PCG-750 / PCG-752 Vacuum Gauge Controller User Manual

Sommaire

  1. Overview pp. 2-3
    Legal notices and table of contents
  2. Procedure pp. 4-5
    Wiring and connection instructions for the controller
  3. Overview pp. 6-7
    Manual conventions, definitions and terminology
  4. Warning pp. 8-9
    Safety symbols and general safety information
  5. Warning pp. 10-11
    Improper use warnings and model-specific safety guidelines
  6. Services pp. 12-13
    Disposal instructions and service contact information
  7. Spec pp. 14-15
    Product description and technical data tables
  8. Spec pp. 16-19
    Detailed specifications, warnings and technical parameters
  9. Spec pp. 20-21
    Dimensional drawings and mounting locations
  10. Spec pp. 22-23
    Gas correction factors and installation notes
  11. Procedure pp. 24-27
    Caution notes, FCC 68 and Profibus connector wiring
  12. Procedure pp. 28-29
    Operating instructions and LCD display description
  13. Procedure pp. 30-33
    Technical details on switching functions and gas dependence
  14. pp. 34-39
    Operating notes and RS232C diagnostic port information
  15. Procedure pp. 40-41
    Profibus operation and deinstallation instructions
  16. Procedure pp. 42-44
    Maintenance, repair and measurement output notes
  17. Procedure pp. 45-46
    Troubleshooting guide and spare parts tables
  18. pp. 47-51
    Vacuum Products Division contacts and book index
  19. p. 15 table, 13 rows
  20. p. 19 table, 33 rows
  21. p. 46 table, 8 rows
  22. p. 17 table, 12 rows
  23. p. 18 table, 20 rows
  24. p. 29 table, 3 rows
  25. p. 5 table, 11 rows

Table 1 Part Numbers

Part NumberDescription
AGPS45INTPower supply, PVG-PCG, international plug.
PCG750KF16PCG-750 Pirani/capacitance diaphragm gauge, tungsten filaments, NW 16KF.
PCG750KF16RSReplacement sensor, PCG-750 Pirani/capacitance diaphragm gauge, tungsten filaments, NW 16KF.
PCG750KF16SD1PCG-750 Pirani/capacitance diaphragm gauge, tungsten filaments, with setpoints, LCD display, NW 16KF, mbar.
PCG750KF16SD2PCG-750 Pirani/capacitance diaphragm gauge, tungsten filaments, with setpoints, LCD display, NW 16KF, Torr.
PCG750KF16SD3PCG-750 Pirani/capacitance diaphragm gauge, tungsten filaments, with setpoints, LCD display, NW 16KF, Pascal.
PCG750KF16SPPCG-750 Pirani/capacitance diaphragm gauge, tungsten filaments, with setpoints, Profibus, NW 16KF, mbar.
PCG752KF16PCG-752 Pirani/capacitance diaphragm gauge, nickel-plated filaments, NW 16KF.
PCG752KF16RSReplacement sensor, for PCG-752 Pirani/capacitance diaphragm gauge, nickel-plated filaments, NW 16KF, recommended for corrosive applications.
PCG752KF16SD1PCG-752 Pirani/capacitance diaphragm gauge, nickel-plated filaments, with setpoints, LCD display, NW 16KF, mbar.
PCG752KF16SD2PCG-752 Pirani/capacitance diaphragm gauge, nickel-plated filaments, with setpoints, LCD display, NW 16KF, Torr.
PCG752KF16SD3PCG-752 Pirani/capacitance diaphragm gauge, nickel-plated filaments, with setpoints, LCD display, NW 16KF, Pascal.
PCG752KF16SPPCG-752Pirani/capacitance diaphragm gauge, nickel-plated filaments, with setpoints, Profibus, NW 16KF, mbar.

