-ray Limit 2 Torr
Acceleration due to gravity mfr 9.8066
Angstrom unit mfr 10 m
Avogadro's number mfr 6.02214076
Boltzmann's constant mfr 1.38064852
Equivalent of heat mfr 4.185
Natural log base mfr 2.7183
Velocity of light mfr 2.9979
Velocity of sound mfr 343
Magnetic flux density mfr 1
Pressure conversion mfr 1 Torr
Pumping speed conversion mfr 1 L/s
Throughput conversion mfr 1 cc/sec
Outgassing rate mfr 2
Pumping speed (nude ion gauge) mfr 0.5 litres/sec
Degas duration mfr 25 minute
Description of the UHV-24/UHV-24p Ionization Gauge mfr 16
Installation mfr 18
Technical Specifications mfr 19
Application Notes mfr 21
Degas mfr 23
Gas Correction Factors mfr 26
Accessories and spare parts mfr 29
X3004-64403 mfr Filament replacement kit kit
UHV-24 pressure range mfr 1 Torr
UHV-24p x-ray limit mfr 5 Torr
UHV-24p sensitivity mfr 20
degas duration mfr 25 minute
nude gauge pumping speed mfr 0.5 litres/sec
Dimensions, Weight & Mechanical Design
Atomic Mass Unit 1.6605 grams
Electrical, Power & Drive Specifications
Emission Current 4 mA
Filament Voltage mfr 3...4.5 current
Filament Current mfr 2.5...3.5 current
Filament voltage mfr 28 VDC
Grid voltage mfr 180 VDC
Degas grid voltage mfr 600 VDC
Degas emission current (total) mfr 67 DC
Degas filament voltage mfr 7 V
Ion current at 10-9 Torr mfr 10 A
Normal operating electron current mfr 4 mA
Degas Grid Voltage mfr 600 VDC
Degas Emission Current (total) mfr 67 DC
Degas Filament Voltage mfr 7 V
Degas Power Requirements (40 W) mfr 600 V
normal operating electron current mfr 4 mA
degas power mfr 40 W
Materials, Media & Process Compatibility
Filament material mfr Tungsten or yttria-coated iridium filaments fila
Filament supports material mfr SST SST
Grid material mfr SST SST
Feedthrough material mfr SST, Alumina, Nickel Alloy Alloy
Filament-to-grid distance mfr 0.050...0.070 in
Filament Supports material mfr SST SST
Maintenance: Filament Replacement mfr 24
Tungsten Filament Replacement mfr 24
Yttria-coated iridium filament Replacement mfr 25
filament-to-grid distance mfr 0.050...0.070
Operating Conditions, Environment & Protection
Bakeout Temperature 450 °C
Standards, Safety & General Data
Standard pressure mfr 101325 Pa

Description

Construction and series positioning

Behavior under load and in different modes

Installation, connection and operation

Integration & Operating Notes

Control & I/O
Switch emission current with pressure range

Use 4 mA normally, 10 mA below 5×10⁻¹⁰ Torr, and ≤0.1 mA above 1×10⁻⁴ Torr. Higher emission also increases gauge pumping, which can bias readings in small chambers.

User Manual, Operation p.17; Application Notes p.21
Process Limits
Cap bakeout at +450 °C and ramp to avoid thermal shock

Repeated or extended bakeout must not exceed +450 °C. Thermal shock can damage the gauge; ramp temperature changes gradually.

User Manual, Bakeout, p.23
Electrical
Ground controller and chamber separately to a common ground

The ground must support at least 10 A. Do not place a ground wire between the vacuum chamber and the controller chassis; improper grounding can expose personnel to 160–900 V.

User Manual, Grounding the Gauge Controller, p.11
Process Limits
Do not operate with combustible gases or mixtures

The gauge is not intrinsically safe; the hot filament and high-voltage discharge can be ignition sources. Do not turn on the gauge while pumping or backfilling combustible gases or mixtures.

User Manual, Use with Combustibles and Mixtures, p.12
Maintenance
Replace filaments with correct spacing and tension

Place the long bracket so the filament sits 0.050–0.070 in from the grid, then push short brackets fully in before tightening. Incorrect spacing or tension changes sensitivity and can damage the filament. Handle only by the ceramic base with nylon gloves.

User Manual, Filament Replacement, pp.24–25
Maintenance
Use correct replacement filament kit for installed type

Yttria-coated iridium kit X3004-64403; UHV-24 tungsten kit 9710018. Each kit contains two filament assemblies and an Allen wrench. Order the kit matching the installed filament type.

User Manual, Accessories and Spare Parts, p.29
Cooling & Media
Choose filament material for gas environment

Yttria-coated iridium runs cooler and outgases less but generates more particles and has shorter life in hydrogen and halogens. Prefer tungsten where hydrogen or halogens are present.

User Manual, Application Notes, p.21

Selection & Integration Q&A

When should I choose the UHV-24p over the UHV-24 or the 571/572 Bayard-Alpert tubes?

Choose UHV-24p when measurements below 2×10⁻¹⁰ Torr are required: its x-ray limit is 5×10⁻¹² Torr and usable range extends to 5×10⁻¹¹ Torr, versus 2×10⁻¹¹ Torr and 2×10⁻¹⁰ Torr for UHV-24. The 571/572 are separate Bayard-Alpert tubes; this nude gauge is for UHV, not rough vacuum.

What controller and cabling does this gauge require for reliable UHV readings?

An external controller must supply filament 3–4.5 V at 2.5–3.5 A, collector 0 V, filament 28 VDC, grid +180 VDC, and a degas mode (40 W, grid up to +600 VDC, 67 mA DC, 7 V filament). Use short PTFE-insulated cables; at 10 pA signals, leakage and noise dominate, so ferrite beads may be needed.

What are the mounting, bakeout and grounding limits for this nude gauge?

Mount in any position in an open location with filaments horizontal; remove the collector protective sleeve first. Maximum bakeout is +450 °C, ramped, no thermal shock. Ground controller and vacuum system separately to a common ground rated ≥10 A; never run a ground wire between chamber and controller chassis.

What maintenance and spare parts are needed, and what is not included in base delivery?

Filaments are field-replaceable: yttria-iridium kit X3004-64403 or UHV-24 tungsten kit 9710018 (each contains two assemblies and an Allen wrench). Degas is factory-performed; post-install degas (~25 min with XGS controllers) is only needed below 1×10⁻⁸ Torr. Controller, cables and filament kits are not included with the gauge.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
UHV-24/UHV-24p Ionization Gauge User Manual
Manual ·36 pp.
View PDF