BINDER FP 115
BINDER ·Made in Germany · In Active Production

BINDER FP 115

BINDER FP 115 — Drying ovens with Temperature 10…300 °C.

The BINDER FP 115 is a forced-convection drying and heating oven with a 116 L interior and a working range of +10 to 300 °C. It is the mid-size 230 V single-phase model in the FP series, positioned between the FP 56 and FP 260, and delivers 1.4 kW with 32 air changes per hour. It is intended for reproducible drying, heating, and thermal testing of non-explosive, non-corrosive loads in laboratory and industrial processes.

Temperature
10…300 °C

BINDER FP Series — Model Comparison & Selection

Model Temperature, °CAir change rate, 1/hourDepth, mmHeating up time to 150 c, minHeight, cmInterior volume, L
FP 115 Current 10–300326361973.2116
FP 260 10–30097861996.2259
FP 56 10–300805961562.260
FP 720 12–300581625159741

Full specifications

Core Performance & Functional Specifications
Temperature 10...300 °C
Air change rate 32 x/h
Heating time to 150°C 19 min
Interior volume 116 L
Load per rack 30 kg (FP115-230V / FP115UL-120V)
Nominal power 1.4 kW
Permitted load 150 kg (FP115-230V / FP115UL-120V)
Temperature fluctuation 0.3 K
Temperature uniformity 1.4 K
Rear wall clearance 160 mm (FP115-230V / FP115UL-120V)
Side wall clearance 100 mm (FP115-230V / FP115UL-120V)
Article Number mfr 9010
Recovery time after door was opened for 30 s at 150°C mfr 5 min
Dimensions, Weight & Mechanical Design
Depth 636 mm (FP115-230V / FP115UL-120V)
Height 732 mm (FP115-230V / FP115UL-120V)
Width 714 mm (FP115-230V / FP115UL-120V)
Net weight of the unit (empty) mfr 55 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 10.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Thermal behavior and load characteristics

Installation, control, and operation

Integration & Operating Notes

Mounting
Maintain rear and side clearances to prevent overheating

The chamber requires a minimum clearance of 160 mm at the rear and 100 mm at the sides for heat dissipation. Installing it in a recess or closer to walls can cause the unit to overheat, potentially damaging components and affecting temperature control. Ensure the installation location is indoors and well-ventilated.

Operating Manual FP (E3.1), chap. 3.4, p. 22
Electrical
Supply voltage must stay within ±10% of nominal

The FP 115 is rated for 230 V ±10% at 50/60 Hz. Out-of-tolerance supply can cause the controller to malfunction or the heating to underperform, and may compromise safety. Verify the site voltage before connection and use a dedicated circuit rated for at least 10 A (unit fuse 10.0 A).

Operating Manual FP (E3.1), chap. 1.6, p. 9
Process Limits
Load materials must be free of corrosive or explosive substances

The chamber is not explosion-proof. Operating temperature must remain below the flash point or sublimation point of the charge, and no explosive mixture with air may form. Acids, halides, and other corrosive ingredients can damage the stainless steel, aluminum, and copper components; such damage is excluded from warranty.

Operating Manual FP (E3.1), chap. 1.8 and 1.9, pp. 11-12
Control & I/O
External devices on Ethernet/USB must comply with EN 61010-1

Only connect IT equipment that meets EN 61010-1:2010 or EN 60950-1:2006 to the Ethernet and USB interfaces. Non-compliant devices can compromise the electrical safety of the system. Use the zero-voltage alarm contact for external monitoring rather than directly switching mains loads.

Operating Manual FP (E3.1), chap. 1.9, p. 13
Maintenance
Annual maintenance by authorized personnel is mandatory

Schedule maintenance at least once a year; warranty is void if performed by non-authorized personnel. The controller displays 'Service due!' after one year of operation and repeats weekly until confirmed. Also perform annual electrical safety tests per EN 50678/VDE 0701 and EN 50699/VDE 0702.

Operating Manual FP (E3.1), chap. 21.2 and 21.1, p. 65
Maintenance
Replace door gasket only when chamber is cold

Attempting to replace the door gasket while the chamber is hot can damage the gasket material and cause leaks. Allow the chamber to cool to ambient before service. Also, never remove the rear panel or allow water to penetrate the chamber during cleaning.

Operating Manual FP (E3.1), chap. 21.2, p. 65

Selection & Integration Q&A

When should I choose the FP 115 over the FP 56 or FP 260?

Choose the FP 115 when your load fits within 116 L and requires up to 300 °C with 1.4 kW single-phase power. The FP 56 offers smaller capacity (56 L) for lower throughput, while the FP 260 provides larger volume but requires a higher-power circuit. The FP 115 balances capacity and electrical demand for typical lab-scale drying.

What are the requirements for connecting the FP 115 to a building management system or alarm panel?

Use the zero-voltage switching contact via the SUB-D socket 'ALARM / SWITCH CONTACT' on the rear. The contact is rated for low-voltage signals and can be wired to a BMS or alarm panel. For analog temperature output, the optional ANALOG OUTPUT socket is required. Ensure any connected device complies with EN 61010-1.

Can the FP 115 be installed in a corner or recessed area?

No. The chamber must not be installed in unventilated recesses. Maintain at least 160 mm clearance at the rear and 100 mm at the sides for heat dissipation. The installation site must be indoors and well-ventilated to prevent overheating and ensure temperature uniformity.

What maintenance and accessories are required beyond the base unit?

Annual maintenance by BINDER-authorized personnel is mandatory to preserve warranty, including electrical safety tests per EN 50678/50699. The base delivery does not include a Pt100 object temperature sensor, HEPA fresh air filter, or APT-COM 4 software; these must be ordered separately if needed. The service reminder appears after one year and must be confirmed.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model FP 115 en
Datasheet ·590,9 КБ ·7 pp.
View PDF
PDF
FP (E3.1) 06 2026 en
Manual ·7,9 МБ ·93 pp.
View PDF