Peter Huber Kältemaschinenbau Ministat 125w Kompakter Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Ministat 125w Kompakter Kälte-Badumwälzthermostat

Peter Huber Kältemaschinenbau Ministat 125w Kompakter Kälte-Badumwälzthermostat — Thermostatic baths.

The Ministat 125w is a compact refrigerated bath circulator with a 2.7 l bath, a variable-speed pressure/suction pump, and a water-cooled R-290 first stage, covering -25 to 150 °C with ±0.02 K stability. It is the water-cooled modification in the Ministat series, requiring a 5–20 °C cooling water supply at ≥2 bar. It serves as a precise temperature source for external circuits and bath immersion duties.

Direct drop-in replacements

Eppendorf
MixMate ®
47%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Ministat Series — Model Comparison & Selection

Model Max delivery flow rate, liter/minutePower supplyTemperature range, °CWeight, kgAmbient temperature limit, °CController
Ministat 125w Kompakter Kälte-Badumwälzthermostat Current 22110-120V 1~ 60Hz-25–150265–40Pilot ONE
Ministat 230w Kompakter Kälte-Badumwälzthermostat 16208-240V 1~/2~ 50/60Hz-40–200365–40Pilot ONE
Ministat 240w Kompakter Kälte-Badumwälzthermostat 22220-240V 1~/2~ 50/60Hz-45–200415–40Pilot ONE

Full specifications

Core Performance & Functional Specifications
Heating capacity 1 kW
Maximum delivery flow rate 22 l/min (pressure pump; 0, 7 bar)
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range -25...150 °C
Temperature stability mfr ±0, 02 K
Dimensions (W x D x H) mfr 225 x 370 x 429 mm
Delivery capacity suction pump max. mfr 16 l/min; 0, 4 bar
Bath opening WxD/BD mfr 178 x 80 / 120 mm
Bath volume mfr 2, 7 l
Filling volume min. mfr 2, 1 l (min)
Dimensions, Weight & Mechanical Design
Weight 26 kg
Electrical, Power & Drive Specifications
Fuse rating 16 A
Power supply 110-120V 1~ 60Hz
Current consumption max. mfr 11, 5 A (max)
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Refrigerant R-290 (A3, H220) (1st stage)
Maximum viscosity 50 mm²/s
Refrigerant quantity 1st stage mfr 0, 48 kg (1st stage)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 11 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 2 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Cooling capacity mfr 100:0, 3; 20:0, 3; 0:0, 2; -10:0, 15; -20:0, 1 k (at °C bath temperature)
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Hydraulics and media

Installation and control

Integration & Operating Notes

Cooling & Media
Site cooling water must be 5–20 °C at ≥2 bar

The water-cooled first stage is designed for inlet water between 5 and 20 °C with at least 2 bar differential pressure and no more than 6 bar. Verify the house supply before installation; warmer water reduces cooling capacity and raises the required flow rate.

Datasheet, Specifications (installation
Cooling & Media
Insulate cooling-water pipes below dew point

If the cooling water temperature can fall below the condensation point, the internal cooling water pipes must be insulated to prevent condensation, corrosion and electrical risk. The circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used.

Datasheet, Specifications (installation
Electrical
Voltage and frequency tolerances are coupled

The supply may deviate ±5% in voltage only with a simultaneous ±2% frequency tolerance, and combinations such as -5% voltage with +2% frequency are not allowed. Size the 16 A circuit for the 11.5 A maximum current and confirm the country-specific plug at ordering.

Datasheet, Specifications (compliance/installation
Process Limits
Bath fluid viscosity capped at 50 mm²/s

The pump can handle bath fluid up to 50 mm²/s kinematic viscosity. At the low end of the -25…150 °C range, fluid choice must keep viscosity within this limit or circulation and heat transfer will degrade.

Datasheet, Specifications (operation
Mounting
IP20 unit needs a dry, ventilated location

With IP20 protection and a 5–40 °C ambient limit, the unit must be sited away from water spray and in a ventilated area. Because the refrigerant is R-290 (A3, 0.48 kg), the installation area must account for flammability and ignition sources.

Datasheet, Specifications (installation/safety
Control & I/O
External Pt100 and digital interfaces available

The Pilot ONE controller provides Ethernet, USB host and device, and RS232, plus a Pt100 connection for an external sensor. These allow remote monitoring and control of the bath or an external circuit.

Datasheet, Specifications (interfaces

Selection & Integration Q&A

When should I choose the Ministat 125w over the air-cooled Ministat 230/240w siblings?

Pick the 125w when a 5–20 °C cooling water supply at ≥2 bar is available and heat rejection to room air is unwanted. The air-cooled siblings suit sites without cooling water; the 230 is noted for its 59 dB(A) noise signature.

What cooling-water supply and piping must I provide for the 125w?

G1/2 male connections, inlet water 5–20 °C, differential pressure ≥2 bar and ≤6 bar. Consumption is 11 l/h at 15 °C water with 0 °C flow; insulate internal pipes if water can drop below the dew point, and use water compatible with copper, 1.4401, MS, PA, PPE, PTFE and EPDM.

What are the electrical and environmental limits for installing this unit?

110–120 V 1~ 60 Hz, 11.5 A max, 16 A fuse; voltage ±5% only with simultaneous frequency ±2%, and some combinations are not allowed. Ambient 5–40 °C, IP20, EN55011 Class A Group 1; specify the country-specific plug at ordering.

What maintenance and accessories are not included in the base delivery?

The base unit ships without a country-specific power plug unless specified, and without external Pt100 sensor or interface cables. Bath fluid and any displacement rack are also not included; note that a rack reduces usable bath capacity from 2.7 l to 1.3 l.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Ministat 125w Immersion Circulator
Datasheet ·95,3 КБ ·2 pp.
View PDF