Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 708-DS Dissolution Apparatus

Agilent Technologies 708-DS Dissolution Apparatus — Dissolution testing with Pumping speed 3×10⁻⁵ m³/h, Temperature 37 °C.

The Agilent 708-DS is a 6-position (G7910A) or 8-position (G7911A) dissolution apparatus with a 1 L vessel volume and motorized drive unit lift. It operates at 37 °C with spindle speeds of 10–250 RPM and a water bath range of ambient +5 to 55 °C. It is used for USP Apparatus 1 and 2 dissolution testing with optional automated sampling and online HPLC analysis.

Pumping speed
3×10⁻⁵ m³/h
Temperature
37 °C

Agilent Technologies Dissolution Apparatus Series — Model Comparison & Selection

Model Temperature, °CCoverslip typeWeight, kgOperating altitude, mPumping speed, m³/hOperating temperature, °C
708-DS Dissolution Apparatus Current 370.154.420003×10⁻⁵—
400-DS Apparatus 7 Delayed StartUse the drop-down arrows592000—55

Full specifications

Core Performance & Functional Specifications
Pumping speed 0.5 mL/min
Temperature 37 °C
Column temperature 50 °C
Coverslip type 0.1 mL
Filter Pore Size 10 μm
Full Flow Filter Pore Size 10 μm
Paddle Volume 1 L
Printer Optional internal thermal printer
Sample volume 5 µL
Syringe Filter Pore Size 0.45 μm
Vessel Volume 1 L
Vial Volume 2 mL
Dissolution medium mfr 0.1 mL
Stirring rate mfr 75 rpm
Needle wash mfr 3 s
Max-Light cartridge cell (DAD 1) mfr 60 mm
Max-Light cartridge cell (DAD 2) mfr 3.7 mm
Basket mesh mfr 40 mesh
Thermal printer paper pack size mfr 10 rolls/pk
Safety standards mfr IEC, EN, CSA, UL UL
Evaporation mfr 1 conditions
Spindle - Speed range mfr 10 RPM
Spindle - Speed Accuracy mfr 1 RPM
Spindle - Speed Selection mfr Via touch screen screen
Sampling mfr Manifold option available for automated sampling
Drive unit lift mfr Motorized drive drive
Optional features mfr Manual drive unit lift, Dosage Delivery Module ,
Instrument Control via PC mfr 1.7
Maintenance and safety-related features mfr Spindle hours are tracked in FW FW
GLP features mfr Errors can be tracked via print-out or in the so
Housing mfr All materials are recyclable recyclable
Solvents mfr A) Water B) Acetonitrile Acetonitrile
Paracetamol content mfr 200 mg
Caffeine content mfr 25 mg
Chlorphenamine content mfr 2.5 mg
Complete dissolution stirring time mfr 90 minutes
Syringe filter diameter mfr 25 mm
Vial dimensions mfr 12 mm
Cap size mfr 12 mm
Condition mfr Performance may vary depending on environmental
Spindle Speed mfr 10.0 RPM
Test Duration mfr 10...999
Final Spin Duration mfr 10...001
Tester ID mfr 12 characters
Comm ID mfr 01...99
Sample Point Alarm mfr 05...000
Number Active mfr Sets the number of vessels used in a test. The r
Dimensions, Weight & Mechanical Design
Weight 54.4 kg
Upper receptor shaft length mfr 21 in.
Bent cannula gauge length mfr 7.75 in.
Dimensions (height × width × depth) mfr 680 mm
Cannula gauge length mfr 7.75 in.
Capsule weight wire length mfr 50 ft.
Cannula length mfr 7.75 in.
Electrical, Power & Drive Specifications
Line voltage mfr 115
Line frequency mfr 60 Hz
ISM Classification mfr 1 A
Power consumption mfr 10.0 A
Needle Wash mfr 3 A
Interfaces, Connections & Control
Local User Interface mfr LCD display with integral touch screen screen
Serial Number mfr 15 characters
Materials, Media & Process Compatibility
Capsule weight wire diameter mfr 0.032 in.
Spindle shaft material mfr Stainless steel steel
Dissolution media volume mfr 900 mL
Media Volume mfr 100...2000 mL
Operating Conditions, Environment & Protection
Operating Altitude 2000 m
Humidity mfr 80 % r.h.
Water bath - Probe accuracy mfr 0.1 °C

Description

Construction and series positioning

The 708-DS is a benchtop dissolution apparatus available in two configurations: the 6-position G7910A and the 8-position G7911A. It uses 1 L vessels and supports USP Apparatus 1 (rotating basket, 40-mesh) and Apparatus 2 (rotating paddle). The drive unit lift is motorized as standard, with a manual lift option. The unit weighs 54.4 kg dry and measures 680 × 622 × 590 mm (H × W × D). Within the Agilent dissolution series, the 708-DS is the standard multi-position apparatus, while the 400-DS Apparatus 7 and BIO-DIS are reciprocating cylinder systems for extended-release and special dosage forms, and the NanoDis System is for nanoparticle dissolution.

