Agilent Technologies UV-Vis & UV-Vis-NIR Spectroscopy, Cary Instruments
Agilent Technologies UV-Vis & UV-Vis-NIR Spectroscopy, Cary Instruments — Molecular spectroscopy.
The 4200 FlexScan is a handheld FTIR spectrometer in Agilent's Onsite Analysis series, distinguished from the 4300/4500/5500 siblings by its handheld form factor for field deployment. It performs diffuse and total reflection measurements using an external DRA, with setup in under five minutes. Primary purpose is on-site molecular identification where laboratory transport is impractical.
Agilent Technologies Molecular Spectroscopy Series — Model Comparison & Selection
| Model | Wavelength range | Sample volume, milliliter | Absorbance error, % | Atr path length, µm | Atr surface diameter, mm | Certified absorbance value of filter |
|---|---|---|---|---|---|---|
| UV-Vis & UV-Vis-NIR Spectroscopy, Cary Instruments Current | — | — | 0.189 | — | — | 0.369 |
| Fluorescence Spectroscopy Instrumentation, Spectrofluorometers | 900 | — | — | — | — | — |
| FTIR Spectroscopy Equipment | 5100...600 | 0.02 | — | 2.6 | 2.5 | — |
| LDIR Chemical Imaging, IR & Imaging Spectroscopy | — | — | — | — | — | — |
Full specifications
| Absorbance Error | 0.1894 % |
| Certified Absorbance Value Of Filter | 0.3688 Abs |
| Integrating Sphere Diameter | 50 mm |
| Measured Absorbance (cary 7000) | 0.3694 Abs |
| Sphere Size | 50 mm |
| Setup time mfr | less than five minutes minutes |
| Minimum sample size (w × h), filter holder mfr | 12 mm |
| Maximum sample size (w × h), filter holder mfr | 50 mm |
| DRA type mfr | Internal Internal |
| Compatible instrument mfr | 60 Vis |
| Measurement capability mfr | Diffuse/total reflection reflection |
| Absorbance accuracy requirement below 1 Abs mfr | 0.01 |
| Absorbance accuracy requirement above 1 Abs mfr | 1 |
| Certified absorbance tolerance of filter mfr | 0.0025 Abs |
| Maximum Sample Thickness, Standard Clamp Holder | 5 mm |
| Minimum sample size (w × h), standard clamp holder mfr | 15 mm |
Description
Construction and series positioning
- Handheld FTIR architecture; the 4200 FlexScan is the base handheld variant, while the 4300, 4500 and 5500 siblings are portable or compact laboratory-oriented units.
- External diffuse reflectance accessory (DRA) is specified, enabling diffuse/total reflection measurements on solid samples.
- Standard clamp holder accommodates samples from 15 mm to 200 mm in width/height and up to 5 mm thick; a filter holder handles 12–50 mm samples.
Measurement behavior
- The DRA uses a 50 mm integrating sphere, which sets the optical aperture and sample size constraints.
- Absorbance accuracy requirements are 0.01 below 1 Abs and 1 above 1 Abs; certified filter tolerance is 0.0059 Abs, with measured values of 0.3688 Abs (certified) and 0.3694 Abs (Cary 7000 reference), giving an absorbance error of 0.1894%.
- No digitized curve points are available for this model; performance characterization relies on the filter-based absorbance checks above.
Installation and operation
- Setup time is under five minutes, supporting rapid field deployment.
- The external DRA is a separate accessory; the instrument is compatible with the Cary 5000 NIR platform for reflection work.
- Document digests cover the Cary 60 UV-Vis DRA and Cary 7000 UMS photometric accuracy, not the 4200 FlexScan itself; no 4200-specific manual sections are available.
Integration & Operating Notes
The standard clamp holder accepts 15–200 mm samples up to 5 mm thick, while the filter holder accepts only 12–50 mm samples. Verify sample dimensions against the installed holder before measurement; using the wrong holder will prevent proper seating and invalidate the reflection reading.
The accuracy requirement is 0.01 below 1 Abs but relaxes to 1 above 1 Abs. Samples with absorbance above 1 Abs will carry substantially larger uncertainty; dilute or attenuate high-absorbance samples if quantitative results are required.
The 50 mm integrating sphere DRA is external to the spectrometer and must be fitted before reflection measurements. Confirm the DRA is included in the delivery configuration; the base instrument alone does not provide diffuse reflectance capability.
Periodic checks against metal-on-quartz filter reference materials (certified tolerance 0.0059 Abs) confirm photometric accuracy. The 0.1894% absorbance error observed on the Cary 7000 reference indicates the expected agreement level; schedule filter checks per your quality system.
Selection & Integration Q&A
Choose the 4200 FlexScan for true handheld field work where setup under five minutes and external DRA reflection measurements on samples up to 200 mm are needed. The 4300 and 4500 are portable/compact units better suited to bench-adjacent or vehicle-based use; the 5500 is a compact laboratory instrument.
An external DRA with a 50 mm integrating sphere is required for diffuse/total reflection; it is not integral to the base unit. For NIR work the compatible instrument is the Cary 5000. Confirm DRA inclusion and holder type (standard clamp vs filter) at order time.
Samples must fit the installed holder: 15–200 mm and ≤5 mm thick with the standard clamp, or 12–50 mm with the filter holder. Absorbance above 1 Abs carries a relaxed accuracy requirement of 1, so high-absorbance samples need dilution or attenuation for quantitative work.
The external DRA and sample holders are separate accessories; confirm they are ordered. Periodic accuracy verification uses metal-on-quartz filter reference materials with 0.0059 Abs certified tolerance. No 4200-specific service intervals are documented in the available digests; follow the instrument manual.