Agilent Technologies RNA QC, DNA QC, Fragment Analyzer
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies RNA QC, DNA QC, Fragment Analyzer

Agilent Technologies RNA QC, DNA QC, Fragment Analyzer — Automated electrophoresis.

The Agilent Fragment Analyzer is a 12-capillary automated electrophoresis system for RNA and DNA QC, running 12 samples per day with 2 µL per well and a 50-minute analysis. Its 22 cm effective length and CCD detection support sizing from 75 bp with 5% accuracy and quantification over two orders of magnitude, delivering RQN quality scores on a 1–10 scale.

Agilent Technologies Fragment Analyzer Systems Series — Model Comparison & Selection

Model Measurement rangePlate capacityAverage sizing of 2 200 nt ivt mrna, ntLower marker size of hs rna kit, ntMaximum ivt mrna sample size analyzed, ntSample volume per well, microliter
RNA QC, DNA QC, Fragment Analyzer Current —12———2
5200 Fragment Analyzer ——23701560002
5300 Fragment Analyzer ——2370156000—
Fragment Analyzer Capillary Arrays —12———2
Ultra-high-throughput fragment analysis 5400 Fragment Analyzer 0.5nine plates2370156000—

Full specifications

Core Performance & Functional Specifications
Analysis time 50 min
Analysis Time6 55 min
Analysis Time7 40 min
Detector type Charged Coupled Device
Plate capacity 12 samples
Sample Input Volume 2 µL
Sample volume 2 µl
Sample Volume Per Well 2 µL
Separation resolution mfr 3 bp
Dilutions per sample mfr single single
Gel capacity mfr Two gel gel
Input concentration dynamic range mfr two orders of magnitude magnitude
RQN quality score scale mfr 1...10
Sizing range mfr 200 nt
Sizing accuracy mfr 20
Sizing precision mfr 20
Quantification accuracy mfr 30
Quantification precision mfr 20
Quantification range (DNA Fragment) mfr 0.1 ng/μL
Quantification range (DNA Smear) mfr 5 ng/μL
Guaranteed shelf life mfr 4 months
Kit size mfr 500 samples
DNA fragment concentration range mfr 0.5 ng/μL
Number of capillaries mfr 12 capillaries
RQN scale mfr 1...10
Capillary arrays mfr 12 arrays
n–1 resolution mfr 60 nt
Samples separated per run mfr 12 samples
Maximum queued samples mfr 288 samples
Capillary counts mfr 12 capillaries
sample separation time mfr 30 minutes
samples per run mfr 12 samples
capillary array options mfr 12 arrays
resolution range mfr 60 nt
Maximum DNA concentration mfr 600 sample
Limit of detection (S/N > 3) or sensitivity mfr 50 pg/μL
Input concentration range mfr 50 pg/μL
Quantitative range mfr 50 pg/μL
Maximum DNA concentration per fragment mfr 600 pg/μL
Maximum DNA concentration per total sample mfr 5 ng/μL
Final sample volume (Plasmid DNA Kit) mfr 24 μL
Dimensions, Weight & Mechanical Design
Effective Length 22 cm
Length 47 cm
Effective length (FA 12-Capillary Array Ultrashort) mfr 22 cm
Total length (FA 12-Capillary Array Ultrashort) mfr 47 cm
Effective length (FA 12-Capillary Array Short) mfr 33 cm
Total length (FA 12-Capillary Array Short) mfr 55 cm
Effective length (FA 12-Capillary Array Long) mfr 55 cm
Total length (FA 12-Capillary Array Long) mfr 80 cm
Standards, Safety & General Data
Standard Warranty One-year

Description

Construction and series positioning

Analytical behavior

Operation and control

Integration & Operating Notes

Process Limits
Input concentration must stay within two orders of magnitude

Samples outside the two-orders-of-magnitude input concentration dynamic range fall outside the validated quantification window; DNA fragment quantification begins at 0.5 ng/µL and DNA smear quantification at 5 ng/µL, so dilute concentrated samples or concentrate dilute ones before loading to keep results within the specified accuracy.

Factory Specification Sheet
Maintenance
Track reagent kit shelf life against the 4-month guarantee

Reagent kits are room-temperature stable but carry a guaranteed shelf life of 4 months; inventory must be rotated so that kits are consumed within this window, since expired reagents invalidate the RQN and quantification results without necessarily producing an instrument error.

Consumables and Analysis Kits datasheet, 5994-0407EN

Selection & Integration Q&A

When should I choose this 12-capillary Fragment Analyzer rather than the 5200, 5300 or 5400 models?

Pick the 12-capillary configuration for about 12 samples per day at 2 µL per well and 50 minutes per run. Move to the 5200 for 12–96 samples one to three times daily, the 5300 for one to two 96-well plates up to three times daily, or the 5400 when more than two plates per day with robotic automation are required.

Which consumables and adjacent parts must be ordered alongside the instrument?

The base instrument requires FA 12-capillary arrays (ultrashort, short or long) and application-specific DNA/NGS reagent kits such as DNF-473, DNF-474, DNF-476, DNF-477, DNF-905, DNF-910, DNF-915 or DNF-920; kits are sized for 500 samples and are room-temperature stable for 4 months, so plan array and kit inventory against the 12-sample daily throughput.

What sample concentration and volume constraints apply when preparing plates for this system?

Load 2 µL per well into the 12-capillary array and keep input concentration within the two-orders-of-magnitude dynamic range; DNA fragment quantification starts at 0.5 ng/µL and DNA smear at 5 ng/µL, so samples below these levels must be concentrated before the 50-minute run.

What maintenance and reagent-rotation practice does the documentation require?

Reagent kits must be consumed within their 4-month guaranteed shelf life despite room-temperature storage, and capillary arrays are consumable items replaced per the array type in use; no electrical, environmental or service-interval data are given in the verified facts, so the instrument manual must be consulted for the maintenance schedule.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Fragment Analyzer Systems: Reliable Results for Nucleic Acid Quality Control
Brochure ·4,8 МБ ·16 pp.
View PDF
PDF
Build Consistency into Your Lab's Workflow
Brochure ·12,3 МБ ·12 pp.
View PDF
PDF
Agilent Fragment Analyzer Consumables
Datasheet ·269,6 КБ ·5 pp.
View PDF