Agilent Technologies 20.60 CFM
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 20.60 CFM

Agilent Technologies 20.60 CFM — Vacuum technologies with Motor power 0.119 kW, Pumping speed 3.6 m³/h, Ultimate pressure 3.3 mbar.

The Agilent IDP 20.60 CFM is an oil-free dry scroll vacuum pump in the IDP family, configured for 100–120 VAC operation and rated at 60 L/min pumping speed with a 3.3 mbar base pressure. Its 0.16 HP motor and scroll architecture suit clean, hydrocarbon-free backing duty in analytical and laboratory vacuum systems.

Motor power
0.119 kW
Pumping speed
3.6 m³/h
Ultimate pressure
3.3 mbar

Agilent Technologies Idp Pumps Series — Model Comparison & Selection

Model Motor power, kWUltimate pressure, mbarPumping speed, m³/hBase pressure, mbarSupply frequency, HzOperating temperature, °C
20.60 CFM Current 0.1193.33.63.350—
26.48 CFM 0.1193.33.63.350—
6.0 CFM ——————
IDP-4 0.1192.43.63.3—5
Scroll Pumps for Helium Recirculation & Cryogenic Applications 0.119—3.63.3——

Full specifications

Core Performance & Functional Specifications
Motor power 0.16 HP
Pumping speed 60 L/min
Ultimate pressure 3.3 mbar
Base Pressure 3.3 mbar
IDP-3, 1φ, 220 V, 50/60 Hz mfr 3
IDP-3, 1φ, 115 V, 60 Hz mfr 3
IDP-3, 1φ, 100 V, 50/60 Hz mfr 3
IDP-3, 24 VDC mfr 3
IDP-3 tip seal replacement kit mfr 3 TS
Exhaust silencer kit mfr 110
Inlet trap mfr 25
Vibration isolation kit mfr 110 VIBISOKIT
Description mfr Part number number
IDP-7 mfr Vibration isolation kit kit
IDP-7 with inlet isolation valve mfr 3807
IDP-7/10 tip seal replacement kit mfr 3807
Purge kit mfr 3807
Gas ballast kit mfr 3807
IDP-10 mfr 3810
IDP-10 with inlet isolation valve mfr 3810
IDP-15 mfr 3815
IDP-15 with inlet isolation valve mfr 3815
IDP-15 tip seal replacement kit mfr 3815
IDP-35 mfr 3835
IDP-35 with Gauge mfr 3835
IDP-35 with Inlet Valve mfr 3835
IDP-35 with Inlet Valve and Gauge mfr 3835
Inlet HEPA filter, NW40 mfr 40
HEPA filter, NW25 mfr 25
IDP-45 mfr 3845
IDP-45 with Gauge mfr 3845
IDP-45 with Inlet Valve mfr 3845
IDP-45 with Inlet Valve and Gauge mfr 3845
inlet valve closing time mfr 20 milliseconds
inlet valve opening time mfr 10 seconds
Pirani gauge pressure range mfr -4...1000 mbar
IDP-3 mfr Vibration isolation kit kit
IDP-4 mfr Replacement filter element element
inlet valve opening time after power restore mfr 10 seconds
tip seal replacement duration mfr half an hour hour
IDP-4 maintenance duration mfr 15...30 minutes
integrated Pirani gauge pressure range mfr -4...1000 mbar
VPI valve isolation time mfr 20 ms
Electrical, Power & Drive Specifications
Supply frequency 50 Hz
Operating voltage mfr 100 VAC

Description

Construction and series positioning

Behavior and operating envelope

Installation and control

Integration & Operating Notes

Electrical
Match the 100–120 VAC motor variant to the site supply

This modification is the 100–120 VAC version; connecting it to a 200–240 VAC line, or ordering the wrong variant for a 230 V site, will damage the motor. Confirm supply voltage and frequency (50 or 60 Hz) before commissioning, since the family is documented at both operating points.

Factory Specification Sheet; IDP Dry Scroll Pump Family Brochure 5991-7583EN
Process Limits
3.3 mbar is the base pressure floor, not a working pressure

The inlet should be sized so the process gas load keeps the pump above its 3.3 mbar base pressure; operating continuously at the ultimate pressure with high gas throughput shifts the effective pumping speed below the 60 L/min rating. Use the family performance data for the actual inlet pressure of the application.

Factory Specification Sheet
Maintenance
Dry scroll tip seals are a wear item; schedule replacement

Because the pump is oil-free, the scroll tip seals and bearings are the consumable elements; base pressure drift upward from 3.3 mbar is the usual indicator that seal replacement is due. Plan seal and bearing service against the family maintenance schedule rather than waiting for a pressure failure.

IDP Dry Scroll Pump Family Brochure 5991-7583EN

Selection & Integration Q&A

When should I choose this 20.60 CFM modification instead of the 26.48 CFM or the smaller 5.4/6.0 CFM siblings?

Pick this one when the required backing speed is around 60 L/min at 3.3 mbar and the site is 100–120 VAC. The 26.48 CFM sibling shares the same 60 L/min and 3.3 mbar figures but is a different frame; the 5.4 and 6.0 CFM units suit lower-throughput chambers. Confirm the voltage variant before ordering.

What supply and adjacent hardware does this pump need from the system integrator?

A 100–120 VAC supply at the correct frequency (50 or 60 Hz) matched to the motor variant, plus inlet and exhaust fittings and any vibration isolation, none of which are included in the base delivery. The pump is a backing stage, so the high-vacuum pump and its flange interface must be specified separately.

What maintenance and consumables should be budgeted beyond the base pump?

Scroll tip seals and bearings are the wear items on this oil-free design; base-pressure drift above 3.3 mbar signals seal service. Inlet fittings, exhaust line, isolation mounts and any cabling are not in the base delivery and must be ordered as accessories. Follow the family brochure service schedule.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Agilent IDP Dry Scroll Vacuum Pumps
Brochure ·8,5 МБ ·10 pp.
View PDF
PDF
Agilent IDP-4 Dry Scroll Vacuum Pump
Brochure ·10,7 МБ ·11 pp.
View PDF