The Ministat 125w is a compact refrigerated bath circulator with a 2.7 l bath, a variable-speed pressure/suction pump, and a water-cooled R-290 first stage, covering -25 to 150 °C with ±0.02 K stability. It is the water-cooled modification in the Ministat series, requiring a 5–20 °C cooling water supply at ≥2 bar. It serves as a precise temperature source for external circuits and bath immersion duties.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Ministat Series — Model Comparison & Selection
| Model | Max delivery flow rate, liter/minute | Power supply | Temperature range, °C | Weight, kg | Ambient temperature limit, °C | Controller |
|---|---|---|---|---|---|---|
| Ministat 125w Kompakter Kälte-Badumwälzthermostat Current | 22 | 110-120V 1~ 60Hz | -25–150 | 26 | 5–40 | Pilot ONE |
| Ministat 230w Kompakter Kälte-Badumwälzthermostat | 16 | 208-240V 1~/2~ 50/60Hz | -40–200 | 36 | 5–40 | Pilot ONE |
| Ministat 240w Kompakter Kälte-Badumwälzthermostat | 22 | 220-240V 1~/2~ 50/60Hz | -45–200 | 41 | 5–40 | Pilot ONE |
Full specifications
| Heating capacity | 1 kW |
| Maximum delivery flow rate | 22 l/min (pressure pump; 0, 7 bar) |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | -25...150 °C |
| Temperature stability mfr | ±0, 02 K |
| Dimensions (W x D x H) mfr | 225 x 370 x 429 mm |
| Delivery capacity suction pump max. mfr | 16 l/min; 0, 4 bar |
| Bath opening WxD/BD mfr | 178 x 80 / 120 mm |
| Bath volume mfr | 2, 7 l |
| Filling volume min. mfr | 2, 1 l (min) |
| Weight | 26 kg |
| Fuse rating | 16 A |
| Power supply | 110-120V 1~ 60Hz |
| Current consumption max. mfr | 11, 5 A (max) |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Refrigerant | R-290 (A3, H220) (1st stage) |
| Maximum viscosity | 50 mm²/s |
| Refrigerant quantity 1st stage mfr | 0, 48 kg (1st stage) |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 11 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 2 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Cooling capacity mfr | 100:0, 3; 20:0, 3; 0:0, 2; -10:0, 15; -20:0, 1 k (at °C bath temperature) |
| Safety class | III / FL |
Description
Construction and series positioning
- The Ministat 125w is a water-cooled refrigerated bath circulator: a 2.7 l bath (178 x 80 mm opening, 120 mm depth) with a Pilot ONE controller, a variable-speed pressure and suction pump, and an R-290 (A3) first-stage refrigerant circuit cooled by site water rather than air.
- Within the Ministat series it is the water-cooled variant; the air-cooled siblings (Ministat 230, 230w, 240w) differ in cooling architecture and, for the 230, in noise signature. Choose the 125w where a cooling water supply is available and heat rejection to room air is undesirable.
- Heating capacity is 1 kW; cooling capacity falls from 0.3 kW at 100 °C and 20 °C to 0.2 kW at 0 °C, 0.15 kW at -10 °C and 0.1 kW at -20 °C, defining the usable load at low temperatures.
Hydraulics and media
- The pressure pump delivers up to 22 l/min at 0.7 bar; the suction pump up to 16 l/min at 0.4 bar. Pump speed is adjustable, so external circuit pressure drop can be matched to the duty point.
- Bath fluid kinematic viscosity must not exceed 50 mm²/s; this constrains fluid selection at the low end of the -25…150 °C range.
- With a displacement rack installed, usable bath capacity drops from 2.7 l to 1.3 l; minimum filling volume is 2.1 l.
Installation and control
- Cooling water: G1/2 male connections, inlet 5–20 °C, differential pressure ≥2 bar, maximum pressure 6 bar; consumption 11 l/h at 15 °C water with 0 °C flow. Cooling power drops and water flow demand rises as inlet water warms; pipes must be insulated if water can fall below the dew point.
- Electrical: 110–120 V 1~ 60 Hz, maximum current 11.5 A, 16 A fuse; voltage tolerance ±5% only with simultaneous frequency tolerance ±2%, and some combinations (e.g. -5% voltage with +2% frequency) are not permitted. A country-specific plug must be specified at ordering.
- Control and I/O: Pilot ONE controller with Ethernet, USB host and device, RS232, and a Pt100 external sensor connection. Protection is IP20; ambient range 5–40 °C. The unit is EN55011 Class A, Group 1, and the R-290 charge (0.48 kg) requires attention to ignition sources and ventilation.
Integration & Operating Notes
The water-cooled first stage is designed for inlet water between 5 and 20 °C with at least 2 bar differential pressure and no more than 6 bar. Verify the house supply before installation; warmer water reduces cooling capacity and raises the required flow rate.
If the cooling water temperature can fall below the condensation point, the internal cooling water pipes must be insulated to prevent condensation, corrosion and electrical risk. The circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used.
The supply may deviate ±5% in voltage only with a simultaneous ±2% frequency tolerance, and combinations such as -5% voltage with +2% frequency are not allowed. Size the 16 A circuit for the 11.5 A maximum current and confirm the country-specific plug at ordering.
The pump can handle bath fluid up to 50 mm²/s kinematic viscosity. At the low end of the -25…150 °C range, fluid choice must keep viscosity within this limit or circulation and heat transfer will degrade.
With IP20 protection and a 5–40 °C ambient limit, the unit must be sited away from water spray and in a ventilated area. Because the refrigerant is R-290 (A3, 0.48 kg), the installation area must account for flammability and ignition sources.
The Pilot ONE controller provides Ethernet, USB host and device, and RS232, plus a Pt100 connection for an external sensor. These allow remote monitoring and control of the bath or an external circuit.
Selection & Integration Q&A
Pick the 125w when a 5–20 °C cooling water supply at ≥2 bar is available and heat rejection to room air is unwanted. The air-cooled siblings suit sites without cooling water; the 230 is noted for its 59 dB(A) noise signature.
G1/2 male connections, inlet water 5–20 °C, differential pressure ≥2 bar and ≤6 bar. Consumption is 11 l/h at 15 °C water with 0 °C flow; insulate internal pipes if water can drop below the dew point, and use water compatible with copper, 1.4401, MS, PA, PPE, PTFE and EPDM.
110–120 V 1~ 60 Hz, 11.5 A max, 16 A fuse; voltage ±5% only with simultaneous frequency ±2%, and some combinations are not allowed. Ambient 5–40 °C, IP20, EN55011 Class A Group 1; specify the country-specific plug at ordering.
The base unit ships without a country-specific power plug unless specified, and without external Pt100 sensor or interface cables. Bath fluid and any displacement rack are also not included; note that a rack reduces usable bath capacity from 2.7 l to 1.3 l.