Peter Huber Kältemaschinenbau KISS K6 Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau KISS K6 Kälte-Badumwälzthermostat — Thermostatic baths with Noise level 57 dB.
The KISS K6 is a refrigerated bath circulator with a 4.5 L bath and a non-variable pressure/suction pump, covering -25 to 200 °C with ±0.05 K stability. Its 0.2 kW cooling at 20 °C falls to 0.05 kW at -20 °C, so it suits small sub-ambient and high-temperature duties. It serves as a temperature-controlled source for external circuits via M16x1 connections.
Direct drop-in replacements
Peter Huber Kältemaschinenbau KISS Series — Model Comparison & Selection
| Model | Noise level, dB | Power supply | Pump connection | Temperature range, °C | Weight, kg | Filling volume, m³ |
|---|---|---|---|---|---|---|
| KISS K6 Kälte-Badumwälzthermostat Current | 57 | 208-240V 1~/2~ 50/60Hz | M16x1 AG male | -25–200 | 25 | 0.002 |
| KISS K12 Kälte-Badthermostat | — | 110-115V 1~ 50/60Hz | (Option, #19607) | -20–200 | 28 | — |
| KISS K20 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.01 |
| KISS K25 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 39 | 0.015 |
| KISS K6s Kälte-Badumwälzthermostat | 57 | 208-240V 1~/2~ 50/60Hz | M16x1 AG male | -25–200 | 25 | 0.002 |
Full specifications
| Noise level | 57 dB(A) |
| Filling volume | 2 l (min) |
| Maximum delivery flow rate | 14 l/min (pressure pump max, 0, 25 bar) |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | no |
| Pump type | Pressure and suction pump, speed not variable |
| Temperature range | -25...200 °C |
| Temperature stability mfr | ±0, 05 K |
| Alarm message mfr | optic, acoustic |
| Dimensions (W x D x H) mfr | 210 x 400 x 546 mm |
| Heating capacity mfr | 1, 6 - 2, 1 kW |
| Delivery capacity suction pump max. mfr | 10, 5 l/min; 0, 17 bar |
| Bath opening WxD/BD mfr | 140 x 120 / 150 |
| Bath volume mfr | 4, 5 l |
| Weight | 25 kg (net) |
| Power supply | 208-240V 1~/2~ 50/60Hz |
| Current consumption max. mfr | 10 + 1, 4 A (max) |
| Controller | KISS |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Cooling capacity mfr | 20:0, 2; 0:0, 15; -10:0, 1; -20:0, 05 °C / kW |
| Safety class | III / FL |
Description
Construction and series positioning
The KISS K6 is the smallest refrigerated bath in the KISS series, with a 4.5 L bath volume and a working range of -25 to 200 °C. Unlike the larger K12, K15, K20 and K25 siblings, it uses a single-stage air-cooled refrigeration circuit with R-290 (A3, H220) refrigerant. The pump is a pressure and suction type with fixed speed, delivering up to 14 l/min at 0.25 bar on the pressure side and 10.5 l/min at 0.17 bar on the suction side. Wetted parts are stainless steel or plastics, and the pump connections are M16x1 male. The unit is rated IP20 and safety class III/FL per DIN 12876, with adjustable overtemperature protection.
Behavior under load
Cooling capacity is strongly temperature-dependent: 0.2 kW at 20 °C, 0.15 kW at 0 °C, 0.1 kW at -10 °C and 0.05 kW at -20 °C, all referenced to a 20 °C room. Heating capacity is 1.6–2.1 kW. Temperature stability is ±0.05 K. The fixed-speed pump means flow cannot be trimmed; external pressure drop must be matched by circuit design. The plastic pump is not resistant to acetone or polyglycols, so fluid selection must respect wetted-material limits. For sub-zero operation, water should be mixed with glysantine or ethylene glycol for freeze protection.
Installation and operation
The unit requires 208–240 V, 1~ or 2~, 50/60 Hz with a 10–16 A fuse; voltage tolerance is ±5% and frequency ±2%, but the combination of -5% voltage with +2% frequency is not allowed. A country-specific plug must be specified at ordering. Ambient limits are 5–40 °C, indoor dry use only. The air-cooled condenser needs clearance and cool intake air. Control is via the KISS controller with OLED display, USB device and RS232 interfaces, and optic/acoustic alarm. Class A Group 1 per EN55011.
Integration & Operating Notes
The unit accepts 208–240 V at 50/60 Hz with ±5% voltage and ±2% frequency tolerance, but the combination of -5% voltage and +2% frequency is explicitly not allowed. Verify the site feed against this combined envelope before connection, and provide a 10–16 A fuse.
Published cooling capacity (0.2 kW at 20 °C down to 0.05 kW at -20 °C) is valid at 20 °C room temperature and may drop as ambient rises. Size the heat load against the capacity at the actual setpoint and expected room temperature, not the 20 °C figure.
The pump is made of plastic and is not resistant to acetone or polyglycols (depending on manufacturer). Select heat transfer fluid accordingly; for sub-zero duty, mix water with glysantine or ethylene glycol for freeze protection rather than using pure water.
IP20 rating and 5–40 °C ambient limit restrict the unit to indoor dry locations. The refrigeration machine is air-cooled, so the installation site must provide adequate clearance and cool intake air; confined or hot spaces will reduce cooling capacity.
Single and two-phase devices are delivered with a power cable and a country-specific plug. The plug type must be stated when ordering; otherwise the unit may arrive with a plug unsuitable for the installation country.
The refrigerant is R-290 (A3, H220), classified as flammable. Installation location and room ventilation must be reviewed for ignition sources and refrigerant concentration limits according to local safety regulations.
Selection & Integration Q&A
Choose the K6 for small bath volumes (4.5 L) and duties within its cooling curve (0.2 kW at 20 °C, 0.05 kW at -20 °C). The larger siblings offer more cooling and bath volume; the K20/K25 also run quieter at 50 dB(A) versus 57 dB(A) for the K6.
The pump connections are M16x1 male. The fixed-speed pressure/suction pump delivers up to 14 l/min at 0.25 bar pressure and 10.5 l/min at 0.17 bar suction. External circuit pressure drop must be matched to this fixed flow; the pump speed cannot be adjusted.
Operate indoors, dry, between 5 and 40 °C ambient, IP20. The air-cooled condenser needs clearance and cool intake air. Cooling capacity is rated at 20 °C room and derates at higher ambient. Supply must be 208–240 V, 1~/2~, 50/60 Hz with a 10–16 A fuse.
The base delivery includes the unit with power cable and country-specific plug (must be specified). Heat transfer fluid is not included; for sub-zero use, mix water with glysantine or ethylene glycol. Wetted parts are stainless steel or plastics; avoid acetone and polyglycols with the plastic pump.