BINDER ED 23
BINDER ED 23 — Drying ovens with Temperature 5…300 °C.
The BINDER ED 23 is a 20 L natural-convection drying and heating chamber with a +5 to 300 °C working range, distinguished within the ED series by its compact 222 × 330 × 495 mm interior and 0.8 kW single-phase heating. It serves small-batch drying, heating, and thermal conditioning where forced air would disturb lightweight or powdery samples.
BINDER ED Series — Model Comparison & Selection
Full specifications
| Temperature | 5...300 °C |
| Air change rate | 13 x/h |
| Heating time to 150°C | 28 min |
| Interior volume | 20 L |
| Load per rack | 12 kg (ED023-230V / ED023UL-120V) |
| Nominal power | 0.8 kW |
| Permitted load | 25 kg (ED023-230V / ED023UL-120V) |
| Temperature fluctuation | 0.5 K |
| Temperature uniformity | 2.8 K |
| Rear wall clearance | 100 mm (ED023-230V / ED023UL-120V) |
| Side wall clearance | 50 mm (ED023-230V / ED023UL-120V) |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 150°C mfr | 5 min |
| Depth | 520 mm (ED023-230V / ED023UL-120V) |
| Height | 495 mm (ED023-230V / ED023UL-120V) |
| Interior height | 330 mm (ED023-230V / ED023UL-120V) |
| Interior width | 222 mm (ED023-230V / ED023UL-120V) |
| Width | 435 mm (ED023-230V / ED023UL-120V) |
| Net weight of the unit (empty) mfr | 27 kg |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 10.0 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- The ED 23 is the smallest standard model in the ED series; siblings ED 56, ED 115, ED 260, and ED 400 share the same +5 to 300 °C range but offer larger interiors. Choose the ED 23 when the 20 L chamber (222 mm W × 330 mm H × 495 mm D) accommodates the workload and when natural convection is preferred over fan-driven airflow.
- Heating is provided by a 0.8 kW single-phase element at 230 V / 50 Hz, protected by a 10.0 A unit fuse. The chamber is delivered as the ED023-230V variant (Art.-No. 9010-0190) with a CEE 7/7 plug; the UL variant ED023UL-120V (Art.-No. 9010-0192) is a separate model.
- The interior is rated for a total permitted load of 25 kg, with a maximum of 12 kg per rack.
Thermal behavior
- Temperature fluctuation is 0.5 K and spatial uniformity is 2.8 K across the working range, supported by a 13 x/h air-change rate. Heating from ambient to 150 °C takes 28 min, and recovery after a 30 s door opening at 150 °C is 5 min.
- Because airflow is by natural convection, the chamber is suited to samples that must not be disturbed by a fan; however, loading patterns should respect the 12 kg per-rack limit and avoid blocking the low-level air path.
Installation and operation
- The unit requires 100 mm rear clearance and 50 mm side clearance to walls or adjacent equipment. Net weight is 27 kg empty; overall external width is 435 mm.
- Electrical connection is to a 230 V, 1-phase, 50 Hz supply; the unit fuse rating is 10.0 A. No external controller or drive is specified in the supplied documentation; operation is via the unit’s own controls.
Integration & Operating Notes
Insufficient clearance restricts natural convection around the chamber and can cause overheating of the outer casing or inaccurate temperature control. Verify the bench or rack can support the 27 kg empty weight plus the 25 kg maximum load before positioning.
Exceeding the per-rack limit can deform shelves and obstruct the natural-convection path, degrading the 2.8 K uniformity. Distribute heavy samples across multiple racks and keep the total below 25 kg to avoid overloading the chamber floor.
After a 30 s door opening at 150 °C, the chamber returns to setpoint in approximately 5 min. Frequent access during critical drying stages will extend cycle time and may affect uniformity; plan loading and sampling to minimize door events.
Selection & Integration Q&A
Pick the ED 23 when the 20 L interior (222 × 330 × 495 mm) fits the batch and natural convection is acceptable. Move to ED 56 or larger only when sample volume or rack count exceeds the 25 kg total / 12 kg per-rack limits.
The standard ED023-230V needs 230 V, 1-phase, 50 Hz with a CEE 7/7 plug and 10 A fusing. A 120 V circuit requires the separate UL variant ED023UL-120V (Art.-No. 9010-0192), which is not a field conversion.
Base delivery is the chamber with its fitted plug and 10 A fuse; racks, additional shelves, and calibration accessories are not listed as included. Confirm rack quantity and any validation service with the order, as the data sheet does not specify them.