BINDER FD 720
BINDER FD 720 — Drying ovens with Temperature 12…300 °C.
The FD 720 is a forced-convection drying and heating oven in BINDER's Avantgarde.Line, with a 741 L interior and a +12 to 300 °C working range. As the largest FD size, it is the only variant with a 400 V 3N~ supply and a fixed 1.8 m power cable. It is intended for tempering, drying and heat treatment of solid, pulverized or bulk materials where airflow and temperature uniformity matter.
BINDER FD Series — Model Comparison & Selection
Full specifications
| Temperature | 12...300 °C |
| Air change rate | 5 x/h |
| Heating time to 150°C | 25 min |
| Interior volume | 741 L |
| Load per rack | 45 kg |
| Nominal power | 4.5 kW |
| Permitted load | 315 kg |
| Temperature fluctuation | 0.6 K |
| Temperature uniformity | 2.5 K |
| Rear wall clearance | 160 mm |
| Side wall clearance | 100 mm |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 150°C mfr | 6 min |
| Depth | 870 mm |
| Height | 1590 mm |
| Width | 1165 mm |
| Net weight of the unit (empty) mfr | 166 kg |
| Power frequency | 50 Hz |
| Rated voltage | 400 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 3 N/PE |
Description
Construction and positioning
- The FD 720 uses forced convection (fan-driven airflow) rather than the natural convection of the ED series, giving active air exchange (5 x/h) and faster recovery after door opening (6 min after 30 s at 150 °C).
- Within the FD series, the 720 is the largest chamber (741 L, 1165 × 1590 × 870 mm net) and the only one requiring a three-phase 400 V 3N~ supply; siblings FD 23, 56, 115 and 260 run on 230 V 1N~ and cover smaller volumes.
- The interior, pre-heating chamber and door inner surfaces are stainless steel 1.4301 (AISI 304); the unit is IP 20 and not classified as a medical device.
Behavior under load and in operation
- Temperature uniformity is 2.5 K and fluctuation 0.6 K; heating from ambient to 150 °C takes 25 min at 4.5 kW nominal power.
- The controller displays temperature with 1 °C resolution (FD series), and the safety controller set-point range is 10–300 °C; set it 2–5 °C above the process set-point.
- On this two-door 720 size, the maximum adjustable temperature set-point depends on ambient temperature, so the achievable top temperature is not fixed at 300 °C under all room conditions.
- Loading is limited to half the usable volume with at least 10 % wall clearance; rack load is 45 kg and total permitted load 315 kg.
Installation, connection and control
- The 720 comes with a fixed power cable of at least 1800 mm and an internal 16 A miniature circuit breaker; other FD sizes use an IEC connector with shockproof plug.
- The chamber must be installed indoors on a non-flammable surface with 160 mm rear and 100 mm side clearance, at 18–40 °C ambient, ≤70 % r.h. non-condensing, and below 2000 m altitude.
- Lifting requires a fork lifter applied only from the rear at the middle of the chamber; do not lift by the door or handle.
- The controller supports temperature set-point entry with ramp and timer functions; fan speed is fixed on FD (adjustable only on BF/FED).
Integration & Operating Notes
The FD 720 draws 4.5 kW on a 400 V 3N~ 50 Hz supply and protects itself with an internal 16 A miniature circuit breaker rather than an external chamber fuse. Verify that the upstream circuit can deliver the nominal current continuously and that supply fluctuations stay within ±10 % of nominal; a fixed 1.8 m cable means the receptacle must be within reach of the installation position.
Maintain at least 160 mm behind and 100 mm at the sides of the chamber. The unit must not be set up in non-ventilated recesses; insufficient clearance prevents heat dissipation and can trip the overtemperature protection. The floor must be non-flammable and level, and the unit must be moved with a fork lifter from the rear only.
On two-door ED, FD and FED 720 chambers the maximum adjustable temperature set-point is not a fixed 300 °C; it depends on the ambient temperature in the room. When planning a process near the upper limit, confirm the achievable set-point for the actual ambient conditions and keep the safety controller set-point 2–5 °C above the process value.
Samples must remain inside the usable volume with at least 10 % of the interior dimensions as clearance to walls, and the usable volume must not be loaded by more than half. Exceeding this prevents sufficient airflow and homogeneous temperature distribution; rack load is limited to 45 kg and total permitted load to 315 kg.
The chamber has an overtemperature protective device class 1 (trip at 330 °C for FD 720) and a safety controller class 2. After a trip, disconnect the chamber, let it cool below the reset threshold or safety controller set-point, inspect if appropriate, reset the alarm on class 2, then restart. Do not simply raise the set-point to clear the alarm.
Maintenance must be performed at least once a year by licensed electricians or experts authorized by BINDER, and an annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended. Replace the door gasket only when the chamber is cold; otherwise the gasket may be damaged. A repeatedly tripping chamber fuse indicates a fault and should be investigated rather than replaced again.
The FD series can be hot-air sterilized at 190 °C for at least 30 minutes, and racks can be removed and autoclaved. All inflammable goods must be removed from the interior before sterilization. Use BINDER neutral cleaning agent (art. no. 1002-0016) and disinfectant spray (art. no. 1002-0022); do not apply neutral cleaner to zinc-coated hinge parts or the rear wall.
Selection & Integration Q&A
Choose the FD 720 when the batch needs the 741 L interior and the 315 kg total load; it is also the only FD size with a 400 V 3N~ supply and a fixed 1.8 m cable. The FD 115 and FD 260 run on 230 V 1N~ and suit smaller loads where a three-phase supply is not available.
It needs 400 V 3N~ 50 Hz, 4.5 kW nominal, with an internal 16 A miniature circuit breaker; supply fluctuation must stay within ±10 %. Because the 720 has no external chamber fuse, size the upstream circuit for the full 4.5 kW and avoid sharing it with loads that cause large voltage dips.
Plan annual maintenance by BINDER-authorized personnel plus annual electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702. Consumables include the door gasket (replace only when cold), neutral cleaning agent art. 1002-0016 and disinfectant spray art. 1002-0022; racks are removable and autoclavable.