No photo available yet
Agilent Technologies TwisTorr 304 FS
Agilent Technologies TwisTorr 304 FS — Turbomolecular Vacuum Pumps with Motor power 0.3 kW.
The TwisTorr 304 FS is a 60,000 rpm turbomolecular pump with a three-phase 54 Vac motor, delivering 250 l/s N2 and 220 l/s H2 pumping speed. It is distinguished within its series by a 300 VA motor and a 3-minute run-up, with cooling mandatory via water or an optional fan. It serves as the high-vacuum stage backed by a recommended forepump, achieving base pressure below 1×10⁻¹⁰ mbar with a ConFlat flange.
Direct drop-in replacements
Full specifications
| Motor power | 300 VA |
| Base Pressure With Recommended Forepump | 1 mbar |
| Pumping Speed Ar | 250 l/s |
| Pumping Speed H2 | 220 l/s |
| Pumping Speed He | 255 l/s |
| Pumping Speed N2 | 250 l/s |
| Rotational speed | 60000 rpm |
| Run-up time | 3 minutes |
| Max gas throughput at full rotational speed N2 mfr | 170 sccm |
| Max gas throughput at full rotational speed Ar mfr | 110 sccm |
| Compression ratio N2 mfr | 1 |
| Compression ratio He mfr | 1 |
| Compression ratio H2 mfr | 1.5 |
| Compression ratio Ar mfr | 1 |
| Operating position mfr | Any Any |
| Lubricant mfr | Permanent lubrication lubrication |
| Installation category mfr | II II |
| Compression ratio Total N2 mfr | 1 |
| Compression ratio Total He mfr | 1 |
| Compression ratio Total H2 mfr | 1.5 |
| Compression ratio Total Ar mfr | 1 |
| Max gas throughput N2 mfr | 170 sccm |
| Max gas throughput Ar mfr | 110 sccm |
| Max oper. power: mfr | 150 cooling |
| Pump CFF 6’’ mfr | 7.5 kg |
| Pump CFF 8’’ mfr | 9.7 kg |
| Fan extension cable mfr | 9699949 |
| Vibration isolator, CF 6" mfr | 9699334 |
| Vibration isolator, CF 8" mfr | 9699335 |
| Purge valve 10 SCCM NW16KF – M12 mfr | 9699239 |
| Purge valve 10 SCCM 1⁄4 Swagelok – M12 mfr | 9699240 |
| Purge valve 20 SCCM NW16KF – M12 mfr | 9699241 |
| Purge valve 20 SCCM 1⁄4 Swagelok – M12 mfr | 9699242 |
| Purge valve 10 SCCM 1⁄4 Swagelok – 1⁄4 Swagelok mfr | 9699232 |
| Purge valve 20 SCCM 1⁄4 Swagelok – 1⁄4 Swagelok mfr | 9699236 |
| Active Gauges mfr | Ask Agilent for details details |
| Maximum vent pressure mfr | 2 bar |
| Shelf life mfr | 12 months |
| Fixing torque (ISO-K M10 clamps) mfr | 22 Nm |
| Weight | 5.5 kg |
| Stand-by power mfr | 10 W |
| Max operating power mfr | 150 W |
| Max input pwr: mfr | 300 VA |
| Stand-by pwr: mfr | 10 W |
| Inlet flange | 8 |
| Foreline flange mfr | 16 NW |
| Compression ratio Foreline/Side Port N2 mfr | 1 |
| Compression ratio Foreline/Side Port He mfr | 2.5 |
| Compression ratio Foreline/Side Port H2 mfr | 1 |
| Compression ratio Foreline/Side Port Ar mfr | 1 |
| Controller AG Rack base mfr | 3506 |
| Controller AG Rack with RS232/485 mfr | 3506 |
| Controller AG Rack with RS232/485 & Profibus mfr | 3506 |
| Controller V304 onboard 24Vdc mfr | 3507 |
| Controller V304 onboard 100-240Vac mfr | 3507 |
| Mains cable NEMA Plug, 3m long mfr | 9699958 |
| Mains cable European Plug, 3m long mfr | 9699957 |
| Serial cable and Navigator Software mfr | 9699883 |
| Inlet screen CFF 6’’ mfr | 9699302 |
| Inlet screen CFF 8’’ mfr | 9699304 |
| Air cooling kit for On-board Controller mfr | 3500 |
| Air cooling kit for Rack AG Controller mfr | 3500 |
| Bracket for On-Board Controller side mounting mfr | 3500 |
| Vent flange, NW 10 KF / M8 mfr | 9699108 |
| Vent valve for AG Rack Controller mfr | 9699844 |
| Vent valve for on-board Controller mfr | 9699834 |
| TwisTorr 304FS foreline flange KF25 mfr | 3500 |
| Required flange torque capability mfr | 1000 Nm |
| Coolant Water Minimum Flow | 50 l/h |
| Coolant water pressure mfr | 3...5 bar |
| Bakeout Temperature | 80 °C |
| Max Altitude | 3000 m |
| Noise Pressure Level | 50 dB |
| Storage temperature | -40 °C |
| Cooling requirements mfr | Natural convection convection |
| Pollution degree mfr | 2 |
| Max gas throughput at full rotational speed (with recommended forepump) N2 — Air cooling mfr | 170 sccm |
| Max gas throughput at full rotational speed (with recommended forepump) Ar — Air cooling mfr | 110 sccm |
| Max gas throughput at full rotational speed (with recommended forepump) N2 — Water cooling mfr | 170 sccm |
| Max gas throughput at full rotational speed (with recommended forepump) Ar — Water cooling mfr | 110 sccm |
| Water cooling kit mfr | 9699337 |
| Plastic water cooling kit mfr | 9699347 |
| Thermistor trip high mfr | 60 °C |
| Pump ISO 100 mfr | 5.5 kg |
| Pump ISO 160 mfr | 5.7 kg |
| Inlet screen, ISO 100 mfr | 3500 |
| Inlet screen, ISO 160 mfr | 3500 |
| Vibration isolator, ISO 100 mfr | 9699344 |
| Vibration isolator, ISO 160 mfr | 9699345 |
| ISO 100 K mfr | 10 clamps |
| ISO 160 K mfr | 10 clamps |
Description
Construction and series positioning
The TwisTorr 304 FS is a single-flow turbomolecular pump with a 60,000 rpm rotor driven by a three-phase motor at 54 Vac and 1010 Hz. It is the standard (non-split-flow) variant; the split-flow X3500-64010 is a separate configuration with its own performance table. The pump weighs 5.5 kg and can be mounted in any position, supported by its high-vacuum inlet flange. The inlet flange is available in ISO 100, ISO 160, CFF 6" and CFF 8" versions; the foreline is KF 16 NW as standard, with KF25 optional. The pump is permanently lubricated and requires no scheduled maintenance.
