BINDER KBF 470 — Climate chambers with Temperature 0…70 °C.

The BINDER KBF 470 is a 472 L constant climate chamber with a 0...70 °C temperature range and 10...80 % RH humidity control, housed in a 930 mm wide cabinet. It is the mid-size standard model in the KBF series, distinguished from smaller 130/260 and larger 1060/1600 siblings by its 40 kg per-rack load capacity and 530 W max heat compensation. It serves as a programmable test station for reproducible temperature and humidity conditioning of samples in laboratory and industrial qualification workflows.

Temperature
0…70 °C

BINDER KBF Series — Model Comparison & Selection

Model Temperature, °CInterior volume, LLoad per rack, kgMax heat compensation, WNominal power, kWTemperature uniformity, K
KBF 470 Current 0–70472405301.40.2
KBF 1060 0–701070406401.60.3
KBF 130 0–70127302701.60.2
KBF 1600 0–701610406401.70.4
KBF 260 0–70262404101.40.3
KBF 720 0–70727406401.60.3

Full specifications

Core Performance & Functional Specifications
Temperature 0...70 °C
Interior volume 472 L (KBF470-230V / KBF470UL-120V)
Load per rack 40 kg (KBF470-230V / KBF470UL-120V)
Maximum heat compensation 530 W
Nominal power 1.4 kW (KBF470-230V / KBF470UL-120V)
Temperature fluctuation 0.1 K
Temperature uniformity 0.2 K
Article Number mfr 9020
Recovery time after door was opened for 30 s at 25°C and 60% RH mfr 2 min
Recovery time after door was opened for 30 s at 40°C and 75% RH mfr 4 min
Dimensions, Weight & Mechanical Design
Height 1951 mm (KBF470-230V / KBF470UL-120V)
Width 930 mm (KBF470-230V / KBF470UL-120V)
Electrical, Power & Drive Specifications
Power frequency 50/60 Hz (KBF470-230V / KBF470UL-120V)
Rated voltage 220...240 V
Unit fuse 16.0 A (KBF470-230V / KBF470UL-120V)
Operating Conditions, Environment & Protection
Humidity fluctuation 1.5 % RH
Humidity range 10...80 % RH

Description

Construction and series positioning

Behavior under load and in humidity mode

Installation, control and operation

Integration & Operating Notes

Process Limits
Humidity control is restricted to 10...70 °C and 10...80 % RH

With humidity active the working range narrows to 10...70 °C; the full 0...70 °C range is available only in dry operation. Humidity is controlled between 10 and 80 % RH with 1.5 % RH fluctuation. Programming a humidified setpoint below 10 °C or outside the RH band will not be reached and may trigger humidity-system alarms.

Data Sheet Model KBF 470 en.pdf; Operating Manual KBF/KBF PRO (E7), Section 18 Humidification / Dehumidification System
Process Limits
Keep sample heat load below 530 W and rack load below 40 kg

The 530 W maximum heat compensation is the total heat that samples or powered equipment inside the chamber may introduce without disturbing temperature and humidity control. Each rack is rated for 40 kg. Exceeding either limit invalidates the specified 0.1 K fluctuation and 0.2 K uniformity and can drive the chamber into alarm.

Factory Specification Sheet; Operating Manual KBF/KBF PRO (E7), Section 24 Technical Description
Maintenance
Plan annual maintenance and twice-yearly condenser cleaning

Regular maintenance must be performed at least once a year, including the steam humidifier, and the condensers should be cleaned at least twice a year by a qualified technician (more often in dusty ambient air). After 8760 operating hours or two years the controller shows a maintenance message that recurs every two weeks until reset by BINDER Service. Refrigeration work requires personnel trained per EN 13313:2010.

Operating Manual KBF/KBF PRO (E7), Sections 23.1 and 23.2 Maintenance and Service
Maintenance
Replace the door gasket only while the chamber is cold

The door gasket must be replaced with the chamber at ambient/cold condition; replacing it while warm can damage the gasket and compromise the seal, which in turn affects recovery time and humidity stability. Schedule gasket replacement during planned downtime before a conditioning run.

Operating Manual KBF/KBF PRO (E7), Section 23.2 Maintenance and Service

Selection & Integration Q&A

When should I choose the KBF 470 rather than the 260 or 1060 sibling?

Pick the 470 when 472 L interior volume and 40 kg per-rack load with 530 W heat compensation fit the sample batch; the 260 is smaller, while the 1060 and 1600 add capacity but not a wider standard temperature range (all 0...70 °C).

What supply and protection does the KBF 470-230V need, and can I use the UL variant instead?

The 230 V model needs 220...240 V, 1~, 50/60 Hz, 1.4 kW, protected at 16.0 A. The UL variant (9020-0496) is a separate 120 V unit with NEMA 5-20 plug; choose it only where 120 V is the site supply, as the two are not interchangeable.

What installation clearances and stability measures apply to this size?

As a size 260-and-larger chamber it requires the wall spacer to maintain the specified wall distance, and the 470 specifically uses the flexible tilt protection kit. Both are needed for correct airflow and safe loading.

What maintenance and accessories are not included in the base delivery?

Annual maintenance (including steam humidifier) and twice-yearly condenser cleaning are user responsibilities; the maintenance reminder appears at 8760 h or two years. Options such as APT-COM 4, analog outputs, zero-voltage alarm contacts, internal LED light, door lock, Pt 100 object sensor and external water cans are ordered separately.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KBF 470 en
Datasheet ·635,5 КБ ·9 pp.
View PDF
PDF
KBF+KBF PRO (E7) 09 2025 en
Manual ·16,5 МБ ·171 pp.
View PDF