Table 2 Technical Specifications (continued)

SpecificationsValue
Data rate≤12 Mbaud
Node address
Local (Adjustable via hexadecimal <ADDRESS>, <MSD>, <LSD> switches)00 … 7Dhex (0 … 125dec)
Default setting0.1Chex
Via Profibus (hexadecimal <ADDRESS> switches set to >7Dhex (>125dec)00 … 7Dhex (0 … 125dec)
Profibus connectionD-Sub, 9-pin,female shielded,
Cablespecial Profibus cable
Cable length, system wiringaccording to Profibus specifications
Materials exposed to vacuum
Vacuum connectionstainless steel 1.4435
Filament PCG-750W
PCG-752Ni
Feedthroughglass
Orificestainless steel
Diaphragmceramic
Further materialsNi, NiFe, stainless steel 1.4301, SnAg
Internal volume4.7 cm3
Permissible pressure (absolute)≤5 bar
Bursting pressure (absolute)10 bar
Permissible temperatures
Operation+10 °C … +50 °C
Vacuum connection 4)≤80 °C
Filament<160 °C
Storage–20 °C … +65 °C
Relative humidity
Year's mean≤65% (no condensation)
During 60 days≤85% (no condensation)
Mounting orientationany
Useindoors only, altitude up to 2000 m NN
Degree of protectionIP 40
Weight
w/o fieldbus interface115 g …130 g
with fieldbus interface230 g … 250 g

4) For horizontal mounting orientation only. During bakeout, measurement range, accuracy and repeatability may deviate from specifications

Table 7 PCG-750 Spare Parts

Sensor for gauges with tungsten (W) filamentOrdering No.
PCG-750PCG750KF16
DN 16 ISO-KF
PCG750KF16RS
PCG750KF16SD1
PCG750KF16SD2
PCG750KF16SD3
PCG750KF16SP

Table 2 Technical Specifications

SpecificationsValue
Measurement range5×10-5 … 1500 mbar
Measurement principle101) mbar … 1500 mbar diaphragm capacitive sensor 1 … 101) mbar 5×10-5 … 1 mbar crossover range thermal conductance acc. to Pirani
Accuracy (N2)5×10-4 … 1×10-3 mbar ±50% of reading 1×10-3 … 100 mbar ±15% of reading 100 … 950 mbar ±5% of reading 950 … 1050 mbar ±2.5% of reading
Repeatability (N2)1×10-3 … 1100 mbar ±2% of reading
Output signal (measurement signal)Voltage range 0 … +10.23 V Measurement range +0.61 … +10.23 V Error signal 0 V (default) Voltage vs. pressure 1.286 V/decade, logarithmic
Output impedance2 × 4.7 Ω, short circuit-proof
Load impedance>10 kΩ
Response time<30 ms
Gauge identification71.5 kΩ
HV adjustmentat <10-5 mbar
ATM adjustmentat >100 mbar
Switching functionsSP1, SP2 Setting range 5.0×10-5 … 1500 mbar Hysteresis 2) 10% of threshold Switching characteristics 2) Low Trip Point Type 1 floating contact (n.o.) per switching function Contact rating closed <30 VAC/DC, ≤1 A resistive open LED lit solid Switching time LED off <30 ms

1) Crossover range for air, O2, CO and N2 10 mbar, 100 mbar in heavy gases.

Table 2 Technical Specifications (continued)

SpecificationsValue
Diagnostic portJack connector 2.5 mm, 3-pin
SupplyWARNING The gauge may only be connected to power supplies, instruments, or control devices that conform to the requirements of a grounded protective extra-low voltage (SELV). The connection to the gauge has to be fused. 3).
Supply voltage at the gauge+15 … +30 VDC
Ripple≤1 Vpp
Power consumption without fieldbus≤2.5 W
Power consumption with fieldbus≤3 W
Fuse to be connected 3)1 AT
Electrical connection
Sensor cableshielded 0.14 mm2/conductor
Cable length RS232C operation≤100 m
Cable length FCC 68≤30 m
Grounding concept Vacuum connection to signal commonSee Paragraph "Power Connection” connected via 10 Ω, 10 nF
RS232C
Transmission rate57600 baud (default)
Data formatbinary 8 data bits one stop bit no parity bit no handshake
see Paragragh “Power Connection"
For further information on the RS232C interface, see Paragragh “Diagnostic Port (RS232C interface)”
Profibus interface
Specification, data format, communication protocolRS485
Interface, physical

2) The hysteresis and the switching characteristics can be programmed via the serial interface or the diagnostic port. 3) Agilent controllers fulfill this requirement.

