Peter Huber Kältemaschinenbau KISS K6 Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau KISS K6 Kälte-Badumwälzthermostat

Peter Huber Kältemaschinenbau KISS K6 Kälte-Badumwälzthermostat — Thermostatic baths with Noise level 57 dB.

The KISS K6 is a refrigerated bath circulator with a 4.5 L bath and a non-variable pressure/suction pump, covering -25 to 200 °C with ±0.05 K stability. Its 0.2 kW cooling at 20 °C falls to 0.05 kW at -20 °C, so it suits small sub-ambient and high-temperature duties. It serves as a temperature-controlled source for external circuits via M16x1 connections.

Noise level
57 dB

Direct drop-in replacements

Eppendorf
MixMate ®
58%
match
⚖️ Compare

Peter Huber Kältemaschinenbau KISS Series — Model Comparison & Selection

Model Noise level, dBPower supplyPump connectionTemperature range, °CWeight, kgFilling volume, m³
KISS K6 Kälte-Badumwälzthermostat Current 57208-240V 1~/2~ 50/60HzM16x1 AG male-25–200250.002
KISS K12 Kälte-Badthermostat —110-115V 1~ 50/60Hz(Option, #19607)-20–20028—
KISS K20 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.01
KISS K25 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200390.015
KISS K6s Kälte-Badumwälzthermostat 57208-240V 1~/2~ 50/60HzM16x1 AG male-25–200250.002

Full specifications

Core Performance & Functional Specifications
Noise level 57 dB(A)
Filling volume 2 l (min)
Maximum delivery flow rate 14 l/min (pressure pump max, 0, 25 bar)
Pump connection M16x1 AG male
Pump speed adjustable no
Pump type Pressure and suction pump, speed not variable
Temperature range -25...200 °C
Temperature stability mfr ±0, 05 K
Alarm message mfr optic, acoustic
Dimensions (W x D x H) mfr 210 x 400 x 546 mm
Heating capacity mfr 1, 6 - 2, 1 kW
Delivery capacity suction pump max. mfr 10, 5 l/min; 0, 17 bar
Bath opening WxD/BD mfr 140 x 120 / 150
Bath volume mfr 4, 5 l
Dimensions, Weight & Mechanical Design
Weight 25 kg (net)
Electrical, Power & Drive Specifications
Power supply 208-240V 1~/2~ 50/60Hz
Current consumption max. mfr 10 + 1, 4 A (max)
Interfaces, Connections & Control
Controller KISS
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Cooling capacity mfr 20:0, 2; 0:0, 15; -10:0, 1; -20:0, 05 °C / kW
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The KISS K6 is the smallest refrigerated bath in the KISS series, with a 4.5 L bath volume and a working range of -25 to 200 °C. Unlike the larger K12, K15, K20 and K25 siblings, it uses a single-stage air-cooled refrigeration circuit with R-290 (A3, H220) refrigerant. The pump is a pressure and suction type with fixed speed, delivering up to 14 l/min at 0.25 bar on the pressure side and 10.5 l/min at 0.17 bar on the suction side. Wetted parts are stainless steel or plastics, and the pump connections are M16x1 male. The unit is rated IP20 and safety class III/FL per DIN 12876, with adjustable overtemperature protection.

Behavior under load

Cooling capacity is strongly temperature-dependent: 0.2 kW at 20 °C, 0.15 kW at 0 °C, 0.1 kW at -10 °C and 0.05 kW at -20 °C, all referenced to a 20 °C room. Heating capacity is 1.6–2.1 kW. Temperature stability is ±0.05 K. The fixed-speed pump means flow cannot be trimmed; external pressure drop must be matched by circuit design. The plastic pump is not resistant to acetone or polyglycols, so fluid selection must respect wetted-material limits. For sub-zero operation, water should be mixed with glysantine or ethylene glycol for freeze protection.

