BINDER BD-S 115
BINDER ·Made in Germany · In Active Production

BINDER BD-S 115

BINDER BD-S 115 — Incubators with Temperature 5…70 °C.

The BD-S 115 is a 118 L natural-convection incubator with a stainless steel 1.4016 (AISI 430) interior and a polyester-coated RAL 9003 housing, controlled by an R-S PID controller with 0.1 °C display resolution. It covers 5–70 °C with 0.3 K fluctuation and 0.5 K uniformity, and is distinguished from the BD-S 56 sibling by its larger chamber volume. It is intended for incubation of cultures at a standard 37 °C set-point.

Temperature
5…70 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER BD-S Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LInterior width, mmLoad per rack, kg
BD-S 115 Current 5–7061.673.511855020
BD-S 56 5–7057.762.56240015

Full specifications

Core Performance & Functional Specifications
Temperature 5...70 °C
Interior volume 118 L
Load per rack 20 kg
Nominal power 0.35 kW
Permitted load 75 kg (BDS115-230V / BDS115UL-120V)
Temperature fluctuation 0.3 K
Temperature uniformity 0.5 K
Back wall clearance 160 mm (BDS115-230V / BDS115UL-120V)
Side wall clearance 100 mm (BDS115-230V / BDS115UL-120V)
Article Number mfr 9090
Recovery time after door was opened for 30 s at 37°C mfr 15 min
Dimensions, Weight & Mechanical Design
Depth 616 mm (BDS115-230V / BDS115UL-120V)
Height 735 mm (BDS115-230V / BDS115UL-120V)
Interior width 550 mm (BDS115-230V / BDS115UL-120V)
Width 765 mm (BDS115-230V / BDS115UL-120V)
Net weight of the unit (empty) mfr 47 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 6.3 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behaviour under load and in operation

Installation, connection and control

Integration & Operating Notes

Electrical
Supply must stay within ±10 % of nominal at the rated frequency

The type plate tolerance is ±10 % at the indicated power frequency (230 V / 50 Hz or 60 Hz). Verify mains quality before installation; out-of-tolerance supply can affect controller operation and heating performance. Size the branch circuit for the 6.3 A chamber fuse on 230 V models.

Type plate, p. 8; Instruction Manual chap. 4.3, p. 19-20
Process Limits
No explosion protection: verify charging material flash point

The chamber has no explosion protection measures. Do not introduce combustible or explosive substances, and confirm that the operating temperature lies below the flash point or sublimation point of the charging material. Acids and halides can also corrode the 1.4016 stainless steel interior.

Instruction Manual chap. 1.9, p. 10-11
Control & I/O
Configure the class 3.1 safety controller before first run

BD-S incubators carry a class 3.1 safety controller per DIN 12880 in addition to the standard class 1 overtemperature device. Set the safety controller set-point and confirm alarm response before loading the chamber; the class differs from ED-S/FD-S ovens (class 2), so do not transfer settings between chamber types.

Instruction Manual chap. 2, p. 14; chap. 9, p. 24
Maintenance
Annual maintenance and repeat electrical inspection required

Plan regular maintenance at least once a year, plus an annual repeat electrical inspection and tests before initial start-up and after maintenance or repair, recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702. Disconnect the power plug before maintenance; do not remove the rear panel; the chamber must not become wet.

Instruction Manual chap. 12.1-12.2, p. 29-30
Maintenance
Replace the door gasket only when cold

Hot gasket replacement may damage the gasket. Allow the chamber to cool before replacing the door gasket, and keep the chamber dry during maintenance to avoid electrical hazard from water ingress.

Instruction Manual chap. 12.2, p. 30
Mounting
Wait one hour after cold storage before start-up

Permissible transport and storage ambient is -10 °C to +60 °C with max. 70 % r.h. non-condensing. After storage in a cold location, wait at least one hour at ambient temperature until the chamber is completely dry before start-up, to prevent condensation-related electrical hazard.

Instruction Manual chap. 3.2-3.3, p. 17

Selection & Integration Q&A

When should I choose the BD-S 115 over the BD-S 56, and when is a forced-convection FD-S the better fit?

Pick the BD-S 115 when the 118 L chamber and 20 kg per-rack load are needed at 5–70 °C with 0.1 °C display resolution and natural convection. The BD-S 56 shares the same temperature range but a smaller chamber. Choose FD-S only if forced-air circulation is required; BD-S is the incubator line for 37 °C culture work.

What supply and plug does the 230 V BD-S 115 need, and what changes for the UL variant?

The 230 V variant (Art.-No. 9090-0022) uses a CEE 7/7 grounded plug, 1N~ current type, and a 6.3 A chamber fuse; supply must stay within ±10 % of nominal at the rated frequency. The UL 120 V variant (Art.-No. 9090-0023) uses a NEMA 5-15P plug with a 12.5 A chamber fuse.

Where can the chamber be installed, and what ambient and duty limits apply?

Do not install in unventilated recesses; observe the prescribed minimum distances and provide sufficient ventilation for heat dispersal. Transport/storage ambient is -10 °C to +60 °C at max. 70 % r.h. non-condensing, and after cold storage wait at least one hour before start-up. Lift at the four lower corners with two people, never by the door.

What maintenance and service schedule applies, and what is not included in base delivery?

Regular maintenance at least once a year, plus annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702, including before initial start-up and after repair. Disconnect the plug before service; replace the door gasket only when cold. Racks and accessories are separate items; the base unit ships without them.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model BD S 115 en
Datasheet ·394,6 КБ ·5 pp.
View PDF
PDF
BD ED FD S (E1) 06 2026 en
Manual ·10,8 МБ ·65 pp.
View PDF