Behavior under load and media

Spindle speed is selectable from 10 to 250 RPM via the touch screen, with accuracy of ±1% above 25 RPM and ±2% from 10 to 25 RPM. The water bath maintains ambient +5 to 55 °C with probe accuracy of ±0.1 °C; the dissolution medium is typically held at 37 °C. The bath temperature probe has a resistance of 2252 Ω at 25.0 °C and 1355 Ω at 37.0 °C, allowing field verification with an ohmmeter. The heater/circulator must never be turned on before the bath is filled with purified water to the correct level.

Installation, control and operation

Line voltage is 115/230 V~ ±10% at 60/50 Hz with no external voltage selector or accessible fuses; power consumption is 2.0/1.0 A without heater and 10.0/5.0 A with the Wide Input Range Heater. The instrument requires 10 cm clearance on either side, 20 cm at the rear, and 30 cm above when fully raised. It operates in ambient 5–40 °C and up to 80% RH non-condensing, indoors up to 2000 m altitude. Control is via LCD touch screen or PC software (Dissolution Workstation 1.7+, Cary WinUV Dissolution 5.3+, Dissolution Workflow Manager 1.0 SR1). Optional features include Dosage Delivery Module, automated sampling manifold, AutoTemp, resident sampling cannulas, handheld temperature probe, and internal thermal printer.

Integration & Operating Notes

Electrical
Circuit must handle 10 A with heater at 115 V

The Wide Input Range Heater draws up to 10.0 A at 115 V (5.0 A at 230 V), while the base instrument draws 2.0/1.0 A. Size the branch circuit and any UPS for the heater-inclusive current, and note that protection is electronic and internal—no externally accessible fuses are provided.

Operator's Manual, Specifications page 13; Power Considerations page 10
Mounting
Bench must support 54.4 kg and allow 30 cm above

The dry weight with vessels and paddles is 54.4 kg. Provide 10 cm clearance on either side, 20 cm at the rear, and at least 30 cm above the unit when fully raised for ventilation and access to liquid and electrical connections. Operating height is 680 mm; clearance height is 991 mm.

Operator's Manual, Bench Space page 12; Physical specifications page 13
Process Limits
Bath setpoint cannot go below ambient +5 °C

The water bath temperature range is ambient +5 to 55 °C. Setpoints below ambient +5 °C are not supported. For 37 °C dissolution testing, ensure the ambient temperature does not exceed 32 °C to stay within the bath's control range.

Operator's Manual, Performance specifications page 14
Control & I/O
Remove retaining ties before power-on self-test

Retaining ties secure the drive unit during shipment. They must be removed before turning on the instrument; failure to do so can cause significant damage during the initialization self-test. Verify removal before first power-up and after any relocation.

Operator's Manual, Turn On the 708-DS page 26
Maintenance
Bi-annual lubrication and belt inspection required

Every six months, lubricate the linear bearing upright supports with 3-in-1 oil and inspect the spindle drive belt for wear, fraying, and tension. These tasks can be performed without opening the main cover. Maintenance intervals may vary with instrument usage.

Operator's Manual, Bi-Annual Maintenance page 90
Maintenance
Rinse sample lines after each test to prevent carryover

When a sampling manifold is configured, flush the sample tubing after every test using a rinse tubing kit (17-1341) or rinse tray. Routine rinsing prevents carryover and blockage. Corrosive media require immediate rinsing with deionized water; laboratory dishwashers must not be used for shafts, baskets, or sample lines.

Operator's Manual, Rinsing Sample Lines page 92; Sample Line / Paddle / Basket Care

Selection & Integration Q&A

When should I choose the 708-DS over the 400-DS Apparatus 7 or BIO-DIS?

Choose the 708-DS for standard USP Apparatus 1 (basket) or 2 (paddle) dissolution testing with 1 L vessels, available in 6-position (G7910A) or 8-position (G7911A) configurations. The 400-DS Apparatus 7 and BIO-DIS are reciprocating cylinder systems for extended-release or special dosage forms; the NanoDis System is for nanoparticle dissolution.

What are the power and fusing requirements for installing the 708-DS?

The 708-DS accepts 115/230 V~ ±10% at 60/50 Hz with no external voltage selector and no externally accessible fuses. Power draw is 2.0/1.0 A without heater and 10.0/5.0 A with the Wide Input Range Heater. Provide an external branch circuit rated for the heater-inclusive current; internal protection is electronic.

What maintenance tasks and consumables are required beyond the base delivery?

Bi-annual maintenance includes lubricating linear bearing upright supports with 3-in-1 oil and inspecting the spindle drive belt. Consumables not included: rinse tubing kit (17-1341) or rinse tray for sample lines, thermal paper rolls (58 mm wide × 13.65 m max) for the optional printer, and replacement temperature probes (tolerance ±0.2 °C).

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Entire Workflow from Dissolution
Brochure ·633,4 КБ ·5 pp.
View PDF
PDF
Top Supplies for Agilent 708-DS Dissolution Systems
Brochure ·170,3 КБ ·4 pp.
View PDF
PDF
Agilent 708-DS Dissolution Apparatus Operator's Manual
Manual ·120 pp.
View PDF