Performance and gas behavior
Pumping speeds are 250 l/s for N2 and Ar, 255 l/s for He, and 220 l/s for H2. Maximum gas throughput at full speed is 170 sccm for N2 and 110 sccm for Ar. Compression ratios are modest for light gases (2.5 for He, 1.5 for H2 total) and higher for N2, reflecting the pump's role as a high-vacuum stage rather than a light-gas compressor. Base pressure below 1×10⁻¹⁰ mbar is specified with the recommended forepump, measured per DIN 28 428 in a leak-free test dome 48 hours after bake-out with a ConFlat flange. Start-up to full speed takes less than 3 minutes.
Cooling, control and installation
Cooling is mandatory before operation: either water through channels in the pump body (minimum 50 l/h, +15 to +30 °C, 3–5 bar) or the optional air-cooling fan X3500-68011 (24 Vdc, 147 m³/h). Water cooling is mandatory if ambient temperature exceeds 35 °C. The pump connects to a compatible Agilent controller via a 6-pin connector (pins B, C, D for the 3-phase motor; pin E for ground). A thermistor near the upper bearing stops the pump with Error Code 2 above 60 °C and Error Code 1 below 0 °C; if the sensor is disconnected, the pump will not start. Soft Start mode, disabled by default, can be enabled to improve bearing grease distribution after a protracted stop. The pump must never be operated without the inlet flange connected to the chamber or blanked, and venting must use air or inert gas at less than 2 bar above atmospheric.
Integration & Operating Notes
The pump is supported solely by its high-vacuum flange; the base cannot be used for fixing. The system counter-flange and clamps must be rated for the rotor-failure torque of 1000 Nm. For ISO-K flanges, use four high-strength clamps such as Agilent IC63250DCMZ tightened to 22 Nm for both ISO 100 K and ISO 160 K.
The thermistor near the upper bearing is a safety interlock: if its connector is open, the controller will not start the pump. The sensor also cuts power automatically above 60 °C (Error Code 2) and below 0 °C (Error Code 1). Verify sensor continuity before troubleshooting a no-start condition, and ensure cooling is active before each start.
Natural convection or the optional fan may suffice only up to 35 °C ambient. Above that, water cooling must be used. The water circuit must deliver at least 50 l/h at +15 to +30 °C and 3–5 bar inlet pressure; electrical conductance must be ≤500 µS/cm, otherwise add up to 20% Ethyl-Glycole in closed circuits. During bake-out, water cooling must always be used.
Operating the pump with an open inlet exposes the spinning rotor to debris and can cause mechanical damage. Always attach the inlet to the vacuum chamber or fit a blank flange before starting. Venting must use air or inert gas free from dust or particles at less than 2 bar above atmospheric pressure; excessive vent pressure can damage the bearings.
Soft Start ramps the motor through 300, 400, 800 and 900 Hz steps to improve grease distribution in the bearings after a protracted stop. It is disabled by default and must be enabled via software or Remote I/O command. Use it when the pump has been idle for an extended period, but note that it lengthens the start-up sequence beyond the nominal 3 minutes.
The TwisTorr 304 FS requires no scheduled maintenance and must not be opened by the user. Before any work, disconnect mains, vent the pump, wait for the rotor to stop and for surface temperature to fall below 50 °C. In case of breakdown, contact Agilent service or use the advanced exchange service; a Request for Return sheet must be completed before shipping.
Selection & Integration Q&A
Choose the standard 304 FS for a single high-vacuum port with 250 l/s N2 and a KF 16 foreline. The split-flow X3500-64010 provides separate main and side port performance (Tab. 4) for applications needing two distinct pumping paths; its performance data must not be applied to the standard model.
Use a compatible Agilent controller: X3506-64001/2/3 (AG Rack) or X3507-64002/3 (V304 onboard). The 6-pin connector uses pins B, C, D for the 3-phase motor and pin E for ground; pins A and F are reserved. A separate earth conductor with M5 screw at 2 Nm and <0.1 Ω impedance is recommended.
Yes, any position is allowed, but the pump must be fixed by its inlet flange only (base fixing not permitted). Install indoors, 5–35 °C, 0–90% RH non-condensing, max altitude 3000 m. Above 35 °C ambient, water cooling is mandatory. Avoid dust, aggressive gases, explosive atmospheres and strong magnetic fields without shielding.
Base delivery excludes cooling (water kit 9699337/9699347 or fan X3500-68011), inlet screen, vent valve (9699834 or 9699844), vibration isolators and purge valve. The pump itself requires no scheduled maintenance; any service must be by authorized personnel via advanced exchange, with a Request for Return sheet.