BINDER MKT 720
BINDER ·Made in Germany · In Active Production

BINDER MKT 720

BINDER MKT 720 — Climate chambers with Temperature -70…180 °C.

The BINDER MKT 720 is a 734 L alternating climate chamber with a -70...180 °C range, using a cascade refrigeration system for deep-temperature cycling. As the largest MKT model, it delivers 13 kW nominal power and 4.5 K/min average heating between -40 and 180 °C, with 4.0 K/min cooling between 180 and -40 °C. It is intended for material and component testing requiring rapid temperature changes under program control.

Temperature
-70…180 °C

BINDER MKT Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior height, mmInterior volume, LInterior width, mm
MKT 720 Current -70–180114200.510207341200
MKT 115 -70–18086.5172.5480115600
MKT 240 -70–18092.5193.5700228735

Full specifications

Core Performance & Functional Specifications
Temperature -70...180 °C
Interior volume 734 L
Load per rack 40 kg (MKT720-400V, 50 Hz)
Nominal power 13 kW
Article Number mfr 9020
Dimensions, Weight & Mechanical Design
Depth 1140 mm (MKT720-400V, 50 Hz)
Height 2005 mm (MKT720-400V, 50 Hz)
Interior height 1020 mm (MKT720-400V, 50 Hz)
Interior width 1200 mm (MKT720-400V, 50 Hz)
Width 1580 mm (MKT720-400V, 50 Hz)
Net weight of the unit (empty) mfr 610 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 400 V
Unit fuse 32.0 A
Phase (Nominal voltage) mfr 3 N/PE
Operating Conditions, Environment & Protection
Cooling down time from 180°C to -70°C mfr 120 min
Standards, Safety & General Data
Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr 4.5 K/min

Description

Construction and series positioning

The MKT 720 is the largest chamber in the MKT series, offering 734 L of interior volume with dimensions of 1,200 mm width, 1,020 mm height, and 1,140 mm depth. Unlike the smaller MKT 115 and MKT 240, which share the same -70...180 °C range, the MKT 720 is designed for higher throughput and larger test loads, with a net weight of 610 kg and a load capacity of 40 kg per rack. The chamber is equipped with a 13 kW nominal power system and operates on a 400 V, 3 N/PE, 50 Hz supply with a 32 A unit fuse. The standard model (MKT720-400V, Art.-No. 9020-0387) is distinguished from the -C variant, which includes a voltage and frequency changer for 480 V 60 Hz operation.

Thermal behavior and control

Temperature uniformity ranges from 0.1 to 2.0 K and fluctuation from 0.1 to 0.5 K, depending on the set value. The chamber achieves an average heating rate of 4.5 K/min between -40 °C and 180 °C and an average cooling rate of 4.0 K/min between 180 °C and -40 °C, with a cooling-down time of 120 minutes from 180 °C to -70 °C. These performance figures follow IEC 60068-3-5 and are supported by a cascade refrigeration system using R452a and R23 refrigerants. The MB2 controller provides program control, fixed-value operation, timer and week programs, and a chart recorder display for documentation.

Installation and operation

Installation requires sufficient ventilation and adherence to minimum distances to walls and ceilings, as the chamber must not be installed in unventilated recesses. The chamber is not explosion-protected and must not be operated in hazardous locations or with combustible materials. Electrical connection must be made by qualified personnel, and the chamber includes a rear power switch (for sizes 115 and above) in addition to the main power switch. The Ethernet interface allows networking with APT-COM™ 4 software for remote monitoring and data logging. Optional RS485, analog temperature outputs, and a flexible Pt 100 sensor are available for integration into existing systems.

Integration & Operating Notes

Electrical
Requires 400 V three-phase supply with 32 A fusing

The MKT 720 draws 13 kW nominal power and must be connected to a 400 V, 3 N/PE, 50 Hz supply with a 32 A unit fuse. Verify that the site provides this capacity and that the over-current protection is sized accordingly; a dedicated circuit is recommended to avoid nuisance trips during heating and cooling cycles.

Data Sheet Model MKT 720 en.pdf; Factory Specification Sheet
Process Limits
Not for use in explosive atmospheres or with flammable substances

The chamber has no explosion protection measures. It must not be operated in potentially explosive areas, and no combustible dust or air-solvent mixtures may be present near the chamber. Ensure that test materials do not release flammable vapors at working temperatures, as this could create an ignition hazard.

Operating Manual MK/MKT (E5), Chapter 1.8, page 13
Maintenance
Plan service at 8760 operating hours or two years

A maintenance message appears after 8760 operating hours or two years. After confirmation, the message reappears every two weeks until reset by BINDER Service. Schedule service intervention before this interval to avoid recurring messages and to maintain warranty and performance. Annual electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 is also required.

Operating Manual MK/MKT (E5), Chapter 22.2, page 120; Chapter 22.1, page 118
Control & I/O
Ethernet interface enables networked monitoring with APT-COM™ 4

The chamber is equipped with an Ethernet serial interface and can be networked with BINDER APT-COM™ 4 Multi Management Software (accessory). This allows remote monitoring, data logging, and integration into laboratory management systems. Network configuration is described in chapter 14.5 of the manual; APT-COM™ 4 requires separate registration.

Operating Manual MK/MKT (E5), Chapter 2, page 19; Chapters 14.5 and 20.1

Selection & Integration Q&A

When should I choose the MKT 720 over the smaller MKT 115 or MKT 240?

Choose the MKT 720 when you need a large interior volume (734 L) for testing bulky components or high-throughput batches. The MKT 115 and MKT 240 offer the same temperature range but smaller volumes (115 L and 240 L respectively). The MKT 720 also provides higher nominal power (13 kW) to handle larger loads, but requires a 400 V three-phase supply with 32 A fusing, which may not be available in all labs.

What are the power supply and fusing requirements for the MKT 720, and can it be integrated with a building management system?

The MKT 720 requires a 400 V, 3 N/PE, 50 Hz supply with a 32 A unit fuse. It includes an Ethernet interface for networking with APT-COM™ 4 software, enabling integration with building management or laboratory monitoring systems. Optional RS485 and analog temperature outputs are available for connection to PLCs or data loggers.

What are the installation clearance and environmental limits for the MKT 720?

Install the chamber with sufficient ventilation and observe minimum distances to walls and ceilings as specified in chapter 3.4 of the manual. Do not install in unventilated recesses or damp areas. The chamber must not be operated in explosive atmospheres or hazardous locations. Ambient temperature and humidity limits are given in the technical data; ensure the site meets these to avoid condensation and overheating.

What maintenance is required and what accessories are needed for operation?

A maintenance message appears after 8760 operating hours or two years, requiring service by BINDER. Annual electrical safety testing is mandatory. Base delivery does not include the Ethernet cable for APT-COM™ 4, RS485 interface, analog outputs, or the flexible Pt 100 sensor; these must be ordered separately. Also, the compressed air connection and dryer are optional for specific test profiles.

Dimensional Drawings & Technical Documentation

Dimensional drawings

MKT 720 — drawing 1

Click drawing to inspect vector dimensions

Technical documentation

PDF
Data Sheet Model MKT 720 en
Datasheet ·851,2 КБ ·8 pp.
View PDF
PDF
MK+MKT (E5) 10 2026 en
Manual ·15,9 МБ ·169 pp.
View PDF