Induztr
Engineering Digest Document Viewer Related Models

Agilent Online UV-Vis Dissolution

Contents

  1. Overview pp. 1-2
    Introduces Agilent Online UV-Vis dissolution solutions and benefits
  2. Config p. 3
    Table matching dissolution system models to configurations
  3. Overview pp. 4-6
    Describes Cary 60 online dissolution platform and setup
  4. Spec pp. 7-8
    Details fiber optic probe technology and ideal platform features
  5. Spec pp. 9-10
    Cary WinUV with SCM/SDA software capabilities
  6. pp. 11-12
    Contact and additional resource information
  7. p. 3 table, 23 rows

UV dissolution system selection chart

MulticellFiber Optic
General information
Software packageCary WinUVCary WinUV
Number of apparatus1 to 21 to 2
Spectrophotometer details
Wavelength range190-1100 nm190-1100 nm
Slit width1.5 nm1.5 nm
Lamp typeXenon flashXenon flash
Instrument designScanningScanning
Wavelength accuracy< ± 0.5 nm< ± 0.5 nm
Wavelength reproducibility± 0.1 nm± 0.1 nm
Photometric accuracy± 0.005 Abs (NIST 930D)± 0.005 Abs (NIST 930D)
Photometric noise± 0.0001 Abs± 0.0008 Abs
Baseline flatness< 0.001 Abs< 0.004 Abs
Stray light< 1.0% (198 nm)< 0.05% (220 nm)
< 0.05% (370 nm)< 0.05% (370 nm)
Online sampling system details
Closed loop samplingYesN/A
Path lengths (mm)0.2, 0.5, 1, 2, 5, 101, 2, 5, 10, 20
Automated dosage deliveryYesYes
Temperature monitoringYesYes
SamplingSimultaneousSequential
Sample filtration (min. pore size μm)0.2 or 0.45 μm (with 850-DS and optional filtration module)N/A
Sample archivalYes (with 850-DS Dissolution Sampling Station)N/A

Engineering Facts

13 facts · every claim linked to a page
Selection

The multicell changer supports one or two dissolution apparatus, each vessel having its own flow cell to avoid cross-contamination.

Determines whether a single or dual-apparatus setup is needed and confirms no sample carryover between vessels.

Source
“The multicell changer integrates one or two dissolution apparatus, providing each vessel with an individual flow cell to prevent cross-contamination.”
Selection

Fiber optic probes are mounted on a moveable manifold that ensures precise sampling position and minimizes hydrodynamic disturbance.

In-situ measurement avoids sample transport and disturbance, which matters when choosing between multicell and fiber optic configurations.

Source
“Analyzes samples in situ through fiber optic probes mounted on a moveable manifold that ensures precise sampling position and minimizes hydrodynamic disturbances.”
Selection

The multicell system accommodates eight flow cells per apparatus, allowing a blank, a standard, and six samples.

Defines the maximum number of vessels/standards that can be handled per apparatus in a multicell setup.

Source
“The multicell changer accommodates eight flow cells per dissolution apparatus, allowing for a blank, a standard, and six samples.”
Selection

Multicell flow cells are available in path lengths from 0.2 to 10 mm, while fiber optic tips range from 1 to 20 mm.

Path length selection directly affects absorbance range and sensitivity for the chosen concentration regime.

Source
“Available in either single- or dual-apparatus configurations, the system supports individual flow cells for a range of path lengths, from 0.2 to 10 mm.”
Maintenance

Fiber optic tips with path lengths from 1 to 20 mm are interchangeable without buying new probes.

Reduces consumable cost and allows changing path length without replacing the probe assembly.

Source
“Interchangeable tips with path lengths from 1 to 20 mm are substituted as needed without investing in new probes.”
Operation

The fiber optic system can take readings as often as every 45 seconds, suiting rapid time-point requirements.

Sets the minimum achievable sampling interval when planning dissolution time points.

Source
“Ideal for rapid time point requirements with the ability to take readings as often as every 45 seconds”
Maintenance

Fiber optic probes require only rinsing and wiping for cleaning, simplifying maintenance.

Low cleaning effort reduces downtime and operator error between runs.

Source
“Cleaning is simple, requiring only rinsing and wiping of the fiber optic probes and tips”
Operation

The xenon flash lamp only flashes when a reading is taken, giving zero warm-up time and low energy use.

No warm-up means immediate measurement readiness and lower operating cost.

Source
“The lamp only flashes when a reading is taken, resulting in zero warm-up time and very low electrical energy use and maintenance requirements.”
Maintenance

Average xenon lamp lifetime is about 10 years under typical laboratory usage.

Long lamp life lowers cost of ownership and reduces replacement planning.

Source
“average lamp lifetime is 10 years under typical laboratory usage conditions.”
Performance

The photometric range extends above 3.5 absorbance units with a correlation coefficient of 0.999.

High linear range supports concentrated or turbid samples without dilution.

Source
“the photometric range of the Agilent Cary 60 extends above 3.5 absorbance units with a correlation co-efficient of 0.999.”
Performance

The spectrophotometer can scan the entire 190–1100 nm range in under 3 seconds.

Fast scanning supports high-throughput multi-wavelength dissolution methods.

Source
“able to scan the entire wavelength range (190–1100 nm) in under 3 seconds”
Compliance

A 21 CFR Part 11 compliance package is included for electronic records and secure data storage.

Regulated labs need this for audit trails and electronic signatures.

Source
“21 CFR Part 11 compliance package is included for electronic records and secure data storage.”
Compliance

Cary WinUV with SCM/SDA supports all 21 CFR Part 11 requirements including audit trails and electronic signatures.

Confirms the software configuration needed for regulated electronic record keeping.

Source
“Using Agilent Cary WinUV with SCM/SDA provides support for all compliance requirements mandated by 21 CFR Part 11.”