Installation and operation

The unit requires 208–240 V, 1~ or 2~, 50/60 Hz with a 10–16 A fuse; voltage tolerance is ±5% and frequency ±2%, but the combination of -5% voltage with +2% frequency is not allowed. A country-specific plug must be specified at ordering. Ambient limits are 5–40 °C, indoor dry use only. The air-cooled condenser needs clearance and cool intake air. Control is via the KISS controller with OLED display, USB device and RS232 interfaces, and optic/acoustic alarm. Class A Group 1 per EN55011.

Integration & Operating Notes

Electrical
Supply tolerance envelope excludes -5% V with +2% Hz

The unit accepts 208–240 V at 50/60 Hz with ±5% voltage and ±2% frequency tolerance, but the combination of -5% voltage and +2% frequency is explicitly not allowed. Verify the site feed against this combined envelope before connection, and provide a 10–16 A fuse.

Datasheet, Technical Specifications, page 2
Process Limits
Cooling capacity derates at low temperature and warm ambient

Published cooling capacity (0.2 kW at 20 °C down to 0.05 kW at -20 °C) is valid at 20 °C room temperature and may drop as ambient rises. Size the heat load against the capacity at the actual setpoint and expected room temperature, not the 20 °C figure.

Datasheet, Technical Specifications, page 1–2
Cooling & Media
Plastic pump incompatible with acetone and polyglycols

The pump is made of plastic and is not resistant to acetone or polyglycols (depending on manufacturer). Select heat transfer fluid accordingly; for sub-zero duty, mix water with glysantine or ethylene glycol for freeze protection rather than using pure water.

Datasheet, Technical Specifications, page 2
Mounting
Indoor dry location with 5–40 °C ambient and air-cooled clearance

IP20 rating and 5–40 °C ambient limit restrict the unit to indoor dry locations. The refrigeration machine is air-cooled, so the installation site must provide adequate clearance and cool intake air; confined or hot spaces will reduce cooling capacity.

Datasheet, Technical Specifications, page 2
Electrical
Country-specific plug must be specified at ordering

Single and two-phase devices are delivered with a power cable and a country-specific plug. The plug type must be stated when ordering; otherwise the unit may arrive with a plug unsuitable for the installation country.

Datasheet, Technical Specifications, page 2
Process Limits
Flammable R-290 refrigerant requires ignition-source review

The refrigerant is R-290 (A3, H220), classified as flammable. Installation location and room ventilation must be reviewed for ignition sources and refrigerant concentration limits according to local safety regulations.

Datasheet, Technical Specifications, page 1

Selection & Integration Q&A

When should I choose the KISS K6 over the larger K12, K15, K20 or K25 siblings?

Choose the K6 for small bath volumes (4.5 L) and duties within its cooling curve (0.2 kW at 20 °C, 0.05 kW at -20 °C). The larger siblings offer more cooling and bath volume; the K20/K25 also run quieter at 50 dB(A) versus 57 dB(A) for the K6.

What are the requirements for connecting the K6 to an external circuit?

The pump connections are M16x1 male. The fixed-speed pressure/suction pump delivers up to 14 l/min at 0.25 bar pressure and 10.5 l/min at 0.17 bar suction. External circuit pressure drop must be matched to this fixed flow; the pump speed cannot be adjusted.

What ambient and duty-cycle limits apply to the K6 installation?

Operate indoors, dry, between 5 and 40 °C ambient, IP20. The air-cooled condenser needs clearance and cool intake air. Cooling capacity is rated at 20 °C room and derates at higher ambient. Supply must be 208–240 V, 1~/2~, 50/60 Hz with a 10–16 A fuse.

What maintenance and consumables are required, and what is not included in base delivery?

The base delivery includes the unit with power cable and country-specific plug (must be specified). Heat transfer fluid is not included; for sub-zero use, mix water with glysantine or ethylene glycol. Wetted parts are stainless steel or plastics; avoid acetone and polyglycols with the plastic pump.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
KISS K6 Immersion Circulator
Datasheet ·136,0 КБ ·2 pp.
View PDF