BINDER MKT 115
BINDER ·Made in Germany · In Active Production

BINDER MKT 115

BINDER MKT 115 — Climate chambers with Temperature -70…180 °C.

The MKT 115 is a 115 L dynamic climate chamber with a -70 to 180 °C working range, using a two-stage refrigeration system to reach deep temperatures. It is the smallest MKT model, distinguished from the MKT 240 and MKT 720 by chamber volume, and is intended for rapid temperature-change testing of small to medium loads in a laboratory setting.

Temperature
-70…180 °C

BINDER MKT Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior height, mmInterior volume, LInterior width, mm
MKT 115 Current -70–18086.5172.5480115600
MKT 240 -70–18092.5193.5700228735
MKT 720 -70–180114200.510207341200

Full specifications

Core Performance & Functional Specifications
Temperature -70...180 °C
Interior volume 115 L
Load per rack 30 kg (MKT115-400V)
Nominal power 5.5 kW
Article Number mfr 9020
Dimensions, Weight & Mechanical Design
Depth 865 mm (MKT115-400V)
Height 1725 mm (MKT115-400V)
Interior height 480 mm (MKT115-400V)
Interior width 600 mm (MKT115-400V)
Width 980 mm (MKT115-400V)
Net weight of the unit (empty) mfr 305 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 400 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 3 N/PE
Operating Conditions, Environment & Protection
Cooling down time from 180°C to -70°C mfr 110 min
Standards, Safety & General Data
Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr 5.3 K/min

Description

Construction and series positioning

Thermal behavior and control

Installation and operation

Integration & Operating Notes

Electrical
400 V three-phase supply with 16 A fuse required

The standard MKT115-400V requires a 3 N/PE 400 V 50 Hz supply with a 16.0 A unit fuse and 5.5 kW nominal power. The supply must be sized for the full load and protected accordingly; a single-phase or lower-voltage supply will not run the chamber. For 480 V 60 Hz sites, the -C variant with voltage and frequency changer must be ordered instead.

Data Sheet Model MKT 115 en.pdf; Factory Specification Sheet
Mounting
Do not install in unventilated recesses or damp areas

The chamber dissipates heat during operation and must be installed with sufficient ventilation and the prescribed minimum distances to walls and adjacent equipment. Installation in unventilated recesses or damp areas risks overheating and electrical hazard; the chamber must never become wet during operation, cleaning or maintenance.

Operating Manual MK/MKT (E5), Chapter 1.8, page 12-13
Process Limits
No explosive atmospheres or flammable substances near the chamber

The chamber has no explosion protection measures and must not be operated in potentially explosive areas or with combustible dust or air-solvent mixtures present. Charging materials that release flammable vapors at working temperature must be excluded from the test plan.

Operating Manual MK/MKT (E5), Chapter 1.8, page 13
Control & I/O
Ethernet interface standard; RS485 and analog outputs optional

The MB2 controller provides an Ethernet serial interface for connection to APT-COM™ 4 software. RS485 and analog temperature outputs are available as factory options and must be specified at ordering. Network configuration is performed in the controller settings menu.

Operating Manual MK/MKT (E5), Chapters 2, 14.5 and 20
Maintenance
Annual electrical safety test and post-maintenance test required

An annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required, as well as a test before initial startup and before resuming operation after any maintenance or repair. The chamber must not be returned to service without this test.

Operating Manual MK/MKT (E5), Chapter 22.1, page 118

Selection & Integration Q&A

When should I choose the MKT 115 instead of the MKT 240 or MKT 720?

Choose the MKT 115 when the test load fits within its 115 L interior (600 × 480 mm) and the 30 kg per-rack load limit. The MKT 240 and MKT 720 offer larger interiors for bigger or multiple samples; all three share the -70...180 °C range and similar thermal performance.

What are the installation and duty-cycle limits for the MKT 115?

Install with sufficient ventilation and minimum distances; do not place in unventilated recesses or damp areas. The chamber must not be operated in explosive atmospheres. A maintenance message appears after 8760 operating hours or two years, and annual electrical safety testing is required.

What maintenance and accessories are needed beyond the base delivery?

The base delivery does not include APT-COM™ 4 software, RS485 or analog outputs, Pt 100 sensor, compressed air connection or dryer. These must be ordered as options. Maintenance is required at 8760 h or two years by BINDER Service, plus annual electrical safety tests per EN 50678/50699.

Dimensional Drawings & Technical Documentation

Dimensional drawings

MKT 115 — drawing 1

Click drawing to inspect vector dimensions

Technical documentation

PDF
Data Sheet Model MKT 115 en
Datasheet ·720,5 КБ ·8 pp.
View PDF
PDF
MK+MKT (E5) 10 2026 en
Manual ·15,9 МБ ·169 pp.
View PDF