Shimadzu Aggregates Sizer
Shimadzu ·Made in Japan · In Active Production

Shimadzu Aggregates Sizer

Shimadzu Aggregates Sizer — Particle size analyzers.

The Aggregates Sizer is a laser diffraction particle size analyzer using a violet laser and Wing sensor II with 78 concentric elements, capturing forward-scattered light up to 60°. It measures 100 nm to 10 µm at concentrations from 0.1 to 1000 ppm, with a minimum 30 s interval, serving aggregation analysis in biopharmaceuticals. The standard unit is distinguished from the TC variant by lacking temperature control.

Direct drop-in replacements

0%
match
⚖️ Compare
0%
match
⚖️ Compare
0%
match
⚖️ Compare

Full specifications

Core Performance & Functional Specifications
Detector types Wing sensor II, 78 concentric sensor elements
Forward scatter angle up to 60 °
Measurement range 100 nm to 10 µm
Measurement range (SALD-7500nano) mfr 0.007 to 800 μm (SALD-7500nano)
Measurement range (SALD-2300) mfr 0.017 to 2500 μm (SALD-2300)
Measurement range (SALD-201V) mfr 0.25 to 350 μm (SALD-201V)
Wavelength (SALD-7500nano) mfr 405 nm (SALD-7500nano)
Wavelength (SALD-2300) mfr 680 nm (SALD-2300)
Wavelength (SALD-201V) mfr 670 nm (SALD-201V)
Detector elements (SALD-7500nano) mfr 84 elements (SALD-7500nano)
Detector elements (SALD-2300) mfr 84 elements (SALD-2300)
Detector elements (SALD-201V) mfr 54 elements (SALD-201V)
Sampler (circulation type) (SALD-7500nano) mfr available (SALD-7500nano)
Sampler (circulation type) (SALD-2300) mfr available (SALD-2300)
Sampler (circulation type) (SALD-201V) mfr available (SALD-201V)
Wet measurement batch cell (SALD-7500nano) mfr available (SALD-7500nano)
Wet measurement batch cell (SALD-2300) mfr available (SALD-2300)
Wet measurement batch cell (SALD-201V) mfr available (SALD-201V)
High-concentration sample measurement (SALD-7500nano) mfr available (SALD-7500nano)
High-concentration sample measurement (SALD-2300) mfr available (SALD-2300)
High-concentration sample measurement (SALD-201V) mfr available (SALD-201V)
Dispersion bath capacity (SALD-MS75) mfr 100 to 300 cm3 (SALD-MS75)
Dispersion bath capacity (SALD-MS23) mfr 100 to 300 cm3 (SALD-MS23)
Max. pump flow rate (SALD-MS75) mfr 2000 cm3/min (SALD-MS75)
Max. pump flow rate (SALD-MS23) mfr 2000 cm3/min (SALD-MS23)
Batch cell capacity (SALD-BC23) mfr 12 cm3 (SALD-BC23)
Batch cell capacity (SALD-BC75) mfr 5 cm3 (SALD-BC75)
Particle size divisions (SALD Series software) mfr Fixed 51 or 101 divisions (SALD Series software)
Measurement principle mfr Laser diffraction method by violet laser
Concentration range mfr 0.1 ppm to 1000 ppm ppm
Minimum measurement interval mfr 30 seconds
Micro cell sample amount mfr 0.125 mL
Batch cell sample amount mfr 5 mL
Repeatability CV (PSL 1 μm, 0.5–5 ppm) mfr less than 3 % (PSL 1 μm, 0.5–5 ppm)
Dimensions, Weight & Mechanical Design
Length 680 or more mm
Width Approx. 400 mm
Electrical, Power & Drive Specifications
Required power supply mfr AC100V, 13A, 50/60 Hz

Description

Construction and positioning

Behavior and measurement modes

Installation, connection and operation

Integration & Operating Notes

Process Limits
Keep sample concentration within 0.1–1000 ppm

Measurements outside the stated concentration range of 0.1 ppm to 1000 ppm are not supported by the verified specification; dilution or concentration of the biopharmaceutical sample must be planned before analysis to stay inside this window.

Factory Specification Sheet
Process Limits
Select cell type to match available sample volume

The micro cell requires 0.125 mL and the batch cell 5 mL; the chosen configuration determines the minimum sample volume that must be recovered, which is critical when aggregating biopharmaceutical material is scarce.

Factory Specification Sheet
Control & I/O
Part 11 software is an option, not base delivery

An FDA 21 CFR Part 11 compliant software package is listed as optional; laboratories requiring audit trails and electronic records must order it separately and validate it as part of the system.

Document c060-e010, scope and variants
Maintenance
Temperature control requires the TC variant

The standard Aggregates Sizer has no temperature-control function; samples sensitive to temperature must be run on the Aggregates Sizer TC variant or handled with external temperature management, since the standard unit cannot regulate sample temperature.

Document c060-e010, scope and variants

Selection & Integration Q&A

When should I choose the Aggregates Sizer instead of a SALD-7500nano or SALD-2300?

Pick the Aggregates Sizer for biopharmaceutical aggregation analysis in the 100 nm–10 µm range at 0.1–1000 ppm. The SALD-7500nano covers 0.007–800 µm and the SALD-2300 0.017–2500 µm for broader industrial sizing; choose those when the range extends beyond 10 µm.

What sample-handling accessories must be ordered with the base unit?

The base delivery does not include cells; specify the micro cell (0.125 mL) or batch cell (5 mL) according to sample volume. For regulated labs, add the optional FDA 21 CFR Part 11 software. Temperature-sensitive work requires the TC variant instead.

What are the mounting and environmental limits from the manual?

The supplied document digests contain no sections or actionable facts on mounting, environment or duty cycle; these limits must be taken from the full instruction manual before installation planning.

What maintenance and consumables are not included in base delivery?

Cells (micro 0.125 mL or batch 5 mL) and the optional Part 11 software are not part of the base unit. The digests provide no service intervals or consumable schedules; consult the full manual for maintenance planning.

Dimensional Drawings & Technical Documentation

Dimensional drawings

performance curve

Click drawing to inspect vector dimensions

Technical documentation

PDF
c060-e005
Datasheet ·2,9 МБ ·20 pp.
View PDF
PDF
c060-e010
Datasheet ·8,1 МБ ·16 pp.
View PDF