Table 4 LCD

LCDMeaning
offno supply voltage
lit greenmeasurement / parameter mode
lit rederror

The display can be rotated by 180 ° via the serial interface.

Instructions for Use

РазделСтраница
About this manual6
Validity6
Definitions and terminology7
Definition of Caution, Warning and Note7
Warning Symbols8
Safety9
Proper use9
Improper use10
Safety guideline for PCG-750, PCG-75211
Disposal12
Service13

Engineering Facts

12 facts · every claim linked to a page
Selection

The gauge covers 5×10-5 to 1500 mbar, so it suits both rough and fine vacuum measurement in one device.

Confirms the gauge spans the full required pressure range, avoiding two separate sensors.

Source
“designed for vacuum measurement of gases in the pressure range of 5×10-5 … 1500 mbar”
Safety

Do not use the gauge to measure flammable or combustible gases mixed with oxidants inside the explosion range.

Defines an absolute process-media restriction that must be checked before selection.

Source
“It must not be used for measuring flammable or combustible gases in mixtures containing oxidants”
Selection

Choose the PCG-752 nickel-plated filament version for corrosive applications instead of the tungsten PCG-750.

Drives the material choice between the two gauge variants for aggressive media.

Source
“recommended for corrosive applications”
Installation

The gauge must be electrically connected to a grounded vacuum chamber per EN 61010, using a conductive metallic clamping ring for KF flanges.

Protective grounding is mandatory for safe fault behavior and compliance.

Source
“Electrically connect the gauge to the grounded vacuum chamber.”
Installation

Connect the gauge only to SELV-compliant supplies and fuse the connection.

Defines the electrical supply safety requirement for wiring the gauge.

Source
“The gauge may only be connected to power supplies, instruments, or control devices”
Installation

Ground the cable shield on one side only via chassis ground and use a differential measurement input to avoid ground loops.

Prevents EMC and potential-difference errors that corrupt the measurement signal.

Source
“Connect the cable shield to ground on one side via the chassis ground. Do not”
Operation

Allow at least 10 minutes stabilization after power-up and preferably run the gauge continuously.

Sets realistic expectations for measurement readiness and drift behavior.

Source
“Allow a stabilization period of at least 10 minutes.”
Operation

The setpoints SP1 and SP2 are factory set to the lower measurement range limit and therefore do not switch until configured.

Prevents false assumption that relays work out of the box; must be programmed.

Source
“The setpoints SP1 and SP2 are factory set to the lower measurement range limit”
Operation

The Profibus variant must not be operated via the RS232C interface because it causes data transmission errors.

Avoids a wiring/commissioning mistake on Profibus units.

Source
“This gauge must not be operated with the RS232C interface.”
Maintenance

Zero drift can occur over time or from contamination, so periodically check and adjust the zero under the same ambient conditions and mounting orientation.

Establishes a routine maintenance task to keep accuracy within spec.

Source
“Periodically check the zero and adjust it if necessary”
Maintenance

Gauge failures from contamination or wear, and expendable parts like seals and filament, are not covered by warranty.

Informs the user about warranty limits and the need for careful handling.

Source
“Gauge failures due to contamination or wear and tear, as well as expendable parts”
Installation

The ceramic sensor can be damaged by impacts, so do not drop the product or allow shocks.

Handling precautions during mounting protect the sensitive diaphragm sensor.

Source
“The ceramic sensor may be damaged by impacts.”