BINDER MKT 115
BINDER MKT 115 — Climate chambers with Temperature -70…180 °C.
The MKT 115 is a 115 L dynamic climate chamber with a -70 to 180 °C working range, using a two-stage refrigeration system to reach deep temperatures. It is the smallest MKT model, distinguished from the MKT 240 and MKT 720 by chamber volume, and is intended for rapid temperature-change testing of small to medium loads in a laboratory setting.
BINDER MKT Series — Model Comparison & Selection
Full specifications
| Temperature | -70...180 °C |
| Interior volume | 115 L |
| Load per rack | 30 kg (MKT115-400V) |
| Nominal power | 5.5 kW |
| Article Number mfr | 9020 |
| Depth | 865 mm (MKT115-400V) |
| Height | 1725 mm (MKT115-400V) |
| Interior height | 480 mm (MKT115-400V) |
| Interior width | 600 mm (MKT115-400V) |
| Width | 980 mm (MKT115-400V) |
| Net weight of the unit (empty) mfr | 305 kg |
| Power frequency | 50 Hz |
| Rated voltage | 400 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 3 N/PE |
| Cooling down time from 180°C to -70°C mfr | 110 min |
| Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr | 5.3 K/min |
Description
Construction and series positioning
- The MKT 115 is the entry size of the MKT series, with 115 L interior volume (600 mm wide × 480 mm high) and a net weight of 305 kg. Sibling models MKT 240 and MKT 720 share the same -70...180 °C range but offer larger interiors.
- The standard variant MKT115-400V (Art.-No. 9020-0385) operates on 400 V 3 N/PE 50 Hz with 5.5 kW nominal power and a 16.0 A unit fuse. A -C variant with an integrated voltage and frequency changer exists for 480 V 3~ 60 Hz supplies; the -C suffix denotes that additional hardware and its separate installation instructions.
- The chamber uses a two-stage refrigerant circuit (R452a and R23, per the MKT 240 type-plate example) to achieve -70 °C; the rear power switch is located on the back of the unit from size 115 upward.
Thermal behavior and control
- Average heating rate is 5.3 K/min between -40 °C and 180 °C (IEC 60068-3-5); average cooling rate is 4.2 K/min between 180 °C and -40 °C. A full cool-down from 180 °C to -70 °C takes 110 min.
- Temperature uniformity depends on setpoint and ranges from 0.2 to 1.8 K; fluctuation ranges from 0.1 to 0.6 K. These values apply to the empty chamber under the stated conditions and will degrade with load.
- The MB2 controller supports fixed-value mode, time programs (with ramps, steps, tolerance bands and section repetition) and week programs. Setpoints can be entered via the Setpoints menu or directly on the normal display.
Installation and operation
- The chamber must be installed with sufficient ventilation for heat dispersal and the prescribed minimum distances; it must not be placed in unventilated recesses. It must not be operated in explosive atmospheres or damp locations.
- An Ethernet serial interface is standard for networking with APT-COM™ 4 software. Options include RS485, analog temperature outputs, a flexible Pt 100 object sensor, compressed air connection and compressed air dryer. Water cooling is not available for the MKT 115.
- The over-temperature protective device class 1 trips if the nominal temperature is exceeded by 20 °C due to a chamber failure; a class 2.0 safety device per DIN 12880 is fitted as standard.
Integration & Operating Notes
The standard MKT115-400V requires a 3 N/PE 400 V 50 Hz supply with a 16.0 A unit fuse and 5.5 kW nominal power. The supply must be sized for the full load and protected accordingly; a single-phase or lower-voltage supply will not run the chamber. For 480 V 60 Hz sites, the -C variant with voltage and frequency changer must be ordered instead.
The chamber dissipates heat during operation and must be installed with sufficient ventilation and the prescribed minimum distances to walls and adjacent equipment. Installation in unventilated recesses or damp areas risks overheating and electrical hazard; the chamber must never become wet during operation, cleaning or maintenance.
The chamber has no explosion protection measures and must not be operated in potentially explosive areas or with combustible dust or air-solvent mixtures present. Charging materials that release flammable vapors at working temperature must be excluded from the test plan.
The MB2 controller provides an Ethernet serial interface for connection to APT-COM™ 4 software. RS485 and analog temperature outputs are available as factory options and must be specified at ordering. Network configuration is performed in the controller settings menu.
An annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required, as well as a test before initial startup and before resuming operation after any maintenance or repair. The chamber must not be returned to service without this test.
Selection & Integration Q&A
Choose the MKT 115 when the test load fits within its 115 L interior (600 × 480 mm) and the 30 kg per-rack load limit. The MKT 240 and MKT 720 offer larger interiors for bigger or multiple samples; all three share the -70...180 °C range and similar thermal performance.
Install with sufficient ventilation and minimum distances; do not place in unventilated recesses or damp areas. The chamber must not be operated in explosive atmospheres. A maintenance message appears after 8760 operating hours or two years, and annual electrical safety testing is required.
The base delivery does not include APT-COM™ 4 software, RS485 or analog outputs, Pt 100 sensor, compressed air connection or dryer. These must be ordered as options. Maintenance is required at 8760 h or two years by BINDER Service, plus annual electrical safety tests per EN 50678/50699.